PHCN-85-Controller
pH/ORP Transmitter and Controller with TFT color display
Product Overview
- Cutout Dimensions 91 × 91 mm
- Data Interface RS485
- Display Type TFT
- Display Type Note 3.5", 320 × 240 TFT Color Display, Virtual Pointer or Digital Style, Multi-Parameter
- Enclosure Ratings IP65, NEMA Type 4X
- Height 96 mm
- Length 94 mm
- Measurement Resolution pH: 0.01, ORP: mV, 0.1 mV
- Mounting Orientation Panel Mount
- Number of Relays None
- Operating Humidity 0 to 85% RH
- Operating Temperature, Max 50 °C
- Operating Temperature, Min 0 °C
- ORP Accuracy ± 5 mV
- ORP Measurement Range -999 to +999 mV
- pH Accuracy ± 0.1 pH
- PH Measurement Range 0 to 14.00 pH
- Power Consumption 5.5 W
- Process Temperature, Max 0 °C
- Process Temperature, Min 100 °C
- Relay Note Contact photoelectric relay output, programmable; temperature, pH/ORP output can be set for pH/ORP/temperature or timing ON/OFF or PID control Relay: Resistance load 50 mA maximum AC/DC 30V
- Series ID PHCN-85-Controller
- Storage Temperature, Max 60 °C
- Storage Temperature, Min -20 °C
- Temperature Accuracy ± 0.5 °C
- Temperature Resolution 0.1 °C
- Temperature Sensor Type 1000 Ω RTD
- Width 96 mm
Supported Applications
The Omega PHCN-85 is a panel-mount pH/ORP transmitter and controller designed for high-level water quality management, automatic running systems, and inline monitoring. Documented applications include:
- High Purity Water: Management control of high pure or ultra-pure water in electronics, electric power, pharmaceuticals, fine chemicals, clinical medical science, and life-science research.
- Industrial Process Solutions: Salinity analysis for water treatment, desalination, concentrate management, circulating cooling water, industrial coating, large-scale water treatment equipment, mixing, metallurgy, petrochemicals, and industrial cleaning.
Operating Conditions & Performance
The PHCN-85 measures pH from 0.00 to 14.00 with a resolution of 0.01 and an accuracy of ±0.1. ORP measurement ranges from -1999 to +1999 mV with a resolution of 0.1 mV and display accuracy of ±5 mV. Temperature is measured from 0 to 100°C (32 to 212°F) using a 1000 Ω RTD input, offering a resolution of 0.1°C and an accuracy of ±0.5°C.
Automatic temperature compensation functions within the range of 0 to 50°C (32 to 122°F). The unit operates in ambient temperatures from 0 to 50°C (32 to 122°F) with relative humidity ≤85% RH (non-condensing). Storage conditions are specified as -20 to 60°C (-4 to 140°F) at ≤85% RH. The enclosure provides a protection level of NEMA 4 (IP65) when the rear cover is installed.
Electrical and Output Specifications
- Power Supply: DC 24V ±4V with internal automatic polarity identification; consumption <5.5 W.
- Current Outputs (2): Isolated, adjustable, reversible 4 to 20 mA outputs supporting instrument or transmitter mode. Loop resistance is maximum 400 Ω at DC 24V with a resolution of ±0.1 mA.
- Relay Outputs (3): Contact photoelectric relay outputs programmable for pH, ORP, temperature, timing ON/OFF, or PID control. Relay load is maximum 50 mA AC/DC at 30V.
- Data Communication: RS485 Modbus-RTU protocol.
Configuration Options
The PHCN-85 features a 3.5 inch 320×240 TFT color screen with virtual pointer or digital style interface display and multi-parameter capability. Calibration supports offline/inline selection with automatic buffer recognition for up to 6-point buffer solution calibration.
Configurable Outputs
The three photoelectric switch outputs can be configured as:
- pH, ORP, or temperature output control.
- Timing output (calendar function with reserved timing).
- PID adjustment to drive pulse metering pumps or extension controls.
Accessories and Options
The series supports the following accessories for specific installation needs:- pH Electrodes: PHE-7351-15-PT1K (Industrial pH electrode with temperature, in-line or submersible) and PHE-6510B (Submersible pH electrode).
- ORP Electrode: ORE-6510B (Submersible ORP electrode).
- Mounting Assembly: PHEH-65-10 (Required for ORE-6510B or PHE-6510B).
- Buffer Solutions: PHA-4, PHA-7, and PHA-10 (pH 4.00, 7.00, and 10.00 respectively; 500 mL bottles).
Key Product Differences
The series is offered as a single model configuration: the PHCN-85 panel-mount pH/ORP transmitter and controller. The unit includes EMC design for anti-interference, password protection, and status check menus to view current transmitting status or mV values. Dimensions are 96 H × 96 W × 94 D mm (3.78 × 3.78 × 3.7") with a panel cutout of 91 × 91 mm (3.58 × 3.58"). The device carries CE approval and includes a one-year warranty.
PDFs & Manuals