CYC325-Series
Cryogenic Dual Channel Digital Benchtop Temperature Controller
- Operates Down to 1.2 K with Appropriate Sensors
- 2 Sensor Inputs
- Supports Diode, RTD, and Thermocouple Sensors
- Sensor Excitation (Current Reversal Eliminates Thermal EMF Errors for Resistance Sensors)
CYC325-Series
From
£
6,964.18
Product Overview
- A/D Conversion 24-bit
- Accuracy Selectable
-
Accuracy Detail
PTC RTD (0 to 500 Ω : Range) : ± 0.004 Ω, ± 0.01 % of rdg
PTC RTD (0 to 5000 Ω : Range) : ± 0.004 Ω, ± 0.02 % of rdg
NTC RTD (0 to 7500 Ω : Range) : ± 0.1 Ω, ± 0.04 % of rdg
Thermocouple (± 25 mV) : ± 1 µV,± 0.05 % of rdg
Thermocouple (± 50 mV) : ± 1 µV,± 0.05 % of rdg
- Approvals CE Mark
- Baud Rate 9600 to 57600
-
Control Characteristics
PID Control Settings:
- Proportional (Gain): 0 to 1000 with 0.1 setting resolution
- Integral (Reset): 1 to 1000 (1000/s) with 0.1 setting resolution
- Derivative (Rate): 1 to 200% with 1% resolution
Manual Output: 0 to 100% with 0.01% setting resolution
Zone Control: 10 temperature zones with P, I, D, manual heater out, and heater range
Setpoint Ramping: 0.1 K/min to 100 K/min
Safety Limits: Curve temperature, power up heater off, short circuit protection
- Control Output Heater
- Control Type Manual Heater Output, Open Loop, PID Control
- Data Interface RS232
- Digital Connector Type 9 pin Sub-D
- Display Annunciators Autotune, Control input, Remote
- Display Measuring Units °C, K, mV, V
- Display Size 5.5 mm Digits
- Display Type LCD
-
Display Type Note
Display: 2-line, 20-character, liquid crystal display with 5.5 mm (0.216”) character height
Display Update Rate: 2 rdg/s
Other Displays: Setpoint, heater range, and heater output; user selected
- Electrical Connection Ceramic Isothermal Block
- Filter Averages 2 to 64 input readings
- Form factor Benchtop
- Front Panel Display Readings 1 to 4
- Heater Output Display Percentage of full power or current
- Heater Output Resolution 0.01
- Height 89 mm
- Input Type Thermocouple, RTD, Thermistor
- Input Type Note Temperature, sensor units Input Configuration: Each input is factory configured for either diode/RTD or thermocouple
- Interface Features IEEE-488 Interface Features: SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT0, C0, E1
- Isolation Sensor inputs optically isolated from other circuits but not each other
- Keypad 20-key membrane, numeric and specific functions
- Length 368 mm
- Loops Control Loops : 2
- Measurement Type Temperature
- Number of Control Outputs 2
- Number of Inputs 2
- Operating Temperature, Max 35 °C
- Operating Temperature, Min 15 °C
- Output Detail 2 Heater output
- Output Signal None
- Power Consumption 6 W
- Resolution Sensor dependent, refer to input specifications table
- RTD Type 4-lead differential With 6-Pin DIN Connector
- Sampling Rate 10 Hz (10 updates/second)
- Sampling Rate Note To 10 rdg/s on each input
-
Sensor Input Configuration
Measurement Type: 4-lead differential
Excitation: Constant current with current reversal for RTDs Supported
Sensors: Diodes, silicon, GaAlAs; RTDs, 100 Ω Platinum, 1000 Ω Platinum, germanium, carbon-glass, CernoxTM, and RoxTM Standard Curves: CY7 and CY670, PT-100, PT-1000, RX-102A, RX-202A
Input Connector: 6-pin DIN
- Sensor Units Display Resolution Sensor dependent up to 5 digits
- Series ID CYC325-Series
- Setpoint Setting Resolution Same as display resolution (actual resolution is sensor dependent)
- Settable Parameters Curve Entry, Keypad Lock-out
- Software Support LabVIEWTM driver; consult factory for availability
- Supply Power Note Standard 120 Vac, optional 240 Vac, 6%, -10%, 50 or 60 Hz, 85 VA
- Supply Power Type AC
- Supply Voltage 100 to 240 Vac
- Temp Display Resolution 0.001° from 0 to 99.999°, 0.01° from 100 to 999.99°, 0.1° above 1000°
- Tuning Autotune, PID, PID zones
- Update Rate, Max 10 rdg/s on each input (except 5 rdg/s on input A when configured as thermocouple)
- Weight 4 kg
- Width 216 mm
Supported Applications
The Omega CYC325 Series, including models CYC325, CYC325-T1, and CYC325-T2, is a precision cryogenic temperature controller designed for controlling and measuring temperatures in laboratory and industrial environments. The instrument features a front panel with a digital display and keypad for setup and operation.
Per manufacturer warnings, the equipment is not intended to be used as a 'Basic Component' under 10 CFR 21 (NRC) in nuclear installations or activities, nor is it intended for medical applications or use on humans.
Operating Conditions & Performance
The series operates within an ambient temperature range of 0 °C to 50 °C and supports a process measurement range from 1.5 K to 400 K, with optional sensors available for extended ranges. The instrument provides a resolution of 0.001 K.
- Accuracy: ±0.05% of reading or ±0.05 K, whichever is greater; improved to ±0.01% of reading or ±0.01 K with CalCurve™.
- Response Time: Typical 10 seconds (depends on system thermal mass).
- Humidity Rating: 0% to 90% RH, non-condensing.
- Storage Temperature: -40 °C to 70 °C.
The unit is housed in a metal chassis with a plastic front panel and carries an enclosure rating of NEMA 1. It complies with the European EMC Directive 89/336/EEC as a Class A product (CE Mark).
Configuration Options
The CYC325 Series is configurable for various control and monitoring needs through selectable outputs, excitation currents, and optional accessories.
- Output Signals: 0-20 mA (load resistance 0 to 500 Ω), 4-20 mA, 0-10 V, 0-5 V, or Relay Output.
- Excitation Current: Selectable between 10 µA or 1 mA.
- Power Supply: 100-240 VAC, 50/60 Hz with rear panel connectors for sensor inputs, heater outputs, and power via screw terminals.
- Dimensions (W x H x D): 193 mm x 127 mm x 305 mm; weight is approximately 4.5 kg.
- Options & Accessories: Optional sensors for extended range, calibration services, and software for remote control and data logging are available.
Key Product Differences
The series includes specific model designations: CYC325, CYC325-T1, and CYC325-T2. The manual notes that the full list of accessories is not explicitly detailed beyond general mentions.
Users should note EMC considerations for Class A products in domestic environments; radio interference may occur requiring adequate measures by the user. Additionally, improper cabling installation may defeat EMC protection, and warranty coverage excludes damage from tampering, excessive corrosion, current, heat, moisture, vibration, or misuse outside of Omega's control.