United Kingdom 0800 488 488 | Intl +44 (0) 161 777 6611 | Welcome to our new website, found an issue or bug? Please report it here
OM-CP-PR140-Series
Submersible Pressure Data Logger for High Temperature
- Flush Mount (OM-CP-PR140)
- or 1/8 NPT Pressure Port
- (OM-CP-PR140-NPT)
- Submersible (IP68)
OM-CP-PR140-Series
From
£
1,207.08
Product Overview
- Accuracy Detail Calibration Accuracy : Pressure: ±0.435 psi @ 20 to 140ºC; Time Accuracy : ±1 minute/month at 20ºC; Pressure Repeatability : ±0.5% FSR; ±0.2% typical
- Approvals CE
- Battery life up to 2 yr
- Battery Size and Type 3.6V Li-ion
- Baud Rate 125000
- Calibration Digital Calibration through Software with Montly Calibration interval
- Calibration Accuracy Pressure: ±0.435 psi
- Data Interface USB
- Diameter 25.4 mm
- Frequency Response Pressure Response Time: 0.1 ms
- Humidity Range 0 to 100% RH
- Length 50.8 mm
- Materials Stainless Steel
- Measurement Type Pressure
- Memory 32,700 Readings
- Memory Note 32,700 readings; software configurable memory wrap
- Millivolt Range N/A
- Mounting Flush Mount
- Number of Measurements 1
- Operating Temperature, Max 140 °C
- Operating Temperature, Min -20 °C
- Measurement Pressure, Max 72.5 psi
- Measurement Pressure, Min 0 psi
- Pressure Repeatability ±0.5%
- Pressure Sensor Type Silicon Strain Gage
- Measurement Pressure Type Absolute
- Real Time Recording The device may be used with PC to monitor and record data in real-time
- Resolution Pressure : 0.0015 psi
- Sampling Rate 1 Hz (1 update/second)
- Sampling Rate Note 1 second to 1 every 24 hours
- Series ID OM-CP-PR140-Series
- Software Windows XP SP3 / Vista / 7 and 8 (32 and 64 - bit)
- Start Modes Delayed Start, Immediate Start
- Stop Modes Adjustable Delay Stop
- Supply Power Type Battery
- Time Accuracy ±1 minute/month
- Weight 85 g
Supported Applications
The Omega OM-CP-PR140 is a high-temperature pressure data logger designed for use in autoclave validation and mapping. As part of the NOMAD® family, this rugged, submersible device operates up to 140°C (284°F) within environments requiring IP68 protection.
Operating Conditions & Performance
The series utilizes a semiconductor strain gauge pressure sensor with a measurement range of 0 to 5 bar (0 to 72.5 psia). Calibrated accuracy is ±0.03 Bar (±0.435 psi) within the process temperature span of 20 to 140°C/68 to 284°F, with a resolution of 0.1 mBar (0.0015 psia). Repeatability is specified as ±0.5% FSR; ±0.2% typical.
Performance characteristics include a pressure response time of 0.1 ms (10 to 90% FSR) and a reading rate configurable from 1 second to 1 every 24 hours. The device stores up to 32,700 readings in non-volatile memory with software-configurable wrap-around capability. Time accuracy is ±1 minute/month at 20°C.
Operating environment limits are -20 to 140°C (-4 to 284°F) and 0 to 100% RH, non-condensing; the logger may be used above 60°C (140°F) for up to 8 hrs per 24 hr period. Power is supplied by a user-replaceable 3.6V lithium battery included with typical life of 2 years.
Configuration Options
The series offers configurable process connections and mounting options:
- Housing & Mounting: Flush mount (OM-CP-PR140) or ⅛ NPT pressure port (OM-CP-PR140-NPT).
- Dimensions: OM-CP-PR140 measures 50.8 L x 25.4 mm dia. (2.0 x 1.0"); OM-CP-PR140-NPT measures 58.2 L x 25.4 mm dia. (2.3 x 1.0").
- Materials: Stainless steel enclosure and wetted materials.
- Data Interface: USB connection via the required OM-CP-IFC400 interface cable at 125,000 baud; supports real-time recording with PC software compatible with XP SP3/Vista/7 and 8 (32- and 64-bit).
- Data Formats: psia, bar, mbar, atm, Torr, Pa, kPa, MPa, mmHg, inHg, inH2O, mH2O, foot, meter.
Optional accessories and configurations include the OM-CP-PR140-LUERFITTING (⅛ NPT to female luer lock adaptor), OM-CP-BAT110 replacement battery, NIST calibration certificate, 5 YEAR WARRANTY, and FDA 21 CFR Part 11 compliant IQ/OQ/PQ secure software validation workbook (OM-CP-SVP-SYSTEM). The OM-CP-IFC406 multiplexer interface allows up to 6 data loggers to connect to a single USB port.
Key Product Differences
The series distinguishes between two primary physical configurations based on the process connection:
- OM-CP-PR140: High temperature pressure data logger with flush mount design (50.8 L mm).
- OM-CP-PR140-NPT: High temperature pressure data logger featuring a ⅛ NPT pressure port (58.2 L mm).
Both configurations are available with or without an optional NIST calibration certificate, indicated by the suffix "-CERT" in model numbers such as OM-CP-PR140-CERT and OM-CP-PR140-NPT-CERT.
PDFs & Manuals
Software & Drivers
Accessories
accesorios
Estación de acoplamiento multiplexor de 6 puertos con cable USB
894,51 € / unidad
0 In Stock
Software seguro para uso con registradores de datos de la serie OM-CP
2.526,92 € / unidad
0 In Stock
consumables
Batería de litio de reemplazo de alta temperatura de 3,6 V para registrador de datos de presión de alta temperatura
84,80 € / unidad
8 In Stock