Product Overview
- Additional Features 1 High Accuracy
- Alarm Note High/low alarm thresholds for %RH and temperature selectable in software
- Battery life up to 1 yr
- Battery Size and Type 1/2 AA Li-ion
- Data Interface USB
- Dew Point Accuracy ± 1.1°C
- Display Type LCD
- Form factor Handheld
- Height 25.3 mm
- Humidity Repeatability ± 0.1%
- Input Type Internal
- Measurement Type Humidity, Temperature
- Memory 16379
- Memory Note 16,379 temperature and 16,379 relative humidity readings
- Number of Measurements 2
- Operating Temperature, Max 80 °C
- Operating Temperature, Min -35 °C
- Relative Humidity Accuracy ± 2% RH
- Relative Humidity Accuracy Note Accuracy [20 to 80% RH] (OM-EL-USB-2-LCD-PLUS); ± 2.0% RH typical; ± 4.0% RH max
- Relative Humidity, Max 1
- Relative Humidity, Min 0%
- Relative Humidity Resolution 0.005
- Samples,Max 16379
- Sampling Rate 10 Hz (10 updates/second)
- Software Compatibility Windows XP/Vista/7 and 8 (32-bit and 64-bit)
- Supply Power Type Battery
- Temperature Accuracy ± 0.3 °C (± 0.54 °F)
- Temperature Accuracy Note ± 0.3°C (± 0.6°F) typical; ± 1.5°C (± 3.0°F) max
- Temperature Measurement, Max 80 °C
- Temperature Measurement, Min -35 °C
- Temperature Repeatability ± 0.1 °C
- Temperature Resolution 0.5 °C
- Weight 40 g
- Width 126 mm
The OM-EL-USB-2-LCD is a standalone data logger that measures and stores up to 16,379 relative humidity and 16,379 temperature readings over 0 to 100% RH and -35 to 80°C (-31 to 176°F) measurement ranges. The high contrast LCD can show a variety of temperature and humidity information.
At the touch of a button, the user can cycle between the current temperature and humidity, along with the maximum and minimum stored values for temperature and humidity. In addition, logging and alarm status is shown using two high intensity LEDs. The data logger is supplied with a long-life lithium battery, which can typically allow logging for up to 1 year.
The user can easily set up the data logger and download logged data by plugging the unit into a PCs USB port and using the supplied Windows software. Data logger parameters that can be set up include data logger name, °C or °F temperature units, logging rate (10 sec, 1 min, 5 min, 30 min, 1 hr, 6 hrs,
12 hrs), high and low alarms, immediate, delayed or push-to-start logging, display off, on for 30 seconds after button press, or permanently on, and data rollover (allows unlimited logging periods by overwriting the oldest data when the memory is full). Downloaded relative humidity, temperature and dew point data saved in text format can be graphed, printed and exported to other applications for further analysis.
SPECIFICATIONS
TEMPERATURE
Range: -35 to 80°C (-31 to 176°F)
Repeatability: ±0.1°C (±0.2°F)
Resolution: 0.5°C (1.0°F)
Accuracy (OM-EL-USB-2-LCD); ±0.5°C (±1.0°F) typical; ±2.0°C (±4.0°F) max
Accuracy (OM-EL-USB-2-LCD-PLUS); ±0.3°C (±0.6°F) typical; ±1.5°C (±3.0°F) max
HUMIDITY
Range: 0 to 100% RH
Repeatability: ±0.1% RH (short term)
Resolution: 0.5% RH
Accuracy [20 to 80% RH] (OM-EL-USB-2-LCD); ±3.0% RH typical; ±5.0% RH max
Accuracy [20 to 80% RH] (OM-EL-USB-2-LCD-PLUS); ±2.0% RH typical; ±4.0% RH max
DEW POINT
Accuracy (Overall Error in the Calculated Dew Point,
for RH Measurements Between 40 to 100 %RH @ 25°): ±1.1°C (±2°F)
GENERAL
Memory: 16,379 temperature and 16,379 relative humidity readings
Logging Interval: 10 sec to 12 hrs
Software: Windows XP/Vista/7 and 8 (32-bit and 64-bit)
Operating Temperature Range: -35 to 80°C (-31 to 176°F); at temperatures below -20°C (-4°F)
the LCD display will exhibit slower response times of approx 10 seconds
Alarm Thresholds: High/low alarm thresholds for %RH and temperature selectable in software
Start Date/Time: Selectable in software
Status Indicators (LEDs): Red and green
Power: 1/2 AA, 3.6V lithium battery (included)
Battery Life: 1 year typical (depends on sample rate, ambient temperature and use of LCD display)
Weight: 40 g (1.4 oz)
Dimensions: See dimensional drawing on PDF
At the touch of a button, the user can cycle between the current temperature and humidity, along with the maximum and minimum stored values for temperature and humidity. In addition, logging and alarm status is shown using two high intensity LEDs. The data logger is supplied with a long-life lithium battery, which can typically allow logging for up to 1 year.
The user can easily set up the data logger and download logged data by plugging the unit into a PCs USB port and using the supplied Windows software. Data logger parameters that can be set up include data logger name, °C or °F temperature units, logging rate (10 sec, 1 min, 5 min, 30 min, 1 hr, 6 hrs,
12 hrs), high and low alarms, immediate, delayed or push-to-start logging, display off, on for 30 seconds after button press, or permanently on, and data rollover (allows unlimited logging periods by overwriting the oldest data when the memory is full). Downloaded relative humidity, temperature and dew point data saved in text format can be graphed, printed and exported to other applications for further analysis.
