Need assistance? Call us: 1-800-663-4209
OM-EL-USB-1-LCD-Logger
Omega USB Temperature Data Logger with LCD Display
- -35 to 80°C (-31 to 176°F) Measurement Range
- USB Interface for Set-Up and Data Download
- User-Programmable Alarm Thresholds
- Status Indication Via Red, Green and Orange LEDs
OM-EL-USB-1-LCD-Logger
From
$
191.24
Product Overview
- Alarm Note High/low alarm thresholds for temperature selectable in software
- Battery life up to 1 yr
- Battery Size and Type 1/2 AA Li-ion
- Data Interface USB
- Display Type LCD
- Enclosure Ratings IP67
- Height 2.21 mm
- Input Type Internal
- LED Indicators Red, green and orange
- Length 119.5 mm
- Measurement Type Temperature
- Memory 16379
- Number of Measurements 1
- Operating Temperature, Max 80 °C
- Operating Temperature, Min -35 °C
- Repeatability ± 0.5°C
- Samples,Max 16000
- Sampling Rate 10 Hz (10 updates/second)
- Sampling Rate Note 10 sec to 12 hrs
- Series ID OM-EL-USB-1-LCD-Logger
- Software Compatibility Windows XP/VISTA/7 (32- and 64-bit)
- Supply Power Type Battery
- Temperature Accuracy ± 0.5 °C
- Temperature Accuracy Note Typical: ± 0.5ºC (± 1.0ºF) Max: ± 1.5ºC (± 3.0ºF)
- Temperature Measurement, Max 80 °C
- Temperature Measurement, Min -35 °C
- Temperature Resolution 0.5 °C
- Weight 43 g
- Width 25.3 mm
Supported Applications
The OM-EL-USB-1-LCD is a standalone USB temperature data logger with an integrated high-contrast LCD display, making it well suited to monitoring tasks that need an at-a-glance reading on the device as well as a recorded history for later download. It measures and stores up to 16,378 temperature readings across a working range of -35 to 80°C (-31 to 176°F), so it can cover deep-cold and moderately warm environments from a single unit.
Typical deployments include cold-chain and refrigerated storage, transport and warehouse monitoring, laboratory and cleanroom environments, and general HVAC and process temperature verification. User-programmable high and low alarm thresholds, together with red, green and orange LED status indicators, allow the logger to flag out-of-limit excursions locally without a PC connection.
With the supplied protective cap fitted, the housing is rated to IP67, allowing use in damp, dusty or wash-down settings. Note that these products are not designed for use in, and should not be used for, human applications.
Operating Conditions & Performance
The logger operates over a temperature range of -35 to 80°C (-31 to 176°F). At temperatures below -20°C (-4°F) the LCD display exhibits a slower response of around 10 sec. The device carries the CE mark and, with its protective cap fitted, an IP67 enclosure rating.
- Temperature accuracy: typical ±0.5°C (±1.0°F), maximum ±1.5°C (±3.0°F)
- Repeatability: ±0.5°C (±0.2°F)
- Resolution: 0.5°C (1.0°F)
- Memory: 16,378 temperature readings
- Logging interval: selectable from 10 sec to 12 hrs
Power comes from a replaceable 1/2 AA 3.6V lithium battery. Battery life is up to 1 year typical when run at 25°C with a 1 minute logging rate, though this depends on sample rate, ambient temperature and how the LCD display is used.
Configuration Options
This series is built around a single logger model, the OM-EL-USB-1-LCD, whose behaviour is configured through the supplied Windows software running on Windows XP/VISTA/7 (32- and 64-bit).
- Display modes: the LCD can be set off, on for 30 seconds after a button press, or permanently on.
- Logging rates: selectable at 10 s, 1 m, 5 m, 30 m, 1 hr, 6 hr and 12 hr.
- Start options: immediate, delayed and push-to-start logging.
- Data rollover: allows unlimited logging periods by overwriting the oldest data when the memory is full.
- Alarms and units: high and low alarm thresholds and °C or °F units are all set in software, with red, green and orange LEDs indicating status.
A companion replacement battery, the OM-EL-BATT, keeps the unit serviceable over its working life.
Key Product Differences
The OM-EL-USB-1-LCD lineup is straightforward:
- OM-EL-USB-1-LCD — temperature data logger with LCD display.
- OM-EL-BATT — replacement 3.6V lithium battery.
Each logger ships with a protective cap, a mounting bracket and the Windows configuration software on CD.
PDFs & Manuals
Software & Drivers
Show Ratings & Reviews
Is there a link to download the software?
Thank you for your question. Please see link and scroll down under product specs for software link.
https://www.omega.com/en-us/temperature-measurement/temperature-and-humidity-data-loggers/om-el-usb-1-series/p/OM-EL-USB-1
Date published: 2021-09-13
My om-EL-USB-TC -LCD is not working Did you give me a dead battery it only displays " --"
The logger must be programmed and started with the provided software. Until till it will only display the dashed lines.
Date published: 2019-07-30
Does the device need any annual or other calibration?
Hello
Regarding your request, we do recommend our devices (like this one as well) to be calibrated at least once a year.
Date published: 2021-10-15
I am trying to download data from one of the data loggers but i am unable to connect to the program and I am receiving a error "CRC is invalid" what doed this means?
Thank you for your question. This error indicates low battery. Please replace battery- OM-EL-BATT
Date published: 2025-04-04
What is the responce time?
14 minutes 32 seconds (time to 90%)
Thank you for your question
Date published: 2019-07-10