Product Overview
- Accuracy Detail Reference Junction Compensation Accuracy: Type K, E, J, T, N, Platinel II; ±0.5°C or less; R, S, NiMo-Ni, CR-AuFe, WRe5-WRe26, W-Wre, U, L; ±1.0°C or less
- Case Material Steel
- Data Interface Ethernet, USB
- Data Interface Note Communication interface for high-order instruments RS232C/RS485 switchable , Ethernet and USB equipped as standard; Transfer Systems: Bulk transfer, control transfer; Communication Contents: File transfer by virtual drive connections; Number of Logging Files: 250/numbers of used groups; Logging Data: Measured data; file name (group name), time of day, month and year of recording start, tag, measured data, alarm status/types; Recording Cycle: Up to 3 groups of 12 points/group can be programmed for 100, 200 and 500 ms recording; 5 groups of 44 points/group for recording with 1s or more (total of 100 points programming possible)
- Dielectric Strength Dielectric Strength Between Channels: 1000 Vac or more between each channel (high strength semiconductor relay used)
- Digital Filter Programmable
- Display Note TFT color LCD, LCD Back Light: Auto/manual OFF function LCD Brightness: 4 levels adjustment Display Types: Measured Data Display: Trend screen, data screen, bar-graph screen Historical Trend Display: Simultaneous display with real-time trend is available Information Display: Alarm display, marker list, file list Setting Screen: Alarm, computation, memory, system, maintenance, communication, etc. Trend Screen: 12 colors selectable Display Screen: 5 screens (5 groups) Display Points: Maximum 44 points/screen/ Time Axis Direction: Vertical or horizontal Line Width: 1/3/5 dot selectable Scale Display: 4 scales Tag/Data Display (Show/hide selectable) Data Screen: Display screen, 5 screens (5 groups) Display Points: Maximum 44 points/screen Display Contents: Measured value, channel/tag, unit, alarm status Bargraph Screen: 12 colors selectable Display Screen: 5 screens (5 groups) Display Points: Maximum 44 points/screen Display Direction: Vertical or horizontal Scale Display: 1 scale Information Display: Alarm Display (Alarm activation/released history display) Marker List File List (Group data file list display)
- Display Size 142 mm (5.6 in)
- Display Type LCD
- Electrical Connection Note Power Terminals/Protective Conductor Terminals/Communications Terminals: M4.0 Measuring Input Terminals/Alarm Output Terminals/Remote Contact Terminals: M3.5
- Enclosure Color Gray
- Features Computation Points: 44
- Height 144 mm
- Impedance Input 1 MΩ
- Input Note DC Voltage, Thermocouple Input; Approximately 1 MΩ; Signal Source Resistance: Thermocouple Input (Burnout Disable)/DC Voltage Input (±2V or less): 1 kΩ or less, DC Voltage Input (±5 to ±50V): 100 Ω or less, Resistance Thermometer: Per wire 10 Ω or less (same resistance for 3 wires);Max Rating: DC voltage input (±2V or less)/thermocouple input (burnout disable), ±10 Vdc
- Input Voltage ± 60 Vdc
- Insulation Resistance 20 MΩ @ 500 Vdc
- Length 204 mm
- Measurement Type Universal
- Memory 32 MB to 2 GB Capacity: Compact Flash; 128 MB Compact Flash Card is included as Standard
- Memory Note Compact Flash card, 32 MB to 2 GB Capacity; 128 MB Compact Flash Card is included as Standard
- Mounting 1/2 DIN Panel Mount
- Operating Temperature, Max 60 °C
- Operating Temperature, Min -20 °C
- Programmable Functions alarm, Arithmetic Operations, Comparison Operations, Custom Functions, Integration Operations, Logarithmic, Logical Operations, On/Off Delay, Power Function, Square Root Function, Trig Function
- Relative Humidity 5 to 90% RH
- Relative Humidity Note Reference Operating Condition: 45 to 65% RH Normal Operating Condition: 20 to 80% RH Transportation Condition: 5 to 90% RH (note: no dew condensation)
- Sampling Rate 10 Hz (10 updates/second)
- Storage Temperature, Max 60 °C
- Storage Temperature, Min -20 °C
- Supply Power Note 100 to 240 Vac (universal power supply); 50/60 Hz,Maximum Power Consumption: 50 VA (DO: all points ON, 240 Vac),Reference Operating Condition:,Power Voltage: 100 Vac, ±1.0%,Power Frequency: 50/60 Hz ±0.5%,Normal Operating Condition:,Power Voltage: 90 to 264 Vac,Power Frequency: 50/60 Hz ±2%;Power Failure Protection: Setups and data are backed up by flash memory Clock: Lithium battery backs up RAM (minimum 5 years)
- Supply Power Type AC
- Supply Voltage 100 to 240 Vac
- Vibration Resistance 10 to 60 Hz, 0.5 G (4.9 m/s²) or less ; Impact: 40 G (392 m/s²) or less
- Warm Up Time Longer than 30 minutes
- Weight 2.2 kg
- Width 144 mm
RD9900 Series are network compatible paperless recorders with high performance and high operating function employed high visibility 5.6' TFT color LCD. High speed of sampling rate 100 ms for 12 points and high accuracy of ±0.1% were realized, and measured data is stored into internal memory and maximum 2 GB compact flash card. As it can be monitored by a web browser display on several computers on intranet or internet, FTP transfer of data file and e-mail notification are also available. The recorder can be used for various applications such as data management, research, development, remote and wide range monitoring by utilizing internet environment.
PDFs & Manuals
Accessories
accessories
Show Ratings & Reviews
Can this recorder be outfitted to measure humidity in addition to temperature? If yes, what probe is recommended and how do you program the device?
Hi Yonas,
The RD9906 accepts TC, RTD, and voltage signals. Please forward request to TEMP@OMEGA.COM for assistance selecting humidity sensor for your application.
Best regards
OMEGA
Date published: 2023-06-18
Can temp resolution be set to 0.01C? Our current system from Yokogawa has minimum resolution of 0.1C which is not fine enough.
Thank you for the inquiry. Resolution is limited to 0.1 for temperature inputs. Exception is 1 degree for platinum and tungsten TC's.
Best regards
OMEGA
Date published: 2022-12-05
what are the temperature ranges it reads?
Thank you for the inquiry. Manual page 96 of the manual provides the available sensor types and ranges.
https://assets.omega.com/manuals/M4450.pdf
Best regards
OMEGA
Date published: 2024-05-06
Is there a procedure for field calibration you can provide? Thanks Bruce
Hi Bruce18,
Thank you for the inquiry. Please refer to chapter 16 of the operation manual.
Best regards
OMEGA
Date published: 2022-03-03
Does Omega buy back these machines? Thanks
Thank you for your question. Omega does not buy back any products.
Date published: 2022-07-05