Product Overview
  • Accuracy ± 0.02 %
  • Cable Length 10 ft
  • Hysteresis 0.0002
  • Linearity 0.0001
  • Platform Size 12 x 14 in
  • Series ID LSC7000
  • Supply Voltage 15 V
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 4ms
  • REVIEWS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;

Scale is 2mv/V: when powered with a 12vdc supply, what would the scaling end points be? on a 0-10VDC input: full scale min = ? full scale max = ?

Asked by: Richard23
Richard, if you are powering the load cell with 10 volts DC, at no load you will see 0 mV out and at full load you will see 20 mV out (Nominal) If you are powering the load cell with 12 volts, at no load you will see 0 mV and at full load you will see 24 mV. (Nominal)
Answered by: SAE DVJ
Date published: 2023-06-12
  • y_2026, m_5, d_5, h_11CST
  • bvseo_bulk, prod_bvqa, vn_bulk_3.0.43
  • cp_1, bvpage1
  • co_hasquestionsanswers, tq_1
  • loc_en_US, sid_LSC7000-10, prod, sort_[SortEntry(order=NUM_ANSWERS, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getContent, 3ms
  • QUESTIONS, PRODUCT
en-US
en
Save Configuration