Need assistance? Call us: 1-800-663-4209
OM-CP-XTCTemp2000-Srs
X-Series - Multi-channel Thermocouple Datalogger w/ LCD
Models In Stock
- Large Backlit LCD Display
- User-Friendly Front Panel
- Monitor Multiple Points
- Accepts thermocouple types J, K, T, E, R, S, B, N
OM-CP-XTCTemp2000-Srs
From
$
1,006.51
Models In Stock
Product Overview
- Calibration Digital calibration through software
- Data Interface WiFi
- Data Interface Note USB-A to micro USB cable (included); 460,800 baud
- Electrical Connection Screw Terminal
- Enclosure Aluminium
- Form factor Benchtop
- LED Indicators 1 per channel and 2 status LEDs
- Length 184.15 mm
- Operating Temperature, Max 60 °C
- Operating Temperature, Min -20 °C
- Process Temperature, Max 60 °C
- Process Temperature, Min -20 °C
- Real Time Recording May be used with PC to monitor and record data in real time
- Samples,Max 2000000
- Sampling Rate 1 day
- Sampling Rate Note 4 Hz up to 1 reading every 24 hours
- Series ID OM-CP-XTCTemp2000-Srs
- Software Compatibility Windows® XP/VISTA/7 and 8 (32- and 64-bit)
- Start Modes Adjustable Delay Start, Immediate Start
- Supply Power Type Battery
- Temperature Accuracy ± 0.5 °C
- Temperature Accuracy Note ± 0.5 °C (0 °C to 50 °C) ± 0.9 °F (32 °F to 131 °F)
- Temperature Resolution 0.01 °C
- Temperature Sensor Type Thermocouple
- Thermocouple Resistance, Max 1000 Ω
- Time Accuracy ± 1 minute/month
- Width 65.58 mm
- Wire Gauge 24 AWG
Supported Applications
The OM-CP-QUADTCTEMP2000-A2 and OM-CP-OCTTCTEMP2000-A2 are four and eight channel thermocouple data loggers with an LCD screen. The integrated display, with on-screen minimum, maximum and average statistics, makes the series well suited to applications where readings must be checked at the point of measurement without connecting to a computer.
Because the devices accept J, K, T, E, R, S, B and N type thermocouples across a wide temperature span, they support multi-point temperature monitoring and thermal recording in industrial and laboratory settings. A user configurable screen that allows any combination of channels to be displayed makes them convenient for field checks and walk-up verification.
- Multi-point thermocouple temperature monitoring
- Point-of-measurement readings via the on-board LCD
- Industrial and laboratory thermal recording
- Field verification of live and statistical values
Operating Conditions & Performance
The internal channels measure over a temperature range of -20 °C to +60 °C (-4 °F to +140 °F) with a resolution of 0.01 °C (0.018 °F). Calibrated accuracy is ±0.5 °C (0 °C to 50 °C) and ±0.9 °F (32 °F to 131 °F). The loggers operate in an environment of -20 °C to +60 °C (-4 °F to +140 °F) and 0 %RH to 95 %RH non-condensing.
Remote thermocouple channels use a female subminiature (SMP) connection and accept a maximum thermocouple resistance of 1000 Ω, <100 Ω recommended, with automatic cold junction compensation based on the internal channel. Accuracy is ±0.5 °C for the J, K, T, E and N types and ±2.0 °C for the R, S and B types.
The reading rate ranges from 4 Hz up to 1 reading every 24 hours. Memory capacity with all channels enabled is 524,032 readings per channel on the 4-channel model and 262,016 readings per channel on the 8-channel model. Data is date and time stamped in °C, °F, K, °R, mV, V, and time accuracy is ±1 minute/month.
Each unit uses a 9 V lithium battery (included and user replaceable) rated for 18 months with the display off and 3 months with continuous display use. Connection is through a USB-A to micro USB cable (included) at 460,800 baud, with software programmable immediate start or delay start up to 6 months in advance and digital calibration through software. The series carries CE approval.
Configuration Options
The two members of the series differ in channel count, physical size, and memory depth while sharing the same measurement performance and LCD feature set. The 4-channel OM-CP-QUADTCTEMP2000-A2 measures 7.25 in x 2.70 in x 1.22 in (184.15mm x 65.58 mm x 30.99 mm) and weighs 15.3 oz (435 g), storing 524,032 readings per channel. The 8-channel OM-CP-OCTTCTEMP2000-A2 measures 7.25 in x 2.70 in x 1.68 in (184.15mm x 65.58 mm x 42.672 mm) and weighs 20.1 oz (569 g), storing 262,016 readings per channel.
Both models accept the same thermocouple types. Type J covers -210 to +760 °C, type K covers -270 to +1370 °C, type T covers -270 to +400 °C, type E covers -270 to +980 °C, type N covers -270 to +1300 °C, type R covers -50 to +1760 °C, type S covers -50 to +1760 °C, and type B covers +50 to +1820 °C.
Both variants provide the on-screen minimum, maximum and average statistics, the user configurable screen that displays any combination of channels, automatic cold junction compensation, and real time recording with a PC. The OM-CP-BAT103 replacement battery allows user servicing on either model.
Key Product Differences
- OM-CP-QUADTCTEMP2000-A2 - 4-Channel Thermocouple Data Logger with LCD screen and USB-A to micro USB cable
- OM-CP-OCTTCTEMP2000-A2 - 8-Channel Thermocouple Data Logger with LCD screen and USB-A to micro USB cable
- OM-CP-BAT103 - Replacement 9V lithium battery
PDFs & Manuals
Software & Drivers
Show Ratings & Reviews
Does this meet the requirements for AMS2750G
Thank you for your question. Unfortunately, it does not.
Date published: 2022-12-29
Does this controller have any outputs
Thank you for your question. This not a controller just a Datalogger and it does not have outputs just t/c inputs.
Date published: 2022-08-12
Upon initializing the device the display has locked up and cannot access it. None of the buttons work and software will not reset the device display. I am unable to reset anything with the device and cannot calibrate the display.
Thank you for your inquiry. Please contact our DAS Department at 1-888-826-6342 or at das@omega.com regarding this issue.
Date published: 2025-09-25
Is there a way I can stop/run Datalogger while it is not connected to the computer ?
Thank you for your question. We will require more information to help on this question, please email the DAS team at DAS@dwyeromega.com
Date published: 2024-08-14
I have a channel that says Channel unplugged and Channel connected. so my data goes in and out
Thank you for your question. We will require more information to help on this question, please email the DAS team at DAS@OMEGA.COM
Date published: 2023-04-05
How do i change this data logger from mv to temperature. "om-cp-quadtctemp200-a2" it has ambient temp ,all 4 channels read mv i would like it to read temperature
Thank you for your question. We will require more information to help on this question, please email the DAS team at DAS@OMEGA.COM
Date published: 2023-07-06
I have a Fluke 5502 multi-product calibrator. Is there a calibration procedure available I can use to calibrate this device in-house
Thank you for your question. We will require more information to help on this question, please email the DAS team at DAS@OMEGA.COM.
Date published: 2022-11-23
Can I edit the Channel title to say something more suitable for any application? I am looking to be able to export the information from the Datalogger without having to edit it in the excel file. Is this possible?
Thank you for your question. We will require more information to help on this question, please email the DAS team at DAS@OMEGA.COM
Date published: 2023-03-11