Singapore 1800 622 1156 | Intl +65 6415 5353 | Welcome to our new site - found a bug? Please report it here
CL3515R-Series
Rugged 11 Thermocouple type & mV Calibrator Thermometer
- USB and RS232 Interface with Windows Software
- 11 Types of Thermocouples: K, J, T, E, R, S, N, L, U, B, and C (in °C, °F, or K), and mV Signal Output
- NIST-Traceable Calibration Certificate with 7 Data Points (Included)
- 2 Channels for Thermocouple Temperature Measurement (T1 and T2) with Offset Adjustment
CL3515R-Series
From
SGD
808.94
Product Overview
- Accuracy ± 0.05 %
- Accuracy Detail ± (0.05% of reading + 0.5 °C)
- Additional Features 1 NIST Calibration Certificate
- Battery life 17 hr
- Display Type LCD
- Display Type Note 4-digit LCD with maxreading of 19999
- Electrical Connection Miniature Female Thermocouple Connector
- Height 192 mm
- Length 52.5 mm
- Milli Volt Accuracy ± 0.025% + 1 digit
- Milli Volt Resolution 10 µV
- Millivolt Range -25 to 75 mV
- Number of Channels 2
- Operating Temperature, Min 0 °C
- Relative Humidity, Max 0.7
- Relative Humidity, Min 0%
- Series ID CL3515R-Series
- Storage Temperature, Max 60 °C
- Storage Temperature, Min -20 °C
- Storage Temperature Note -20 to 60°C (-4 to 140°F ),<70%RH
- Supply Voltage 9 Vdc
- Temperature Coefficient 0.1 times the applicable accuracy specification per °C from 0 to 18°C and 28 to 50°C (32 to 64°F and 82 to 122°F)
- Temperature Measurement, Max 2316 °C
- Temperature Measurement, Min 0 °C
- Thermocouple Simulation Resolution Range Thermocouple Simulation Range: Resolution: 0.1° (1° for R/S/B/C); Accuracy*: ± (0.3°C + 10 µV)
- Thermocouple Type J, K,T, E, N, R, S, U, B, C, L
- Type of RTD Signal None
- Weight 320 g
- Width 91 mm
Supported Applications
The Omega CL3515R is a rugged, portable digital calibrator thermometer designed for temperature calibration and measurement using external thermocouples. It supports K, J, T, E, R, S, N, L, U, B, and C type sensors as well as millivolt (mV) inputs. The instrument complies with the N.I.S.T. Monograph 175 Revised to ITS-90 standard.
Key functional capabilities include:
- Dual thermocouple input for simultaneous measurement of two sensors
- Adjustable T/C offset functionality
- Thermocouple simulation output with 0.1° resolution and ±(0.3°C + 10uV) accuracy
- mV signal generation ranging from -25.00mV to 75.00mV with 10uV resolution
Operating Conditions & Performance
The CL3515R operates within an ambient temperature range of 0°C to 50°C (32°F to 122°F) at less than 80% R.H. Storage is permitted from -20°C to 60°C (-4°F to 140°F) with humidity below 70% R.H.
Measurement ranges and accuracy vary by thermocouple type:
- K-TYPE: -200°C to 1372°C or -328°F to 2501°F
- J-TYPE: -210°C to 1200°C or -346°F to 2192°F
- T-TYPE: -250°C to 400°C or -418°F to 752°F
- E-TYPE: -250°C to 1000°C or -418°F to 1832°F
- R/S-TYPE: 0°C to 1767°C or 32°F to 3212°F (Resolution: 1°)
- N-TYPE: -200°C to 1300°C or -328°F to 2372°F
- L/U-TYPE: -200°C to 900°C (L) and -200°C to 600°C (U)
- B/C-TYPE: 600°C to 1820°C (B); 0°C to 2316°C or 32°F to 4200°F (C) (Resolution: 1°)
Accuracy specifications for K/J/T/E/L/U types are ±(0.05% rdg + 0.5°C) from -50°C to 1372°C, and ±(0.05% rdg + 1.0°C) below -50°C down to the minimum range limit. For R/S/B/C types, accuracy is ±(0.05% rdg + 2°C) from 0°C to 1767°C.
Electrical and Environmental Limits
- Safety Warning: Do not use when working voltages at the measurement surface exceed 24V AC or DC.
- Microwave Safety: Do not make temperature measurements in microwave ovens to avoid damage or burns.
- Battery Life: Approximately 17.5 hours with a carbon zinc battery (Standard 9V, NEDA 1604 / IEC 6F22).
Configuration Options
The CL3515R series is configured as a handheld unit accepting standard miniature thermocouple connectors with flat blades spaced 7.9mm center to center.
- Housing Dimensions: 192mm(H) x 91mm(W) x 52.5mm(D)
- Weight: 320g
- Display Configuration: Three display areas (Main, Second, Third). Main and Second are 4½ digit LCDs with a max reading of 19999. The third area displays T1-T2 differential and group settings.
- Data Interface: RS-232 optical interface for PC connection software
Included Accessories
- Two type 'K' thermocouple bead wires
- Two type 'K' calibration bead wires (max insulation temp 260°C/500°F, accuracy ±2.2°C or ±0.75% of reading from 0°C to 800°C)
- 9V battery
- Instruction manual and PC interface cable
Key Product Differences
The series is identified by the model pattern CL3515R followed by the specific thermocouple type (K/J/T/E/R/S/N/L/U/B/C). The instrument features user-selectable temperature scales in Celsius or Fahrenheit.
- Resolution: 0.1° for K, J, T, E, N, L, U types; 1° for R, S, B, C types.
- mV Output Accuracy: ±(0.025% + 1 digit) over the -25.00mV to 75.00mV range.
PDFs & Manuals
Accessories
accessories
8X2X3.5" SOFT CASE
/ each
0 In Stock
cables
Retractable Sensor Cables for Thermocouples, RTDs and Thermistors
SGD72.63 / each
2 In Stock
Retractable Sensor Cables for Thermocouples, RTDs and Thermistors
SGD72.63 / each
10 In Stock
Retractable Sensor Cables for Thermocouples, RTDs and Thermistors
SGD62.79 / each
2 In Stock
Retractable Sensor Cables for Thermocouples, RTDs and Thermistors
SGD65.35 / each
0 In Stock
sensors
General Purpose Thermocouple Probes With Miniature Plug Connector
SGD117.42 / each