Need assistance? Call us: 1-800-663-4209
CDCN91-Controller
Dual Channel Conductivity/ResistivityController
- 2-Channel Input
- Simultaneous Display
- AC Line-Voltage or DC Powered
- Display and/or Control: µS, mS, PPM or PPB (TDS), kohms, Mohms, % Rejection, Difference, Ratio, °C or °F
CDCN91-Controller
From
C$
2,516.36
Product Overview
- Accuracy ± 2 %
- Additional Features 1 Dual Channel; NEMA 4X (IP65) front and back with optional NEMA 4X (IP65) rear cover kit
- Conductivity Measurement 0.055 to 400,000 µS/cm
- Display Type 2 x 16 Alphanumeric LCD
- Display Type Note Alphanumeric 2 x 16 LCD Contrast: User selected, 5 levels Update rate: 1.5 seconds
- Enclosure Ratings IP65, NEMA Type 4X
- Frequency 50/60 Hz
- Materials PBT, Polycarbonate
- Number of Relays None
- Operating Range Note Resistivity: 10 kΩ/cm to 18.26 MΩ/cm (0.055 to 100 μS/cm); TDS: 0.001 to 999999 ppm or ppb (display limit)
- Operating Temperature, Max 120 °C
- Operating Temperature, Min -25 °C
- Output Detail Current Outputs: (3 each) 4 to 20 mA, isolated, fully adjustable and reversible Max Loop Impedance: 150 Ω @ 12V, 450 Ω @ 18V, 750 Ω @ 24V Update Rate: Approx. 100 mS Accuracy: ±0.03 mA @ 25°C, 24 Vdc Open Collector Outputs: (2 each) Isolated, 50 mA sink or source, 30 Vdc max pull-up voltage Operational Settings: Hi, Lo, USP, Pulse, Off Hysteresis: User adjustable Time Delay: 0 to 6400 seconds Maximum Pulse Rate: 400 pulses/minute
- Power Consumption 20 VA
- Relay Note (up to 4 each) SPDT Relays Maximum Voltage Ratings: 5A @ 30 Vdc or 5A @ 250 Vac Operational Settings: Hi, Lo, USP, Pulse, Off Hysteresis: User adjustable Time Delay: 0 to 6400 seconds Maximum Pulse Rate: 400 pulses/minute
- Resistivity Measurement Range 10 kΩ/cm to 18.26 MΩ/cm
- Series ID CDCN91-Controller
- Temperature Accuracy ± 0.5 °C
- Weight 581 g
Supported Applications
The Omega CDCN-91 Series is a dual-channel conductivity/resistivity controller designed for monitoring and controlling process fluids. The series supports applications requiring precise measurement of conductivity, resistivity, Total Dissolved Solids (TDS), and temperature. It features programmable setpoints with adjustable hysteresis to manage alarm and control functions.
Documented application contexts include:
- Water treatment processes
- General process fluid monitoring and control
- Simultaneous display of two independent channels for comparative analysis (e.g., Percent Reject, Difference, or Ratio)
Operating Conditions & Performance
The CDCN-91 Series operates within specific environmental limits to ensure measurement integrity. The ambient operating temperature range is -10 to 55 °C (14 to 131 °F), with a storage temperature of -15 to 80 °C (5 to 176 °F). Humidity tolerance is rated at 0 to 95%, non-condensing.
Performance specifications for the series include:
- Conductivity Range: 0.055 µS/cm to 400,000 µS/cm (displayed as 0.055 µS/cm to 400.00 mS/cm)
- Resistivity Range: 10 kΩ/cm to 18.26 MΩ/cm
- TDS Range: 0.001 to 99,999 ppm or ppb (display limit)
- Temperature Measurement: -25 to 120 °C (-13 to 248 °F) using PT1000 sensors
- Accuracy: ±2% of reading for Conductivity/Resistivity; ±0.5°C for Temperature (0 to 100 °C); ±0.03 mA @ 25°C, 24 VDC for current outputs
- Update Rate: Approximately 1.5 seconds per update cycle
The enclosure is rated NEMA 4X (IP65) on the front and back when equipped with the optional rear cover kit.
Configuration Options
The CDCN-91 Series offers configurable power, output signaling, and mounting options to accommodate diverse installation requirements. The series supports both AC and DC power configurations via distinct model numbers (CDCN-91 for DC; CDCN-91AC for AC/DC).
Power Supply Options
- CDCN-91: 11 to 24 VDC ±10%, regulated, 0.5 A max
- CDCN-91AC: 100 to 240 VAC ±10%, 50-60 Hz, 20 VA (also supports DC input)
Output Signals & Control
- Analog Outputs: Up to three fully scalable and isolated 4 to 20 mA outputs. Max loop impedance is 150 Ω @ 12 V, 450 Ω @ 18 V, or 750 Ω @ 24 V.
- Digital Outputs: Up to four programmable SPDT relays (5 A @ 250 VAC, 5 A @ 30 VDC).
- Open Collector: Two optically isolated outputs available; 50 mA sink or source with a maximum pull-up voltage of 30 VDC.
Sensor & Connection Options
- Sensors: Compatible with Omega CDCE-90 Series Standard and Certified Conductivity/Resistivity Electrodes (cell constants from 0.01 to 20.0).
- Cabling: Three-conductor shielded cable for sensor extensions up to 30 m (100 ft). Liquid tight connector kits (FPM-5000-LTCK) are available.
Housing & Mounting
- Enclosure: PBT case with polyurethane-coated polycarbonate window and sealed 4-key silicone rubber keypad. Includes optional NEMA 4X (IP65) rear cover kit.
- Mounting: Standard panel mounting; optional heavy-duty wall mount bracket (CDCN-91-WMB) or surface mounting bracket (FPM-5000-MB).
Key Product Differences
The primary distinction within the CDCN-91 Series is the power supply configuration, which defines the model number:
- CDCN-91: Designed for DC power applications (11 to 24 VDC).
- CDCN-91AC: Supports both AC and DC power inputs.
The series utilizes a dual-channel architecture allowing simultaneous monitoring of Channel 1 and Channel 2. Display modes vary based on channel configuration, showing either two conductivity values (C1/C2), one conductivity value with its corresponding temperature, or calculated ratios/differences between channels. The device includes output simulation capabilities for system testing.
PDFs & Manuals
Accessories
consumables
installation hardware
TRANSMITTER COVER NEMA 4X
C$126.58 / each
0 In Stock
LIQUID TIGHT CONNECTOR (3) KIT
C$149.35 / each
0 In Stock
MOUNTING BRACKET
C$102.47 / each
0 In Stock
sensors
Conductivity Cells
C$1,288.96 / each
0 In Stock
Show Ratings & Reviews
how do you reset the passcode to access the calibration?
Thank you for your inquiry.
Press and hold ENTER key:
• 2 seconds to select the CALIBRATE menu.
• 5 seconds to select the OPTIONS menu
The Key Code is UP-UP-UP-DOWN keys in sequence.
• After entering the Key Code, the display will show the fi rst item in the selected menu.
Date published: 2023-08-02
which conductivity cells are included by default with unit?
Thanks for your question. It does not come include the conductivity cell. You need to buy the probe separably. Please see CDCE series.
Date published: 2019-08-09