Need assistance? Call us: 1-800-663-4209
OMNI-AMP-IV-Srs-TC-Amp
Rugged Epoxy Encapsulated Thermocouple Amplifier
- 100% Encapsulated
- 2µV/°C Zero Stability
- 0.025°F/°F Reference Junction Stability
- Up or Downscale TC Burn-out
OMNI-AMP-IV-Srs-TC-Amp
From
$
642.26
Product Overview
- Bandwidth 3 dB down at 1.5 Hz 40 dB/decade roll-off
- CMRR 120 dB min @ 60 Hz
- Cold Junction Compensation ± 0.045°C/°C
- Common Mode Voltage +10 to -10 Vdc or 20 Volts peak-to-peak ac
- Electrical Output ± 10 V into 10 kΩ min
- Height 73 mm
- Impedance Input 2 MΩ
- Input Noise 0 to 1.5 Hz: 1.4 µVRMS (RTI)
- Input Range 40 to 1500
- Length 51 mm
- Long Term Stability Zero: ± 1µV/Month;Span: ± 0.05%; Stability of zero point - 2 µV/°C , Span stability - ± 0.01%/°C
- NMRR 80 dB min @ 60 Hz
- Non-Linearity ± 0.005 %
- Operating Temperature, Max 55 °C
- Operating Temperature, Min 0 °C
- Output Signal Voltage
- Programmable Functions Trim Pot
- Protection Output Protection: Continuous short circuit proof, Input Protection: Up to 115 Vac or Vdc common mode or differential Output Port
- Resolution Infinite
- Series ID OMNI-AMP-IV-Srs-TC-Amp
- Storage Temperature, Max 80 °C
- Storage Temperature, Min -20 °C
- Supply Power 115 Vac @ 3 VA max
- Supply Power Type AC
- Weight 455 g
- Width 95 mm
Supported Applications
The Omega OMNI-AMP IV is a high-performance, rugged epoxy encapsulated thermocouple amplifier designed for signal conditioning in industrial environments. It provides amplification and cold junction compensation for Type E, J, T, or K thermocouples using a 10 mV/°F scale factor. The series also accommodates Types R, S, and B when configured with the optional 1 mV/°F scale factor option.
Operating Conditions & Performance
The amplifier operates within an ambient temperature range of 0 to 55°C (32 to 130°F) and supports a storage temperature from -20 to 80°C (-4 to 175°F). It delivers a ±10 VDC output into a minimum load resistance of 10 kΩ. Performance specifications include an input impedance of 2 MΩ (differential), infinite resolution, and nonlinearity of ±0.005%. Stability after a 30-minute warm-up is rated at 2 μV/°C (1 μV/°F) for zero drift and ±0.01%/°C (±0.006%/°F) for span stability, with long-term span stability of ±0.05%. The unit features a frequency response that rolls off 40 dB/decade after dropping 3 dB at 1.5 Hz.
Electrical protection includes input tolerance up to 115 VAC or DC common mode or differential, and the output is continuous short circuit proof. Noise rejection capabilities include a Common Mode Rejection Ratio (CMRR) of 120 dB minimum @ 60 Hz and a Normal Mode Rejection ratio of 80 dB minimum @ 60 Hz.
Key Performance Metrics
- Cold Junction Compensation: ±0.045 °C/°C (±0.025 °F/°F)
- Noise (RTI) 0 to 1.5 Hz: 1.4 μVRMS
- Common Mode Range: +10 to -10 VDC or 20 Volts peak-to-peak AC
- Power Consumption: 3 VA maximum @ 115 VAC (or 230 VAC)
Configuration Options
The OMNI-AMP IV series is configurable across several dimensions to meet specific installation and measurement needs. The unit features rugged terminal strips for electrical connections, requiring a screwdriver for termination.
Housing and Mounting
- Enclosure: Rugged epoxy encapsulated housing (100% encapsulated).
- Dimensions: 73 H x 95 W x 51 mm D (2.87" x 3.75" x 2.0").
- Weight: 455 g (18 oz).
- Mounting Options: Supplied with bolts for standard mounting; a DIN rail mounting kit is available as accessory model number DIN400.
Signal and Power Configuration
- Scale Factor: 10 mV/°F (standard) or 1 mV/°F (optional for Types R, S, B).
- Burnout Direction: Configurable as Up Scale (-01 code) or Down Scale (-02 code).
- Power Supply: Available in 115 VAC @ 3 VA maximum (code -115, range 105 to 125 Volts AC, 50-60 Hz) or 230 VAC (code -230, range 210 to 250 Volts AC, 50-60 Hz).
Key Product Differences
The series is distinguished by the thermocouple type supported and the resulting scale factor. Models ending in "3" (e.g., OMNI-AMP IV-13-115) utilize a 10 mV/°F output for Types J, K, T, and E. Models ending in "4" (e.g., OMNI-AMP IV-14-115) utilize a 1 mV/°F output specifically for Types R, S, and B.
Model numbers further differentiate the burnout direction: codes starting with "1" indicate Up Scale (increase in temperature), while codes starting with "2" indicate Down Scale (decrease in temperature). The final digits denote the power line voltage compatibility (-115 for 105 to 125 VAC, -230 for 210 to 250 VAC).
Wire color coding is provided per thermocouple type: J (White/Red), K (Yellow/Red), T (Blue/Red), E (Purple/Red), and R/S/B (Black/Red). The terminal strip assignments include dedicated terminals for OUTPUT, CMN, AMBIENT OUT, +INPUT, -INPUT, and AC power connections.
PDFs & Manuals
Show Ratings & Reviews
What is the isolation voltage between the thermocouple and the amplifier? For example if you have a thermocouple attached to a sample and charge the sample with a bias, at what point is the isolation compromised?
Hi Marcus,
Thank you for the inquiry. Input protection to 115 VAC or DC. Over 120 dB CMRR at 60 Hz. Manual gives good detail on the hardware.
https://assets.omega.com/manuals/M2074.pdf
Best regards
OMEGA
Date published: 2021-01-19