DVG-64A

0 to -30 inHg, Vacuum

Choose a Variant

Measurement Pressure Type

$353.15

Item# DVG-64A

In Stock

Lead Time (If not in stock): 8 weeks

Product Specs

<< See all Product Models
Brand
Accuracy Class > 0.25%
Accuracy Detail +/-10% or +/-10 Micron, whichever is larger (50 to 2000 Microns) at 75°F
Auto-Shutoff 10 min above 19000 Microns, 1 hr below
Battery Size and Type Alkaline AA
Battery life 600 hr
Display Type Backlit LCD
Height 5.5 in
Length 1.5 in
Measurement Pressure Type Vacuum
Measurement Pressure, Max 190000 Pa
Measurement Pressure, Min 0 Pa
Number of Batteries 3
Operating Temperature, Max 150 °F
Operating Temperature, Min 0 °F
Pressure Accuracy ± 10 %
Process Connection Size 1/4 in
Process Connection Type SAE Flare Fitting
Proof Pressure Up to 800 psi
Sample/Display Rate 0.5 sec
Sensor Type Thermistor
Supply Power Type Battery
Weight 6 oz
Width 2.75 in
See All Specs Show Less
0 to -30 inHg, Vacuum DVG-64A DVG-64A
Rated 4 out of 5 by from 1 Atm. to 1 mTorr with a mechanical gauge. This meter works great in conjunction with a mechanical gauge. It starts working at 20 Torr, just before a mechanical gauge reaches zero. The combination gives a full range down to 1 mTorr. I wish there was a way to stop the auto shut off as the gauge goes through a calibration startup when I turn it back on. The instructions say nothing about the autocalibration. Does calibration work at vacuum? Don't know. A DC input from a external supply would solve this. (Or a comment in the instructions but instructions are hard. We need Monk's brother Ambrose Monk)
Date published: 2020-03-23
  • y_2025, m_8, d_30, h_9
  • bvseo_bulk, prod_bvrr, vn_bulk_3.0.42
  • cp_1, bvpage1
  • co_hasreviews, tv_0, tr_1
  • loc_en_US, sid_DVG-64A, prod, sort_[SortEntry(order=HELPFULNESS, direction=DESCENDING), SortEntry(order=POSITIVE_FEEDBACK, direction=DESCENDING), SortEntry(order=SUBMISSION_TIME, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 6ms
  • REVIEWS, PRODUCT
en-US
en
Save Configuration