TRCIII-Series

High Precision Calibration-Temperature Reference Chamber

  • A 0°C (32°F) Thermoelectric “Refrigerator”
  • Eliminates Old Fashioned “Ice Bath”
  • For Versatile Use in the Factory, Laboratory, Instrument Shop
  • Calibration of All Temperature Instruments and Sensors
TRCIII-Series
From SGD 9,291.15
High Precision Calibration-Temperature Reference Chamber TRCIII-Series TRCIII-Series
loading all models
Product Overview
  • Accuracy ± 0.1 °C (± 0.18 °F)
  • Display Resolution None
  • Form factor Portable Benchtop
  • Frequency 60 Hz
  • Height 305 mm
  • Length 322 mm
  • Number of Reference Wells 6
  • Number of Reference Wells Note 6 wells, 5⁄32 " ID, 33⁄4 " deep; wells are thermally and electrically grounded to each other; accepts 1⁄8 " OD probes
  • Operating Temperature, Max 32 °C
  • Operating Temperature, Min 2 °C
  • Reference Temperature 0 °C
  • Reference Wells Depth 3.75 in
  • Reference Wells Inner Diameter 0.1563 in
  • Series ID TRCIII-Series
  • Stability ± 0.04°C
  • Stabilization Time 2 or 3 hours depending on ambient temperature and thermal load
  • Supply Power Note 115/220 V, 60 Hz, 100 Watts
  • Weight 9.5 kg
  • Width 159 mm

Supported Applications

The Omega TRCIII Series is a high precision ice point™ Calibration Reference Chamber designed to maintain a stable reference temperature of 0°C (32°F) using freezing water. The series serves as an on-site calibration standard for thermocouple, RTD, and thermistor probes.

Operating Conditions & Performance

The TRCIII Series operates within an ambient temperature range of 2 to 32°C (35 to 90°F). The unit maintains a process reference point at 0°C (32°F) with accuracy values stated as ±0.1°C or better. Media compatibility is limited to distilled, deionized water; the unit must not be allowed to freeze completely, as damage will result.

Configuration Options

The series offers configurable power supply options and optional accessories for expanded calibration workflows:

  • Voltage Models: TRCIII (120 Vac operation) and TRCIII-220 (220 Vac operation).
  • Power Requirements: 115/220 V, 60 Hz, 100 Watts.
  • Physical Dimensions: 305 H x 159 W x 322 mm D (12.0 x 6.25 x 12.7").
  • Weight: 9.5 kg (21 lb).

The series supports the following optional configurations and accessories:

  • NIST Traceable Calibration Available.
  • Companion Dry Block Probe Calibrator (Model CL900/950).
  • TRP reference probes including TRP-K, TRP-J, TRP-T, TRP-E, TRP-C, TRP-D, TRP-B, TRP-S, and TRP-R.
  • OMEGACARE™ Extended Warranty Program (extends standard 1-year warranty to a total of 4 years).

The unit is supplied with male and female color-coded thermocouple connectors. White-coded Copper-Copper connectors are specified for electrical connection.

Key Product Differences

Variants within the TRCIII Series differ primarily by input voltage:

  • TRCIII: ice point™ reference 120 Vac operation.
  • TRCIII-220: ice point™ reference 220 Vac operation.

The CAL-3 option provides NIST-traceable calibration. TRP reference probes are available in standard lengths of 12" for Types J, K, T, and E, or 6" for Types R, S, B, C, G, and D; longer leads (e.g., -24, -36) are configurable at additional cost.