Skip to search Skip to main content Skip to navigation

Need assistance? Call us: 1-800-663-4209

We noticed you are in , please visit your local site by clicking the below link for regional offers and support.

XCIB

High Temperature Inconel Overbraided Silica Fiber Insulated Thermocouples

Models In Stock
  • High Temperature Ceramic Insulation with Inconel Overbraid
  • Flexible, Abrasion Resistant
  • Withstand Temperatures up to 1090 °C (2000 °F)
  • Unexcelled Oven and Furnace Profiling
XCIB
From $ 102.91
Models In Stock
High Temperature Inconel Overbraided Silica Fiber Insulated Thermocouples XCIB XCIB
loading all models

Can't find the right model?

Use our custom configurator to fine tune your product

Product Overview
  • Mounting Bolt-on
  • Process Temperature, Max 1038 °C
  • Process Temperature, Min 0 °C
  • Series ID XCIB
  • Sheath Material Inconel® 600
  • Washer Inner Diameter 0.1563 in
  • Washer Length 1 in
  • Washer Width 0.375 in
  • Wire Gauge 20 AWG
  • Wire Type Solid

Supported Applications

The XCIB Series comprises High-Temperature Inconel® Overbraided Silica Fiber Insulated Thermocouples. Each sensor is double-protected with an abrasion-resistant Inconel® overbraid over high-temperature fiber insulation, giving it the mechanical durability and thermal endurance needed for demanding heat-treatment environments.

These thermocouples are designed for oven and furnace profiling, where flexible, rugged sensors must survive repeated exposure to extreme heat. They are available in 5 convenient styles (1–5) for easy application and can be supplied in J, K, E and N thermocouple types to suit a wide range of process control and calibration tasks.

The sheath is Inconel 600®, and the probe diameter is 4.8 mm (3/16"). For assemblies that must ship with a calibration, use the prefix CAL- on the model number.

Operating Conditions & Performance

The XCIB thermocouples withstand 980°C (1800°F) continuous service and 1038°C short-term. The Inconel 600® overbraid and high-temperature fiber insulation are what allow this dual continuous/short-term rating in oven and furnace profiling.

A range of electrical terminations is offered, each with its own temperature limit. Ceramic beaded leads with compensated spade lugs and Nextel® insulated leads with compensated spade lugs are available for direct high-temperature runs, while connectorized options include the OMEGA® OSTW male standard thermocouple connector rated to 220°C (425°F) and the OMEGA® NHXH high-temperature ceramic male connector rated to 650°C (1200°F). Miniature options include the SMPW mini-male connector to 220°C (425°F) max and the SHX mini-male ceramic connector to 650°C (1200°F) max.

Configuration Options

Within the series, models differ along four independent axes: thermocouple type, probe style, termination type and length.

Thermocouple type sets the alloy pair and useful range: J is Iron-Constantan, K is CHROMEGA®-ALOMEGA®, E is CHROMEGA®-Constantan, and N is OMEGALLOY®-NICROSIL-NISIL.

Probe style selects one of the 5 convenient styles (1–5), while termination type chooses how the sensor is connected:

  • 1: Ceramic beaded leads with compensated spade lugs.
  • 2: Type OSTW male standard thermocouple connector to 220°C (425°F).
  • 3: Type NHXH male ceramic connector to 650°C (1200°F).
  • 4: Nextel® insulated leads with compensated spade lugs.
  • 5: Stripped leads.
  • 6: SMPW mini-male connector to 220°C (425°F) max.
  • 7: SHX mini-male ceramic connector to 650°C (1200°F) max.

Length is specified in feet, with 3' or 10' as the standard options. A grounded or ungrounded junction can be specified, and a Nextel-insulated version is available at no extra charge by changing the XCIB prefix to XSIB.

Key Product Differences

Model numbers follow the structure XCIB-<thermocouple_type>-<probe_style>-<termination_type>-<length_in_feet>. For example, XCIB-K-1-3-10 designates a Type K probe, probe style 1, termination Type 3, at a 10' length. Representative catalog numbers include XCIB-K-4-2-10, XCIB-J-4-3-3, XCIB-E-4-1-10 and XCIB-N-4-5-10, along with the ungrounded XCIB-K-4-2-10-UNGROUNDED.

