Product Overview
- Accuracy Class B
- Accuracy Detail Class B per IEC60751
- Air Velocity, Max 1 m/s
- Process Temperature, Min -200 °C
- Series ID 1PT100K-RTD-ELEMENTS
- Temperature Coefficient 3850 ppm/K
- Temperature Sensor Type RTD
- Water Velocity, Max 0.4 m/s
Supported Applications
The Omega Standard Accuracy Ceramic Wire Wound Platinum RTD Elements are Resistance Temperature Detectors (RTDs) designed for extremely low or high temperatures. These elements, constructed by placing a coiled platinum wire into a ceramic core and sealing the ends with glass, are tolerant of thermal shocks. The series is specified for use in dry environments only.
Operating Conditions & Performance
The 1PT100K-RTD-ELEMENTS series operates across multiple temperature ranges depending on the specific model configuration: -200 to 600 °C (-330 to 1110 °F), -200 to 850 °C (-330 to 1560 °F), and -50 to 400 °C (-60 to 750 °F). The elements meet Class B Tolerance (Per IEC751) with an Alpha value of 0.00385.
Response times vary by model size and medium velocity:
- Flowing Water V = 0.4 m/sec: 50% Response ranges from 0.2 to 0.4 sec; 90% Response ranges from 0.3 to 1.4 sec.
- Moving Air V = 1 m/sec: 50% Response ranges from 3.0 to 26.8 sec; 90% Response ranges from 8.8 to 90.0 sec.
The series offers limited vibration resistance, which depends on the mounting method. Wetted materials include platinum wire, a ceramic core, and glass seals.
Configuration Options
This series supports various dimensional configurations to accommodate tight spaces or specific insertion requirements. Standard electrical connections feature 10 mm bare leads; welded insulated extension leads are available as an option. OMEGA's RTD MI Cable is also offered in a wide selection of conductor configurations and sheath materials.
Dimensions and Resistance
The series includes elements with insertion lengths from 8 mm to 30 mm and diameters ranging from 1.0 mm to 4.5 mm. Nominal resistance options include:
- Single Element (1 x): 100 Ω, 500 Ω, and 1000 Ω.
- Dual Element (2 x): 100 Ω configurations are available for specific models such as *2PT100K2517 and *2PT100KN3045.
Tight Space Configurations
Specific models (A2105*, A2132*, A2142*) are designed for tight spaces with leads perpendicular to the body for surface measurements. These variants feature dimensions such as 2 L x 2.4 W x 2 mm D and operate in the -50 to 400 °C (-60 to 750 °F) range.
Key Product Differences
Variants within the series differ primarily by physical dimensions, nominal resistance values (1 x or 2 x), and temperature ranges. Self-heating error varies significantly based on size; for example, in flowing air V = 1 m/sec, self-heating errors range from 0.01 °C/mW to 0.28 °C/mW depending on the model.
Model numbers follow a structured guide indicating dimensions, nominal resistance, temperature range, and self-heating error (e.g., 1PT100KN815). Pricing is dependent on lead style. Dual element models are identified with an asterisk in documentation notes.
PDFs & Manuals
Show Ratings & Reviews
Rated 4 out of
5
by
Javier from
poor packaging
The product was packaged very good and in a secure way for the ship, but the dimension of the boxes were bigger than of necesary . However the PT100 elements were not in the original blister where they could are more protected.
Date published: 2018-05-15
Rated 4 out of
5
by
vbermud from
Bare leads way too short
It would be extremely helpful if the bare leads came longer than 10 mm
Date published: 2020-06-06
Rated 5 out of
5
by
Luz Marina from
Your products hace excelent quality
Im happy with my product and im glad to buy more products from you.
Date published: 2018-10-17
What are the leads of these things made of ? They are very, very soft, and melt when soldered with 60/40 tin lead.
So what is the recommendation to solder, if they melt at soldering temperatures ? I suspect that they are forming another alloy with the tin-lead, with a melting point below the headline number
Date published: 2022-06-23
Is the ceramic body submersible in water?
Thank you for your inquiry rn_intc. No, it is not submersible in water.