Skip to search Skip to main content Skip to navigation

1-800-663-4209 | Welcome to our new website, found an issue or bug? Please report it here

We noticed you are in , please visit your local site by clicking the below link for regional offers and support.

HHPT-51

Handheld Particle Counter With SD Card Data Recorder and USB Interface

Choose a Variant

Measurement Type

$6,342.99 / each

Item# HHPT-51

Only 9 left

Limited Stock

Product Specs

<< See all Product Models
Brand
Accuracy ± 7 %
Approvals Complies with CE, JIS B9921; 1997, ISO 21501-4
Battery Size and Type Li-ion
Calibration PSL particles in air
Charging Time 2 hours (fast charge mode), 27 hours (trickle charge mode)
Coincidence Loss 5% at 2,000,000 particles per cubic ft.
Commodity Code 9027.89.8000
Counter Input Note Counting Efficiency: 50% at 0.3 µm; 100% for particles > 0.45 µm (per JIS B9921:1997)
Data Interface USB
Device Modes audio, Concentration, totalize
Display Type TFT
ECCN EAR99
Height 11 cm
Length 6.7 cm
Measurement Type Air Contamination
Memory 500 sample records (rotating buffer)
Number of Channels 6
Number of Channels Note Channels: 0.3, 0.5, 1.0, 2.0, 5.0, 10.0 µm
OkForExport true
Operating Temperature, Max 40 °C
Operating Temperature, Min 10 °C
Operating Time 4 hrs
Relative Humidity 20% to 90%
Source of Light Laser diode
Storage Humidity, Max Up to 90%
Storage Temperature, Max 50 °C
Storage Temperature, Min -10 °C
Supply Power Note AC adapter, 12 Vdc at 3 A, 100 to 240 Vac, 50 to 60 Hz
Supply Power Type AC
Supply Voltage 12 Vdc
Time Delay 0 to 24 hours
Vacuum Rating Internal pump flow controlled
Variables Logged Date, time, counts, relative humidity, temperature, sample volumes, alarms and labels
Volumetric Flow Rate, Max 2.83 l/min
Weight 875 g
Width 21.8 cm
Zero Count 1 count/5 minute (per JIS B9921:1997)