Product Specs
<< See all Product Models| Brand |
|
| Accuracy | ± 0.08 % |
| Accuratezza | ± 0.08 % |
| Accuratezza Dettagli | 0.08% BSL linearity, hysteresis and repeatability combined Thermal Accuracy: ± 0.50%(Span &Zero) |
| Cicli di pressione | 1 million, minimum |
| Classe di accuratezza | < 0.1% |
| Collegamento elettrico | Cable |
| Compatibilità con i fluidi | Any, Clean dry air / inert gas, Clean process / cooling water, Corrosive chemicals / acids / bases, Fuel gases, HVAC / room air, Hydraulic fluids, Inert / non-reactive gases, Liquids, Oils / lubricants, Organic solvents, Steam, Nitrogen, Non-corrosive Dry Gasses, Oil, Oxygen, Solvents, Water, Water-based fluids |
| Dimensioni attacco al processo | 1/4 in |
| Electrical Connection | Cable |
| Electrical Output | 4 to 20 mA |
| Errore di azzeramento/bilanciamento | ± 0.5% |
| Errore zero/equilibrio Nota | ± 0.5% typical (1% maximum) |
| Funzionalità aggiuntive 1 | None |
| ID serie | PX409-DIF |
| Larghezza di banda | DC to 1 kHz |
| Liquidi o gas compatibili con il lato negativo | Clean, Dry, Non-ionic gases |
| Lunghezza cavo | 2 m |
| Materiali a contatto con il fluido | 316L Stainless Steel |
| Materiali a contatto con il fluido lato positivo | 316L stainless steel |
| Measurement Pressure, Max | 5 psi |
| Measurement Pressure, Min | 0 psi |
| Negative Side differentional compatable fluids or gasses | Clean, Dry, Non-ionic gases |
| Nota sulla stabilità | Long Term Stability (1-Year): ± 0.1% FS typical |
| Peso | 200 g |
| Potenza elettrica | 4 to 20 mA |
| Pressione di linea/statica | 500 psi max |
| Pressione di prova | 4x Full Scale up to 1000 psi |
| Pressione di prova Nota | 4 times range |
| Pressione, max | 5 psi |
| Pressione, min | 0 psi |
| Process Connection Type | NPT Male |
| Range di regolazione dell'ampiezza | ± 0.5% typical (1% maximum) |
| Resistenza agli urti | 50 g, 11 ms half-sine |
| Resistenza del circuito, max | (Vs-9) x 50 Ω |
| Resistenza dell'isolamento | 100 MΩ |
| Stabilità | ± 0.1% |
| Temperatura compensata, max | 85 °C |
| Temperatura compensata, min. | -17 °C |
| Temperatura di esercizio minima | -45 °C |
| Temperatura di esercizio, max | 115 °C |
| Tempo di risposta | <1 ms |
| Tensione di alimentazione | 9 to 30 Vdc |
| Tipo di attacco al processo | NPT Male |
| Tipo di pressione | Differential |
| Uscita segnale | Current |
| Valore peso lordo | 0.15 |
| Vibrazioni | 5-2000-5 Hz, 30 minute cycle,Curve L, Mil-Spec 810 figure 514-2-2, vertical and horizontal axis |
| Country of Origin | United States |
| See All Specs | |
PDFs & Manuals
Accessories
Static Pressure Sensors for Manometers & Gages across Air Filters
From SGD10.49