Skip to search Skip to main content Skip to navigation

United Kingdom 0800 488 488 | Intl +44 (0) 161 777 6611 | Welcome to our new website, found an issue or bug? Please report it here

We noticed you are in , please visit your local site by clicking the below link for regional offers and support.

OM-CP-RFRTDTEMP2000A-Series

Wireless RTD Temperature Data Logger with Display

  • Wireless Two Way Communication
  • 100Ω Pt RTD Input
  • View Data in Real Time
  • 5 Year Battery Life
OM-CP-RFRTDTEMP2000A-Series
From £ 362.90
Wireless RTD Temperature Data Logger with Display OM-CP-RFRTDTEMP2000A-Series OM-CP-RFRTDTEMP2000A-Series
loading all models
Product Overview
  • Alarm Note Programmable high and low limits; alarm is activated when temperature reaches or exceeds set limits, Alarm Delay: A cumulative alarm delay may be set in which the device will activate the alarm (via LED) only when the device has recorded a user specified time duration of dataAudible Alarm Functionality: 1 beep per second for reading alarm above/below threshold
  • Band Frequency 2.405 to 2.475 GHz
  • Band Frequency Note ISM band 2.405 to 2.475 GHz operation (channels 11 through 25) @ 250 kbps data rate
  • Battery life 4 yr, typical with reading interval of 1 min
  • Battery Size and Type Li-ion, 9V
  • Baud Rate 250000
  • Calibration Digital Calibration through Software
  • Calibration Accuracy Temperature: ± 0.05 °C
  • Calibration Date Automated
  • Configuration Omega Data Logging Software, Windows PC
  • Data Format °C, °F, K, Ω, °R
  • Data Interface Wireless, USB
  • Data Interface Note USB to mini USB, 250,000 baud for standalone operation or OM-CP-RFC1000-EXT/ OM-CP-RFC1000-EXT-EUR/OM-CP-RFC1000-IP69K required for wireless operation
  • Display LCD
  • Form factor Wall Mount
  • Height 0.95 in
  • Input Type Internal
  • LED Functionality Green LED blinks every 5 seconds to indicate unit is logging, Blue LED blinks every 15 seconds to indicate unit is in wireless mode, Red LED blinks every 1 second to indicate alarm condition
  • Length 3.25 in
  • Materials Plastic
  • Measurement Type Temperature
  • Memory Modes Yes
  • Number of Measurements 1
  • Operating Temperature, Max 60 °C
  • Operating Temperature, Min -20 °C
  • Output Power, Max +0dBm
  • Pressure Resolution 0.02 mbar
  • Pulse Resolution ≤10 µS
  • Relative Humidity 0 to 95% RH non-condensing
  • Response Time Temperature: 10 min
  • Response Time Note Temperature: 10 min free air
  • RF Frequency 2.45 GHz
  • RF Frequency Note 2.45GHz IEEE 802.15.4 ultra-low power wireless transceiver with fully bi-directional communication
  • RH Resolution 0.001
  • RTD Type 2, 3 or 4-wire 100Ω Pt RTD, α= 0.00385
  • Samples,Max 21504 Samples, Interval Selectable 1 Sec to 24 hrs
  • Series ID OM-CP-RFRTDTEMP2000A-Series
  • Software XP SP3/Vista/7 and 8 (32 and 64-bit)
  • Start Modes Delayed Start, Immediate Start
  • Supply Power Note Ultralife 9V lithium battery included; user replaceable
  • Supply Power Type Battery
  • Temperature Measurement, Max 850 °C
  • Temperature Measurement, Min -200 °C
  • Time Accuracy ± 1 minute/month
  • Weight 4.1 oz
  • Width 3.5 in

Supported Applications

The Omega OM-CP-RFRTDTEMP2000A is an RTD-based, wireless temperature data logger with display. The series identity centers on high accuracy temperature monitoring within laboratories and the pharmaceutical and biotech industries.

Operating Conditions & Performance

This device accepts 2, 3 or 4 wire 100Ω platinum RTD sensors. It features a documented device accuracy of ±0.05°C. Recalibration is recommended annually at 25°C and 60°C.

Configuration Options

The series supports wireless communication via the OM-CP-RFC1000-EXT interface device or plugged-in operation using a USB cable. Power options include battery replacement with the OM-CP-BAT103 battery.

  • Interface: Wireless (via OM-CP-RFC1000-EXT) or USB cable
  • Sensor Input: 2, 3, or 4 wire 100Ω platinum RTD
  • Battery Option: OM-CP-BAT103 battery for replacement

Key Product Differences

The series is identified by the model number guide OM-CP-RFRTDTEMP2000A. The device carries a 5-year warranty and is made in the USA. Corporate quality certifications include CE, ISO 9001 Certified (Stamford, CT), and ISO 9001 Certified (Manchester, UK).

en-GB
en_UK
Save Configuration