- 4 to 20 mA, 0 to 1 Vdc, 0 to 5 Vdc or 0 to 10 Vdc Outputs for Both Temperature and Relative Humidity
- Rugged Industrial Splashproof Enclosure NEMA 13 (IP54)
- Accurate to ± 0.6°C (0 to 50°C) and ± 2.5% (10 to 90% Relative Humidity)
- New Models with Built-In Digital Display to Indicate Temperature and Relative Humidity
HX93-Srs-Temp-RH-Sig-Cond
From
SGD
888.59
Models In Stock
Product Overview
- Cable Diameter 8 mm
- Electrical Connection Liquid-tight nylon with neoprene gland, for 2.50 to 8.00 mm (0.09 to 0.315") diameter cable; internal 6 position terminal block accepts 14 to 22 gauge wire
- Enclosure Ratings IP54, NEMA Type 13
- Length 80.01 mm
- Materials, Wetted Polycarbonate
- Measurement Type Temperature, Humidity
- Relative Humidity Response Time 8 sec
- Repeatability ± 0.1% RH
- Response Time 5 to 30 sec
- Sampling Rate 0.25 Hz (4 seconds between updates)
- Series ID HX93-Srs-Temp-RH-Sig-Cond
- Supply Power Type DC
- Temperature Measurement, Max 75 °C
- Temperature Measurement, Min -30 °C
- Width 82.3 mm
- Wire Gauge 22 to 14 AWG
Supported Applications
The HX93B and HX93BD Series are temperature and relative humidity transmitters that provide remote and on-site monitoring, outputting a linearized current or voltage signal proportional to both measured parameters. Relative humidity output is temperature compensated, and the HX93BD models add a built-in digital display so a reading can be taken directly at the sensor.
Offered in wall-mount, duct-mount, and remote-probe configurations, the series fits HVAC and building automation, clean-room and laboratory environments, storage and process rooms, and OEM equipment where humidity and temperature must be transmitted to a controller or datalogger. A gray polycarbonate housing, stainless steel mesh-type filter, and M12 connector make the units durable and easy to service; the temperature/relative humidity probe is field replaceable (except on the duct-mount model).
The transmitters are CE and UKCA marked with optional NIST calibration. They are not intended for condensing humidity conditions. Duct-mount and remote-probe models include a duct mount kit.
Operating Conditions & Performance
The series measures relative humidity from 0 to 100% and temperature over a standard range of -30 to 75°C (-22 to 167°F) (a special -20 to 75°C (-4 to 167°F) range is also offered). Accuracy is specified by band:
- Temperature: ±0.6°C from 0 to 50°C (32 to 122°F); ±1.25°C from -30 to 0°C (-22 to 32°F) and 50 to 75°C (122 to 167°F)
- Relative humidity: ±2.5% from 20 to 80% RH; ±3.5% from 5 to 20% and 80 to 95% RH; ±4% from 0 to 5% and 95 to 100% RH
- Repeatability: ±0.1% RH, ±0.2°C (0.4°F)
- Resolution: 0.1% RH, 0.1°C (0.1°F)
Temperature response time is 5 seconds minimum to 30 seconds maximum, and relative humidity response time is 8 seconds, with the sensor sampling once every 4 seconds. Input voltage is 30 Vdc maximum, with the HX93B series running on 9 to 30 Vdc and the HX93BV2 series on 12 to 30 Vdc. Cable entry is a liquid-tight nylon gland for 2.50 to 8.00 mm (0.09 to 0.315") diameter cable, and the internal 6-position terminal block accepts 14 to 22 gauge wire.
The probe measures Ø16 mm (Ø0.63 in) with a 71.88 mm (2.83 in) stem, and the housing (without probe) is 80.01 L x 82.30 W x 55.63 D mm (3.15 x 3.24 x 2.19 in). The enclosure is rated NEMA 13 (IP54). HX93BD models feature two 4-digit backlit LCDs updating at one reading per second.
Configuration Options
The lineup is organized along three independent choices, which combine to define each model:
- Display: HX93B (transmitter only) or HX93BD (transmitter/indicator with built-in digital display)
- Output: specify 'C' for 4 to 20 mA, 'V0' for 0 to 1 Vdc, 'V1' for 0 to 5 Vdc, or 'V2' for 0 to 10 Vdc — each spanning 0 to 100% RH and -30 to 75°C (-22 to 167°F)
- Mounting: wall mount (base), duct mount (-D), or remote probe (-RP1)
The output choice sets the electrical interface only; all outputs cover the same full RH and temperature spans. Voltage-output V2 (0 to 10 Vdc) units require the higher 12 to 30 Vdc supply. Duct-mount and remote-probe models ship with a duct mount kit, and the remote-probe (-RP1) versions are the heaviest at 352 g (12.4 oz) versus 251 g (8.9 oz) for the base wall-mount transmitter.
Key Product Differences
Each model number is built as HX93B (or HX93BD for display) plus an output code and a mounting suffix. The base ordering configurations are:
- HX93B(*) — Wall mount temperature and relative humidity transmitter
- HX93B(*)-D — Duct mount temperature and relative humidity transmitter
- HX93B(*)-RP1 — Remote probe temperature and relative humidity transmitter
- HX93BD(*) — Wall mount, digital display transmitter/indicator
- HX93BD(*)-D — Duct mount, digital display transmitter/indicator
- HX93BD(*)-RP1 — Remote probe, digital display transmitter/indicator
Applying the output codes yields the full range of models, including current-output units such as HX93BC, HX93BC-D and HX93BDC, and voltage-output units such as HX93BV0, HX93BV1, HX93BV2 and their -D and -RP1 variants (for example HX93BDV2-RP1).
Accessories include the TH-RP field-replaceable remote probe with 3 m (10') cable and M12 connector, the TH-SP short probe, the HX92-CAL calibration kit with 11 and 75% relative humidity standards, regulated power supplies (PSR-24S, PSR-24L, PSR-24L-230), the TX4-100 shielded transmitter cable, CAL-3-HU NIST-traceable calibration, and the IND-KIT spare duct mount kit.