  • Calibrates from Ambient plus 11°C to 260°C (Ambient plus 20°F to 500°F )
  • Rugged Miniature Handheld, Benchtop and Portable Design
  • Fast, Accurate, Stable Readings 
  • Fast Heat-Up/ Cool-Down Time
CL1000
From SGD 5,652.30
Ambient 11 °C to 260 °C Temperature Dry Block Calibrator CL1000 CL1000
loading all models
Product Overview
  • Accuracy ± 1.5 °C (± 2.7 °F)
  • Additional Features 1 Compact & Portable
  • Cool-Down Time Note 260 to 38°C (500 to 100°F) in 70 minutes
  • Display Resolution 0.1°
  • Frequency 50/60 Hz
  • Heat-Up Time Note 38 to 260°C (100 to 500°F) in 15 minutes
  • Operating Temperature, Max 50 °C
  • Operating Temperature, Min 0 °C
  • Power Consumption 275 W
  • Process Temperature, Max 260 °C
  • Process Temperature, Min 11 °C
  • Process Temperature Note Temperatur Uniformity: [within 2.5 cm (1") from Bottom of Well]: ±0.15°C (±0.3°F)
  • Relative Humidity, Max 0.8
  • Relative Humidity, Min 0%
  • Relative Humidity Note 0 to 80% RH non-condensing
  • Sensor Type RTD
  • Sensor Type Note Platinum RTD, 100 Ω, 0.00385, Class A
  • Series ID CL1000
  • Temperature Stability ±0.15°C
  • Weight 1.7 kg

Supported Applications

The Omega CL1000 Series Mini hot point® Dry Block Probe Calibrator is a portable, rugged benchtop device designed to test and calibrate temperature probes. The unit features a built-in precision PID digital controller and provides a stable thermal environment for calibration tasks in both laboratory and field settings.

Operating Conditions & Performance

The CL1000 series operates within an ambient temperature range of 0 to 50°C (32 to 122°F) with humidity ratings up to 90% non-condensing. The probe well can be set to any temperature between Ambient +11 to 260°C (Ambient +20 to 500°F). Performance specifications include an accuracy of ±1.5°C (±2.75°F) and a resolution of 0.1°. Each unit includes a NIST Calibration Certificate with 2 data points.

Power requirements are available in two configurations: 3.15 A @ 115 Vac ~ (±10%) 50/60 Hz, 275 W for domestic models and 1.6 A @ 230 Vac ~ (±10%) 50/60 Hz, 275 W for European models.

Configuration Options

The series offers configurable well styles to accommodate various probe diameters in both standard and metric sizes. The enclosure is a custom fabricated metal design with overall dimensions of 127 W x 56 H x 155 mm D (5 x 2.2 x 6.1") and weighs 1.7 kg (3.8 lb).

Well Styles

  • CL1000A: Standard well style with holes for 1/16", 1/8" (2), 5/32", and 3/16" probes.
  • CL1000B: Standard well style with holes for 1/16" (2), 1/8" (2), 3/16", and 1/4" (2) probes.
  • CL1000C: Standard well style with holes for 1/16" (2), 1/8" (2), and 5/32" (2) probes.
  • CL1000D: Metric well style with holes for 1.5mm, 2.0mm, 3.0mm, and 4.5mm probes.

Communications Options

An optional RS-232 communication port is available via the -C2 suffix. This option includes pre-wired connectivity with free CN9-SW communications software and a 1.8 m (6') cable included.

Included Accessories

  • Soft Carrying Case
  • Power Cord
  • Operator's Manual
  • NIST Calibration Certificate with 2 Data Points Included

The series also supports the OMEGACARE™ Extended Warranty Program. Custom well styles are available through Omega's Custom Engineering Department.

Key Product Differences

Variants within the CL1000 series differ primarily by voltage input and communication capabilities:

  • Voltage Models: Base models operate on 115 Vac. Adding the suffix "-230V" designates a model for 230 Vac operation.
  • Communication Suffixes: The "-C2" suffix indicates inclusion of RS232 communications and software; this can be combined with voltage options (e.g., CL1000A-230V-C2).

The series is intended for indoor use only. It is not designed for medical or nuclear applications, nor should it be operated in flammable or explosive environments.