Skip to search Skip to main content Skip to navigation

Need assistance? Call us: 1-800-663-4209

We noticed you are in , please visit your local site by clicking the below link for regional offers and support.

BB702-series

Ambient 10 °C to 215 °C Temperature Blackbody Calibrator

  • Model BB702 Calibrates from Ambient +10 to 215°C (Ambient +20 to 420°F)
  • Portable Rugged Design
  • Calibrates Infrared Pyrometers Quickly and Accurately
  • Built in Digital PID Autotune Temperature Controller with Temperature Readout
BB702-series
From $ 4,515.45
Ambient 10 °C to 215 °C Temperature Blackbody Calibrator BB702-series BB702-series
loading all models
Product Overview
  • Accuracy ± 0.5 °C
  • Accuracy Detail ±0.5°C, ±0.25% rdg (±0.9°F, ±0.25% rdg)
  • Additional Features 1 Portable
  • Calibration Reference Sensor Platinum RTD, class A, alpha = 0.00385
  • Cavity Emissivity 0.95
  • Control Mode PID Control
  • Data Interface RS232
  • Display Type LED
  • Frequency 50/60 Hz
  • Height 152 mm
  • Installation Category II
  • Internal Sensing Element Platinum RTD, class A, alpha = 0.00385
  • Length 280 mm
  • Operating Temperature, Max 45 °C
  • Operating Temperature, Min 5 °C
  • Power Consumption 75 W
  • Process Temperature, Max 215 °C
  • Process Temperature, Min 10 °C
  • Series ID BB702-series
  • Stability ±0.1°C
  • Target Size 63.5
  • Warm Up Time 12 min
  • Warm-Up Time Note 12 min from ambient to 204°C (400°F)
  • Width 305 mm

Supported Applications

The Omega BB702 series is a portable, rugged bench-top blackbody calibration source designed to test and calibrate non-contact infrared temperature instruments. The unit features a built-in precision PID digital controller with temperature readout.

Operating Conditions & Performance

The target plate operates within a range of Ambient 10 to 215°C (ambient 20 to 420°F). Accuracy is specified as ±0.5°C, ±0.25% rdg (±0.9°F, ±0.25% rdg), with stability at ±0.1°C (±0.2°F). The target plate measures 63.5 mm (2.5") in diameter and has an emissivity of 0.95.

Internal control and reference sensors are Platinum RTD, class A, alpha = 0.00385. Warm-up time is 12 min from ambient to 204°C (400°F). The unit operates in ambient environmental conditions of 5 to 45°C (41 to 113°F) and humidity levels of 0 to 90% RH, non-condensing.

Configuration Options

The series is offered with two power supply configurations:

  • BB702: 115 Vac, 50/60 Hz, 75 W (Domestic Models)
  • BB702-230VAC: 230 Vac, 50/60 Hz, 75 W (European Models)

The unit includes an RS232 computer interface and a built-in RTD reference probe output. It features a tilt stand for benchtop operation.

Key Product Differences

Differences between models are defined by the voltage rating in the model number suffix:

  • BB702: 115V configuration, range: ambient 10 to 215°C (ambient 20 to 420°F).
  • BB702-230VAC: 230V configuration, range: ambient 10 to 215°C (ambient 20 to 420°F).

The series includes a NIST-traceable calibration certificate with three data points and carries an Installation Category II rating. Dimensions are 152 H x 305 W x 280 mm L (6 x 12 x 11") and weight is 7.2 kg (16 lb).

  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 8ms
  • REVIEWS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;

What is the uncertainty of the target plate's 0.95 emissivity? Or is that uncertainty already incorporated into the plate's temperature accuracy?

Asked by: Hartman Walushka
Hi Hartman, Thank you for the inquiry. Accuracy of the emissivity is included in the calibrator accuracy. Best regards OMEGA
Answered by: NorwalkAE_RS
Date published: 2023-09-07
  • y_2026, m_8, d_23, h_9CST