- Accuracy Detail pH: ±0.01 pH,ORP/ISE:±0.1 mV,Temperature Calibrated Accuracy: ±0.15°C ,@ 25°C ambient (±0.015 Ω); pH EMI Susceptibility: <0.2 pH @ 3V/M, 80 MHz to 1 GHz; ORP/ISE Susceptibility: <10 mV @ 3V/M, 80 MHz to 1 GHz
- Battery life 1 yr, typical with reading interval of 1 min
- Battery Size and Type Alkaline 9V
- Baud Rate 115200
- Calibration Digital calibration through software
- Data Interface USB
- Display Note LCD Dimensions: 1.375" x 2.5" Backlight: Configurable w/auto shut-off and contrast adjustment
- Display Type LCD
- Electrical Connection Note Temperature: Removable screw terminal, pH: Female BNC
- Height 1.25 in
- Impedance Input 1012 Ω
- Indicator Type Power, Status, Memory
- Length 3.3 in
- Materials Aluminum
- Measurement Type Temperature, pH/ORP/ISE
- Memory 131,071 readings
- Memory Note 131,071 readings per channel
- Number of Measurements 2
- Operating Temperature, Max 50 °C
- Operating Temperature, Min -5 °C
- PH Measurement Range -2 to 16 pH
- Real Time Recording The device may be used with PC to monitor and record data in Real-time
- Relative Humidity 0 to 95% RH non-condensing
- Resolution Temperature: 0.01°C, Measurement: 0.01 pH
- Sampling Rate 2 Hz (2 updates/second)
- Sampling Rate Note 1 Reading every 2 Seconds to 1 Every 24 Hours
- Sensor Type pH, RTD
- Series ID OM-CP-PHTEMP2000-Logger
- Software XP SP3/Vista/7 and 8
- Start Modes Delayed Start, Immediate Start
- Supply Power Type Battery
- Temperature Measurement Range -40 to 110 °C
- Time Accuracy ±1 minute/month (at 25°C; RS232 cable not in use)
- USB Cable Length 6 ft
- Weight 440 g
- Width 4.8 in
- Wire Code Option 2 -, 3 - , 4- Wire
Supported Applications
The Omega OM-CP-pHTemp2000 is a battery-powered, stand-alone pH and temperature data logger with an LCD display. It provides access to current readings as well as minimum, maximum, and average statistics for process monitoring. The device connects via a female BNC connector to OMEGA PHE-4200 series pH probes and other common pH, ORP (Oxidation Reduction Potential), and ISE (Ion Selective Electrode) sensors. Temperature compensation is automatic using an RTD probe connected through removable screw terminals.
Applications include environments requiring FDA 21 CFR part 11 compliant software validation via the OM-CP-SVP-SYSTEM package. The logger supports real-time recording and programmable start times up to six months in advance, storing data in non-volatile solid-state memory for retrieval via PC using included Windows software.
Operating Conditions & Performance
The series measures pH/ORP/ISE within a range of -2.00 to 16.00 pH with a resolution of 0.01 pH (0.001 pH) and calibrated accuracy of ±0.01 pH for pH, or ±0.1 mV for ORP/ISE. Temperature measurement utilizes a 100Ω Pt RTD input covering -40 to 110°C (-72 to 230°F), with a resolution of 0.01°C (0.001 Ω) and calibrated accuracy of ±0.15°C @ 25°C ambient (±0.015 Ω).
Operating conditions are specified as -5 to +50°C, 0 to 95% RH (non-condensing). The device features an input impedance of 10¹² Ω typical for pH/ORP/ISE inputs and exhibits EMI susceptibility less than 0.2 pH @ 3V/M, 80 MHz to 1 GHz.
Key Performance Specifications
- pH Range: -2.00 to +16.00 pH
- Temperature Range: -40 to 110°C (-72 to 230°F)
- Memory Capacity: 131,071/channel
- Sample Rate: Configurable from 2 seconds to 24 hours
- Baud Rate: 115,200 via PC serial or USB interface
- Data Format: Date and time stamped °C, °F, R, Ω, pH, V, mV, engineering units specified through software
Configuration Options
The OM-CP-pHTemp2000 series is configurable with various accessories to support specific installation and compliance needs. The primary unit features a black anodized aluminum enclosure measuring 4.8" x 3.3" x 1.25" (122mm x 84mm x 32mm) and weighs 440 g (16 oz).
Available Options & Accessories
- Interface: OM-CP-IFC200 Windows software and 1.8 m (6') USB interface cable.
- Validation Package: OM-CP-SVP-SYSTEM FDA 21 CFR part 11 compliant IQ/OQ/PQ secure software validation workbook and software package (unlimited users, license per computer).
- Power Supply: Operates on a user-replaceable 9V lithium or alkaline battery. Optional AC adaptors include OM-CP-110-PWR-2000 (110 Vac) and OM-CP-220-PWR-2000 (220 Vac).
- Replacement Parts: OM-CP-BAT103 Replacement 9V lithium battery.
- Sensors: Compatible with PHE-4201 general purpose pH electrode and PR-11-2-100-1/4-6-E general purpose 100Ω RTD probe (¼" diameter, 6" sheath).
Sensor Connection Requirements
- pH/ORP/ISE: Requires a female BNC connector. The pH electrode should have an output impedance less than 300 megaohms at the desired temperature.
- Temperature: Supports 2-, 3-, or 4-wire interface for 100Ω platinum RTD probes via removable screw terminals. Most 100 Ohm, 3 wire RTD probes will work if resistance between two same colored wires is less than 1 Ohm.
Key Product Differences
The series includes specific model variations to address calibration documentation and software compliance needs:
- OM-CP-PHTEMP2000: Standard pH and temperature data logger with LCD display.
- OM-CP-PHTEMP2000-CERT: Includes a NIST calibration certificate (optional).
The device is CE approved. A 5-year warranty applies to the unit, which is made in the USA. The backlight uses significant battery life and can be configured to auto shut-off when not in use; typical battery life is 1 year with display off or 30 days with continuous LCD and no backlight.