Skip to search Skip to main content Skip to navigation

Netherlands 07077 03815 | Intl +31 (0) 7077 03815 | Welcome to our new website, found an issue or bug? Please report it here

We noticed you are in , please visit your local site by clicking the below link for regional offers and support.

OM-PR-Series

Submersible Pressure/Temperature Data Loggers

Models In Stock
  • Temperature Range: -20 to 85°C (-4 to 185°F)
  • Pressure Ranges: 35, 150, 350, 550, 2000, 5800 psia
  • Vacuum Ranges: 760 and 380 Torr
  • 150% Over Range Protection
OM-PR-Series
From  625,77
Models In Stock
Submersible Pressure/Temperature Data Loggers OM-PR-Series OM-PR-Series
loading all models
Product Overview
  • Approvals CE
  • Battery life up to 5 yr
  • Battery Size and Type 3.6V Li-ion
  • Calibration Accuracy Temperature : ±2 °C, Pressure : ±1%
  • Control Output 2 Programmable Alarms
  • Data Interface USB
  • Diameter 50 mm
  • Height 56 mm
  • Materials 316L Stainless Steel
  • Measurement Type Pressure, Temperature
  • Measuring Temperature Range, Max 85 °C
  • Measuring Temperature Range, Min -20 °C
  • Memory 32,000 or 64,000 (depends on setup)
  • Millivolt Range N/A
  • Number of Measurements 2
  • Operating Temperature, Max 85 °C
  • Operating Temperature, Min -20 °C
  • Measurement Pressure, Min 0 psi
  • Pressure Repeatability ±0.05%
  • Process Connection Size 1/4 in
  • Process Connection Type NPT Male
  • Proof Pressure 150% Full Scale
  • Record Trigger Note Record Trigger: Two independent triggers Multiple Trigger Modes: Instantaneous, button control, on alarm, time and date (Start and Stop), day of week Record Mode: Fill to end of memory or cyclic, number of samples and time duration
  • Sampling Rate 0.5 Hz (2 seconds between updates)
  • Sampling Rate Note User configured 1 every 2 seconds to 1 every 24 hours
  • Series ID OM-PR-Series
  • Software Windows 2000/XP/Vista/7 and 8 (32-bit and 64-bit)
  • Supply Power Type Battery
  • Weight 900 g

Supported Applications

The OM-PR Series is a family of battery-powered, stand-alone, water-tight pressure and temperature data loggers. Each unit is a compact, self-contained recorder that captures up to 64,000 samples of pressure and/or temperature, making it well suited to unattended field studies where a cabled instrument is impractical.

Because the logger is housed in a rugged submersible 316L stainless steel enclosure rated IP67, it can be deployed directly in demanding, wet, and submerged environments. A ¼ NPT male process connection lets it thread into pipes, tanks, and test fittings, and an optional garden hose adaptor supports water-system work. Vacuum-range models extend the same platform to sub-atmospheric measurement.

Each logger can be set to record both pressure and temperature or a single parameter to maximize storage, and is configured with the included Windows software over a direct USB connection. Two independent triggers and multiple trigger modes — instantaneous, button control, on alarm, time and date, and day of week — let the device start and stop logging automatically around an event of interest.

Operating Conditions & Performance

The OM-PR Series measures temperature over -20 to 85°C (-4 to 185°F) with a resolution of 0.1°C (0.1°F). Temperature accuracy is ±1°C, 0 to 50°C (32 to 122°F) and ±2°C, -20 to 85°C (-4 to 185°F).

On the pressure side, accuracy is ±1% with a resolution of 0.01 psia and repeatability of ±0.05%. Each sensor includes 150% Over Range Protection, rated to 1.5 times maximum, to guard against transient overpressure.

Recording behavior is highly configurable:

  • Sample rate: user-configured from 1 every 2 seconds to 1 every 24 hours.
  • Number of samples: 32,000 or 64,000, depending on setup.
  • Record mode: fill to end of memory or cyclic, by number of samples and time duration.
  • Alarms: 2 user programmable alarms, high or low.

Power is supplied by an included 1/2AA, 3.6V, 1.2Ah lithium battery, and the replaceable internal battery offers up to a 5-year life, rated up to 5 years at a 1 minute sample rate. Communication is by direct USB connection, and the included software runs on Windows 2000/XP/Vista/7 and 8 (32-bit and 64-bit). The enclosure measures 5.6 H x 5 cm dia (2.19 x 1.95") and weighs 0.9 kg (2.0 lb). The series is CE marked, an optional NIST calibration certificate is available, and OMEGACARE extends the standard 1-year warranty to a total of 2 years.

Configuration Options

All OM-PR loggers share the same submersible 316L stainless steel body, ¼ NPT male port, temperature performance, USB configuration, and recording feature set. The models differ chiefly in their calibrated pressure or vacuum range and in whether they ship with a NIST calibration certificate, following the model number structure OM-PR — Pressure Range — NIST Calibration.

Positive-pressure models are offered across an ascending ladder of full-scale ranges:

  • OM-PR35 — 0 to 35 psia.
  • OM-PR150 — 0 to 150 psia.
  • OM-PR350 — 0 to 350 psia.
  • OM-PR550 — 0 to 550 psia.
  • OM-PR2000 — 0 to 2000 psia.
  • OM-PR5800 — 0 to 5800 psia.

Vacuum-range models cover sub-atmospheric measurement: OM-PR760V for the 760 Torr vacuum range and OM-PR380V for the 380 Torr vacuum range. Every range is available with an optional NIST calibration certificate ordered as the matching -NIST model, which does not change the measurement capability but adds documented, traceable calibration.

Key Product Differences

The ordering lineup pairs each pressure or vacuum range with an optional NIST-certified counterpart:

  • OM-PR35 / OM-PR35-NIST — pressure/temperature data logger, 35 psia range.
  • OM-PR150 / OM-PR150-NIST — 150 psia range.
  • OM-PR350 / OM-PR350-NIST — 350 psia range.
  • OM-PR550 / OM-PR550-NIST — 550 psia range.
  • OM-PR2000 / OM-PR2000-NIST — 2000 psia range.
  • OM-PR5800 / OM-PR5800-NIST — 5800 psia range.
  • OM-PR760V — 760 Torr vacuum range.
  • OM-PR380V / OM-PR380V-NIST — 380 Torr vacuum range.

Supporting replacement parts and accessories are available for every model:

  • OM-PR-BATT — replacement 1/2AA, 3.6V lithium battery.
  • OM-PR-ADAPTOR — garden hose adaptor (brass coupling).
  • OM-PR-SEAL — replacement O-ring seal (pack of 3).
  • OM-PR-COVER — locking anti-theft cover.
  • OM-PR-USB-CABLE — replacement 0.9 m (3') USB cable.
en-NL
en_NL
Save Configuration