1-800-663-4209 | Welcome to our new website, found an issue or bug? Please report it here

OM-CP-PR140-Series
Submersible Pressure Data Logger for High Temperature
Models In Stock
- Flush Mount (OM-CP-PR140)
- or 1/8 NPT Pressure Port
- (OM-CP-PR140-NPT)
- Submersible (IP68)
OM-CP-PR140-Series
From
$
1,120.00
Models In Stock

Product Overview
- Accuracy Detail Calibration Accuracy : Pressure: ±0.435 psi @ 20 to 140ºC; Time Accuracy : ±1 minute/month at 20ºC; Pressure Repeatability : ±0.5% FSR; ±0.2% typical
- Approvals CE
- Battery life up to 2 yr
- Battery Size and Type 3.6V Li-ion
- Baud Rate 125000
- Calibration Digital Calibration through Software with Montly Calibration interval
- Calibration Accuracy Pressure: ±0.435 psi
- Data Interface USB
- Diameter 25.4 mm
- Frequency Response Pressure Response Time: 0.1 ms
- Humidity Range 0 to 100% RH
- Length 50.8 mm
- Materials Stainless Steel
- Measurement Type Pressure
- Memory 32,700 Readings
- Memory Note 32,700 readings; software configurable memory wrap
- Millivolt Range N/A
- Mounting Flush Mount
- Number of Measurements 1
- Operating Temperature, Max 140 °C
- Operating Temperature, Min -20 °C
- Measurement Pressure, Max 72.5 psi
- Measurement Pressure, Min 0 psi
- Pressure Repeatability ±0.5%
- Pressure Sensor Type Silicon Strain Gage
- Measurement Pressure Type Absolute
- Real Time Recording The device may be used with PC to monitor and record data in real-time
- Resolution Pressure : 0.0015 psi
- Sampling Rate 1 Hz (1 update/second)
- Sampling Rate Note 1 second to 1 every 24 hours
- Software Windows XP SP3 / Vista / 7 and 8 (32 and 64 - bit)
- Start Modes Delayed Start, Immediate Start
- Stop Modes Adjustable Delay Stop
- Supply Power Type Battery
- Time Accuracy ±1 minute/month
- Weight 85 g
The OM-CP-PR140 is a pressure data logger designed for use in autoclave validation and mapping. This rugged device can withstand temperatures up to 140 °C (284 °F) and is completely submersible (IP68). The OM-CP-PR140 is built with a precision stainless steel pressure gauge. The device has an accuracy of ±0.03 Bar ( ±0.435 psi), which can be achieved over a wide temperature range, from 20 to 140 °C (68 to 284 °F).
The OM-CP-PR140 can be programmed to take readings as often as once per second (1 Hz), and has non-volatile memory that can store up to 32,700 measurements. Simply place the OM-CP-PR140 in the OM-CP-IFC400 docking station, connect to an available USB port on your computer, and the device can be started, stopped or downloaded using our user friendly software. Graphical, tabular and summary data is provided for analysis and can be viewed in psia, mmHg, bar, Torr and kPa. The data can also be automatically exported to Excel® for further calculations. The storage medium is non-volatile solid state memory, providing maximum data security even if the battery becomes discharged.
The OM-CP-PR140 can be programmed to take readings as often as once per second (1 Hz), and has non-volatile memory that can store up to 32,700 measurements. Simply place the OM-CP-PR140 in the OM-CP-IFC400 docking station, connect to an available USB port on your computer, and the device can be started, stopped or downloaded using our user friendly software. Graphical, tabular and summary data is provided for analysis and can be viewed in psia, mmHg, bar, Torr and kPa. The data can also be automatically exported to Excel® for further calculations. The storage medium is non-volatile solid state memory, providing maximum data security even if the battery becomes discharged.
PDFs & Manuals
Software & Drivers
Accessories
accessories
consumables
3.6V High Temperature Replacement Lithium Battery for High Temperature Pressure Data Logger
$111.77
30 In Stock
Show Ratings & Reviews
Does the OM-CP-PR140 measuure vaccum pressure as well as?
The unit is not designed to measure or be used in a vaccuum
Date published: 2022-02-22