Product Overview
  • Electrical Connection Screw Terminal
  • Enclosure Material Stainless Steel
  • Filter Hydrophobic Filter
  • Height 1.25 in
  • Insulation Etched Teflon (PTFE) Colored Leads 
  • Length 4.51 in
  • Load Resistance 4K Ohms Minimum
  • Current, Max 8 mA
  • Measurement Type Humidity, Temperature
  • Mounting Orientation Wall Mount
  • Operating Temperature, Max 60 °C
  • Operating Temperature, Min -40 °C
  • Output Detail 0-5 VDC: 12 - 40 VDC / 18 - 28 VAC
  • Probe Diameter 0.75 in
  • Relative Humidity 0 to 95%
  • Relative Humidity Sensor Type Capacitive with Hydrophobic Filter
  • Repeatability ± 0.05°C
  • Response Time 20 Seconds Typical
  • Series ID RH-Combo-Wall-Plate-Platinum RTD
  • Storage Temperature, Max 65 °C
  • Storage Temperature, Min -40 °C
  • Supply Power Type DC
  • Supply Voltage 24 Vdc
  • Weight 0.235 lb
  • Width 2.76 in
  • Wire Gauge 26 to 16 AWG

Supported Applications

The ACI Relative Humidity Wall Plate with Platinum RTD Series is a relative humidity transmitter designed for use in commercial heating and cooling building management systems. The series utilizes a thermoset polymer capacitive sensing element to deliver proportional analog voltage output signals suitable for electronic controllers.

Documented applications include:

  • Pharmaceutical facilities
  • Hospitals and Operating Rooms
  • Vivariums
  • Clean Rooms
  • Process Control environments
  • Wash Down Environments
  • Stability Chambers

The transmitter features a factory-applied hygroscopic filter to provide resistance against moisture, dust, and other chemicals. A temporary plastic sensor cover is included for protection during wash down applications.

Operating Conditions & Performance

This series operates within an ambient temperature range of -40 to 140°F (-40 to 60°C) with a storage temperature range of -40 to 149°F (-40 to 65°C). The relative humidity measurement range is 0-100%, with an operating RH range of 0 to 95% RH, non-condensing. PCBs are conformally coated for added protection.

Performance specifications include:

  • RH Accuracy @ 77°F (25°C): +/- 2% or 3% from 10 to 95%
  • RH Stability | Repeatability | Sensitivity: Less than 2% drift / 5 years | 0.5% RH | 0.1% RH
  • RH Response Time (T63): 20 Seconds Typical
  • RH Transmitter Stabilization Time: 30 Minutes (Recommended time before doing accuracy verification)

The Platinum RTD temperature sensor follows DIN EN 60751 (IEC 751) standards with a Temperature Coefficient of 3850 ppm/°C. The tolerance class is +/-0.06% Class A, defined by the formula: +/-°C = (0.15°C + (0.002 * |t|)).

Configuration Options

The series supports field selection of output signals and supply voltages via integral DIP switches. Units include 0%, 50%, and 100% test options to verify operation and installation, along with increment and decrement calibration DIP switches for field calibration without replacing the sensing element.

Electrical & Output Configuration

  • RH Supply Voltage: 0-5 VDC: 12 - 40 VDC / 18 - 28 VAC; 0-10 VDC: 18 - 40 VDC / 18 - 28 VAC (Reverse Polarity Protected)
  • RH Output Signal: 3-wire: 0-5 or 0-10 VDC
  • RH Supply Current (VA): 8 mA maximum (0.32 VA)
  • RH Output Load Resistance: 4K Ohms Minimum

Sensor & Housing Options

  • Housing Material: 430 Stainless Steel (Brushed Stainless Steel Finish) designed to mount over a single gang junction box.
  • RH Sensor Type: Capacitive with Hydrophobic Filter; Sintered Filter Material is 304 Series Stainless Steel.
  • Foam Material | Foam Thickness: Cross-linked LPDE (White) | 0.25" (6.35 mm); FMVSS-302 Flammability Rating.
  • Platinum RTD Configurations: Available in RHx-100-xW Series (100 Ohms nominal @ 32°F) and RHx-1K-xW Series (1000 Ohms nominal @ 32°F).

Certifications & Accessories

  • Agency Approvals: CE, UKCA, RoHS, WEEE.
  • NIST Calibration: Three point NIST Calibration Certificates are available (Default Test Points: 20%, 50% & 80%).
  • Accessories: A/SINTERED FILTER (3/8" Sintered Filter) and A/1" VINYL PULL CAP (1" EZ Vinyl Filter Cover).

Key Product Differences

The series is configured via a model number guide: A/ - <SENSOR SERIES> - <ACCURACY> - <MODEL SERIES> - <NUMBER OF WIRES> - <CONFIGURATION> - <OUTPUT SIGNAL> - <NIST>.

Key variations include:

  • Accuracy Selection (RH2/RH3): +/- 2% or +/- 3%
  • Model Series (100/1K): 100 Ohm Platinum RTD or 1K Ohm Platinum RTD
  • Number of Wires (2W/3W): Two Wires or Three Wires with specific wire color codes for each series.
  • Output Signal: Field selectable between 0 to 5 VDC and 0 to 10 VDC via DIP switches.

Accessories

SINTERED FILTER FOR RH DUCT & REMOTE PROBE
$51.00 / each
4 In Stock
  • bvseo_sdk, java_sdk, bvseo-4.0.0