SPECIFICATIONS
TEMPERATURE
Range: -35 to 80°C (-31 to 176°F)
Repeatability: ±0.1°C (±0.2°F)
Resolution: 0.5°C (1.0°F)
Accuracy (OM-EL-USB-2-LCD); ±0.5°C (±1.0°F) typical; ±2.0°C (±4.0°F) max
Accuracy (OM-EL-USB-2-LCD-PLUS); ±0.3°C (±0.6°F) typical; ±1.5°C (±3.0°F) max
HUMIDITY
Range: 0 to 100% RH
Repeatability: ±0.1% RH (short term)
Resolution: 0.5% RH
Accuracy [20 to 80% RH] (OM-EL-USB-2-LCD); ±3.0% RH typical; ±5.0% RH max
Accuracy [20 to 80% RH] (OM-EL-USB-2-LCD-PLUS); ±2.0% RH typical; ±4.0% RH max
DEW POINT
Accuracy (Overall Error in the Calculated Dew Point,
for RH Measurements Between 40 to 100 %RH @ 25°): ±1.1°C (±2°F)
GENERAL
Memory: 16,379 temperature and 16,379 relative humidity readings
Logging Interval: 10 sec to 12 hrs
Software: Windows XP/Vista/7 and 8 (32-bit and 64-bit)
Operating Temperature Range: -35 to 80°C (-31 to 176°F); at temperatures below -20°C (-4°F)
the LCD display will exhibit slower response times of approx 10 seconds
Alarm Thresholds: High/low alarm thresholds for %RH and temperature selectable in software
Start Date/Time: Selectable in software
Status Indicators (LEDs): Red and green
Power: 1/2 AA, 3.6V lithium battery (included)
Battery Life: 1 year typical (depends on sample rate, ambient temperature and use of LCD display)
Weight: 40 g (1.4 oz)
Dimensions: See dimensional drawing on PDF
PDFs & Manuals
Software & Drivers
Show Ratings & Reviews
Rated 4 out of
5
by
henry from
senior field service rep
really great. nice and small and easy to use. the only issue is replacing the batteries. the tab to push down to separate the body is the weak point. this can break and then the 2 pieces will not stay together. a little tape is needed to keep the pieces together
Date published: 2018-09-22
Rated 5 out of
5
by
Bob S from
Works great.
Data logger with great features. Incredibly flexible range of sampling rates and the software plots data for you. Paints a great picture. I am a homeowner with a humidity problem and this is a great tool to help understand the source. I have already recommended it to two HVAC professionals. Bob Schaale at Omega was very helpful by recommending the right product
Date published: 2017-12-20
Are these data Loggers calibrated when received and how do you verify the calibration?
To order these units with a NIST certification you would order OM-EL-USB-2-LCD-NIST or OM-EL-USB-2-LCD-PLUS-NIST. To verify the units you would need a humidity / temperature chamber, unfortunately, we do not sell those. Please email us at das@omega.com if you need any further assistance. Thank you.
Date published: 2022-05-10
hi can i get a standard operating procedure for this make model please?
Hello Sangeetha,
If you navigate to product page and scroll down there is a link to download the manual:
https://www.omega.com/en-us/control-and-monitoring-devices/data-loggers/om-el-usb-2-lcd-series/p/OM-EL-USB-2-LCD?searchterm=OM-EL-USB-2-LCD
Date published: 2022-05-10
I cannot use OM-EL-USB-2-LCD software EasyLog USB, version 7.5 on Windows 10. When I try to start or stop the software, it gives an error: Unhandled exception has occured..... InvalidArgument=Value of '1' is not valid for 'index'. Parameter name: inde
Thank you for your question. The issue you are having has to be reviewed by our Data Acquisition Dept. Please e-mail them directly-DAS@OMEGA.COM.
Date published: 2022-05-10
Is this compatible with a mac? Is it possible to print charts ?
The software requires Windows operating system 7 thru 10 it is not compatible with MAC operating system. The software is basic the best option is to export the data file to Excel and create graph from Excel.
Date published: 2022-05-10
I have a bunch of these and the orings are either missing or dry rotted. Is it possible to order replacement orings from Omega?
Thank you for your question. We will require more information to help on this question, please email the DAS team at DAS@dwyeromega.com
Date published: 2024-07-11
Manual for OM-EL-USB-2. Is there one available? I only see one for OM-EL-USB-1 How would I change the battery?
Hi Chuck74,
Thank you for the inquiry. Link to manual for this model;
https://assets.omega.com/manuals/M4218.pdf
To access and change battery;
- remove protective cover on the USB connector
- the light grey body pulls apart from the dark grey body. You can press in detents lightly to help with removal.
Best regards
Omega
Date published: 2022-07-11
What are the conditions for the typical accuracies shown for standard and plus version?
Thank you for your question. Please download the manual from the product page which will have all the information.
Date published: 2024-02-09
If data logging 4 to 8 hours per log event for say over a week or several weeks (assuming memory not full), is there a log "number" that would show on the device that could be written down to correspond to a particular log event?
Thank you for your question. We will require more information to help on this question, please email the DAS team at DAS@OMEGA.COM
Date published: 2023-08-29