8-Channel USB Thermocouple Data Acquisition Module

  • Measures from -270 to 1820°C (-454 to 3308°F)
  • Accepts Thermocouple Types J, K, T, E, R, S, B and N
  • Automatic Cold Junction Compensation
  • High Accuracy and Resolution
TC-08-TC-DAQ
From SGD 1,131.06
8-Channel USB Thermocouple Data Acquisition Module TC-08-TC-DAQ TC-08-TC-DAQ
loading all models
Product Overview
  • Accuracy ± 0.2 %
  • Analog I/O Note Analog Input Note: Conversion Time: 100 ms (thermocouple and cold junction compensation)
  • Common Mode Voltage ±7.5V
  • Configuration Pentium class processor or equivalent
  • Conversion Rate 100 ms
  • Data Interface USB 1.1, USB 2.0
  • Data Interface Note Supported Operating System: Microsoft Windows®XP SP2 / VISTA/7
  • Electrical Connection Miniature Female Thermocouple Connector
  • Height 34 mm
  • Impedance Input 2 MΩ
  • Input Type Thermocouple, Voltage
  • Input Type Note Thermocouple: J, K, T, E, R, S, B, N
  • Length 201 mm
  • Memory 32 MB minimum
  • Number of Channels, Input Analog: 8
  • Series ID TC-08-TC-DAQ
  • Software TC-08 Recorder and Player
  • USB Cable Length 1.8 m
  • Voltage Measurement Range ±70 mV
  • Width 104 mm

Supported Applications

The Omega TC-08 is an eight-channel USB thermocouple data acquisition module designed to measure temperatures using miniature connector thermocouples. It supports Types B, E, J, K, N, R, S, and T for temperature measurement applications requiring connection to a Windows-based PC (XP SP2 or above, Vista, Win 7). The device is also capable of measuring other sensors utilizing the ±70 mV input range.

Operating Conditions & Performance

The TC-08 provides temperature measurement ranges dependent on the specific thermocouple type connected. Overall capabilities span from -270 to 1820°C (-454 to 3308°F). Specific documented ranges include Thermocouple Type B: 20 to 1820°C (68 to 3308°F); Type E: -270 to 910°C (-454 to 1670°F); Type J: -210 to 1200°C (-346 to 192°F); Type K: -270 to 1370°C (-454 to 2498°F); Type N: -270 to 1300°C (-454 to 2372°F); Types R and S: -50 to 1760°C (-58 to 3200°F); and Type T: -270 to 400°C (-454 to 752°F).

Performance specifications include a temperature accuracy of the sum of ±0.2% of reading and ±0.5°C, with voltage accuracy at the sum of ±0.2% of reading and ±10 µV. The module features automatic cold junction compensation with a conversion time of 100 ms (including thermocouple and cold junction compensation). Resolution is specified as 20-bits, with a noise-free resolution of 16.25-bits.

Electrical Specifications

  • Input Range: ±70 mV (Voltage)
  • Overload Protection: ±30V
  • Maximum Common Mode Voltage: ±7.5V
  • Input Impedance: 2 MΩ
  • Power Supply: From USB port (USB bus powered)

PC Requirements

  • Processor: Pentium class processor or equivalent
  • Memory: 32 MB minimum
  • Disk Space: 10 MB minimum
  • Ports: USB 1.1 or USB 2.0 compliant port (must be connected direct to the port or a powered USB hub)

Configuration Options

The TC-08 series is configured as an eight-channel module with miniature thermocouple input connectors and a supplied USB cable measuring 1.8 m (6'). The unit connects via a USB 1.1 interface to the host PC.

Software & Data Output

The device includes TC-08 Recorder and Player software for data visualization, allowing output in spreadsheet or graph format. A single high-specification PC can support up to 64 devices via the driver, while the included software supports up to 20 USB TC-08 units simultaneously.

Physical Dimensions

The module measures 201 x 104 x 34 mm (7.91 x 4.09 x 1.34"). It is CE marked and carries a 5 YEAR WARRANTY.

Key Product Differences

The series consists of the TC-08 model, described as an "8-channel thermocouple data acquisition module." The resolution capability varies by temperature range; for example, Type K offers 0.1°C resolution from -270 to 1370°C and 0.025°C resolution from -250 to 1370°C.

Users must install the software before connecting the USB TC-08 device. The product is not designed for use in, or should not be used for, human applications.