The XSIB equivalents, such as XSIB-K-4-1-10 and XSIB-J-4-2-3, substitute Nextel® insulation for the same configurations. Custom lengths and sizes are available, and connector plug options can be appended to the model number: add 'M8-S-M' or 'M12-S-M' for a male straight M8 or M12 plug, or 'M8-R-M' for a male right-angled M8 plug.

Rated 5 out of 5 by from Everything I need in a Customer Experience Prompt deliver, perfect packaging and great service
Date published: 2025-08-13
Rated 2 out of 5 by from Out-of-box accuracy issue This item seems well built, and came in time - but was off from room temperature by -25F right out of the box. I'm working with Omega to see if they'll replace it; I understand the need to learn and make corrections - but it's a little frustrating to wait for that process.
Date published: 2024-02-16
Rated 5 out of 5 by from Perfect Item Just what we needed for our application. Fast delivery!
Date published: 2023-07-26
  • y_2026, m_8, d_30, h_14
  • bvseo_bulk, prod_bvrr, vn_bulk_3.0.48
  • cp_1, bvpage1
  • co_hasreviews, tv_0, tr_3
  • loc_en_US, sid_XCIB-E-4-1-10, prod, sort_[SortEntry(order=HELPFULNESS, direction=DESCENDING), SortEntry(order=POSITIVE_FEEDBACK, direction=DESCENDING), SortEntry(order=SUBMISSION_TIME, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 4ms
  • REVIEWS, PRODUCT

Can these be made to meet special limits of error?

Asked by: Andrewww
Hi, This series is available only with standard limits of error.
Answered by: Application Eng JA
Date published: 2026-06-26

What is the response time of the sensor?

Asked by: Ricardo Santillan
Thank you for your inquiry. Please see our technical reference page for more information https://www.omega.com/en-us/technical-reference
Answered by: Application Engineer CB
Date published: 2022-03-18

Do these thermocouples come with a calibration certificate?

Asked by: Teresa12
Thank you for your inquiry. These sensors only come with a Certificate of Conformance (CAL-1). CAL-1 documents that the products you buy comply with Omega's specifications. If you need a NIST certificate of calibration with this product, please send a quote request to sales@omega.com
Answered by: Application Engineer MC
Date published: 2023-05-02

Do the thermocouples have a required installed orientation?

Asked by: enycel
Thank you for your inquiry. No required install orientation. Thermocouples are very flexible in-terms of how they are installed. You try to put them in the best position to get your temperature reading.
Answered by: Engineer CB
Date published: 2020-05-28

Can these be ordered in a custom length? Can the probe end be welded to a surface?

Asked by: JustinM
Yes the probes can be ordered in custom lengths and if you use the configure now feature on the page you can select that on your own. The style 4 connection should allow a weld instead of using the screw hole but heat should be kept to a minimum.
Answered by: ApplicationEngineerANB
Date published: 2024-02-05

Do these thermo couples have UL certification?

Asked by: Mario69
Thank you for your inquiry. Unfortunately, no, they do not have UL certification.
Answered by: NorwalkAE_SP
Date published: 2024-08-28

What type of junction used in XCIB style 2? Grounded or ungrounded? What is is the dimension (diameter) of its sensing element?

Asked by: Anonymous
Style 2 of the XCIB series probes is ungrounded and the bead is approximately 0.080&quot; in diameter
Answered by: ApplicationEngineerANB
Date published: 2020-11-11
  • y_2026, m_8, d_30, h_14CST
  • bvseo_bulk, prod_bvqa, vn_bulk_3.0.48
  • cp_1, bvpage1
  • co_hasquestionsanswers, tq_7
  • loc_en_US, sid_XCIB-E-4-1-10, prod, sort_[SortEntry(order=NUM_ANSWERS, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getContent, 10ms
  • QUESTIONS, PRODUCT
en-US
en
Save Configuration