Product Overview
  • Accuracy ± 2 %
  • Accuracy Detail ± 2% FS, 0 to 9000 FPM (45 M/s)
  • Compliance Note Meets the technical requirements of EU Directive 2011/65/EU (RoHS II).
  • K-Factor 0.81
  • Materials, Wetted 304 Stainless Steel
  • Operating Temperature, Max 1500 °F
  • Port 1 Size 1/4 in
  • Port 1 Type Stem
  • Port 2 Size 1/4 in
  • Port 2 Type Stem
  • Series ID Series-160F-Kit
  • Weight 14 lb

Supported Applications

The Dwyer Series 160F Pitot Tubes are designed for the environmental testing field to measure air or gas stream flow velocity. The straight stainless steel design facilitates insertion into ducts through grills and pressure taps, aiding positioning in hard-to-reach locations where hook-style pitot tubes may not allow access.

Operating Conditions & Performance

The series measures air velocity or airflow within a range of 0 to 9000 FPM (45 M/s) with an accuracy of ±2% FS. The wetted material is 304 SS, and the process temperature limit is 1500°F (815°C). For optimal performance, readings should be taken in a smooth, straight duct section a minimum of 10-15 duct diameters upstream and 5 diameters downstream from the pitot tube. The K-Factor for velocity calculation is 0.81.

Configuration Options

The Series 160F offers multiple probe lengths to accommodate various installation depths, with permanently etched insertion depth graduations on the sensing tube. Available configurations include:

  • Probe Lengths: Models are available in 18" (457.2 mm), 24", 36", 48", and 60" lengths.
  • Dimensions: The probe diameter is 1/4" [6.35] with a process connection of 1/4" OD. Overall length varies by model, ranging from 19-9/32 (489.56) for the base unit to 61-9/32 (1556.54) for the longest variant.
  • Accessories: Optional accessories include a flange mounting plate with 1/2" female NPT, split flanges, mounting glands, step drills, and carrying cases for pitot tubes up to 48".

The Series 160F-KIT includes the 160-18, 160-24, 160-36, and 160-48 models along with a carrying case. The static pressure port is parallel to the sensing tube to allow quick alignment of the tube with airflow.

Key Product Differences

Variants within the series are distinguished primarily by probe length and overall dimension "A":

  • 160F: 18" (457.2 mm) insertion length; 19-9/32 (489.56) overall.
  • 160F-24: 24" (609.6 mm) insertion length; 25-9/32 (642.14) overall.
  • 160F-36: 36" (914.4 mm) insertion length; 37-9/32 (946.94) overall.
  • 160F-48: 48" (1219.2 mm) insertion length; 49-9/32 (1251.74) overall.
  • 160F-60: 60" (1524 mm) insertion length; 61-9/32 (1556.54) overall.

All models meet the technical requirements of EU Directive 2011/65/EU (RoHS II). The weight for individual units is listed as 4.3 oz (122 g).