Need assistance? Call us: 1-800-663-4209
LC60
Single Point Load Cell with IP65 Rating
Models In Stock
- Aluminum-Alloy IP65 Single Point Load Cell.
- Colorless Anodized Aluminum
- Moisture-proof Sealant
- Ideal for OEM Applications
Product Overview
- Approvals OIML R60
- Cable Length 1.5 ft
- Dead load 0% FS
- Electrical Connection Cable
- Enclosure Ratings IP65
- Input Resistance 406 ± 6 Ω
- Insulation Resistance ≥ 5000 MΩ
- Load Cell Type Bending Beam
- Operating Temperature, Max 65 °C
- Operating Temperature, Min -35 °C
- Output Resistance 350 ± 3 Ω
- Overload, Max 150% FS
- Process Temperature, Max 40 °C
- Process Temperature, Min -10 °C
- Safe Overload 120% FS
- Series ID LC60
- Storage Temperature, Max 70 °C
- Storage Temperature, Min -40 °C
- Supply Power Note 5 to 12 Vdc, 18 Vdc max
- Supply Voltage 5 to 12 Vdc
- Wiring/Connector Table Green: (+) Output; White: (-) Output; Black: (-) Input; Red: (+) Input
- Zero/Balance Error ± 2%
Supported Applications
The Omega LC60 series comprises single point load cells designed for force measurement in weighing applications. As indicated by the product identity and variant naming, these sensors are configured as bending beam elements suitable for platforms requiring off-center loading tolerance.
Operating Conditions & Performance
This series is identified with an IP65 rating, indicating protection against dust ingress and low-pressure water jets. Specific performance parameters such as capacity ranges in kilograms (KG), accuracy classes, output signals, or environmental limits are not detailed in the current documentation bundle.
Available Capacities
The series encompasses a range of capacities across multiple model families:
- LC61SP family: 600G, 1KG, 2KG, and 3KG.
- LC62SP family: 6KG, 10KG, 15KG, 20KG, 30KG, 35KG, 40KG, and 50KG.
- LC63SP family: 60KG, 100KG, and 200KG.
Configuration Options
The LC60 series is offered in three distinct model families (LC61SP, LC62SP, LC63SP), each covering specific capacity segments. The variants are designated as single point load cells within the bending beam family.
Model Families and Capacities
- LC61SP: Low-capacity range from 600G to 3KG.
- LC62SP: Mid-range capacity from 6KG to 50KG.
- LC63SP: Higher capacity options including 60KG, 100KG, and 200KG.
Key Product Differences
Variation within the LC60 series is defined by the model prefix (LC61SP, LC62SP, or LC63SP) which corresponds to specific capacity ranges. The suffix in each part number denotes the rated load in grams (G) or kilograms (KG). No additional distinctions regarding housing materials, process connections, output types, or certifications are provided in the available source data.
Show Ratings & Reviews
Rated 5 out of
5
by
Anonymous from
Accurate Load Cell, Reasonable Price
I used this load cell to create my own scale for a custom engineering research project. The results are comparable to a well-maintained scale in our lab but now I can log each the dynamic changes of the load in my data acquisition software of choice.
Date published: 2024-10-02
Rated 5 out of
5
by
Chance BG from
Great repeatability
Product worked better than expected. Was expecting to see much more fluctuation in the load reading.
Date published: 2023-10-05
What is the maximum strain for the LC61SP-2kg?
The maximum calibrated load range is 0 to 2 kilograms. In case of occasional accidental overload the load cell can handle up to 120% of it's calibrated load range.
Date published: 2020-08-22
What is the creep specification of LC61SP-600G, LC61SP-3KG and LC62SP-6KG?
Hello, thank you for contacting Omega Engineering. We do not have the exact creep spec for this load cell , but the nominal value is 0.01 to 0.1 percent of rated output.
For additional resource please see: https://www.nist.gov/document/vdi-312pdf
https://www.nist.gov/document/vdi-312pdf
Date published: 2023-03-17
On aluminum load cells ,LC61SP-3KG. Are connection wires included? What is recommended the supply voltage? Is there a transfer function that can be used to convert Vout measurement to mass? What is the sensitivity? Thanks Victor
Thank you for your question?
Output is a linear response over the sensors 0 to 3KG load range. Sensitivity of the LC61 is 1.0mV/V, +/- 0.1. A 0.45 meter long is attached to the unit. Wiring code= Red (+Exc), Black (-Exc), Green(+ Signal), White (- Signal). Maximum Excitation 18VDC.
Date published: 2022-10-20
What is the bandwidth of LC61SP-600 G? Is it compatible with a DAQ sampling over 2 kHz?
These load cell a simply full wheatstone bridge strain gage bases devices. Typical response time to a load change is 1 to 2 milliseconds. They should be compatible with a high speed DAQ system.
Date published: 2022-06-01
For the LC61SP-2Kg is the output range 0-10V or 0-5V ?
Hello, this is not an amplified output load cell.
Output (mV/V):
LC61: 1.0 ±0.1 mV/V
Date published: 2024-12-17
What is the Full Load Deflection of the LC62SP-50KG
Hello thank you for contacting Omega Engineering. We do not have the exact data for full load deflection, but the nominal range for beam load cells is 0.003 to 0.015"
Date published: 2024-01-17
What is the %RO Accuracy of Item# LC62SP-35KG?
It is OIML accuracy class C3 (0.0230%)
Date published: 2023-03-06
Is the LC61SP-2KG a half bridge or full bridge configuration ?
Full bridge;
Input Resistance : 40 ohm ±6
Output Resistance ?: 350 ohm ±3
Date published: 2024-12-18