Product Overview
- Accuracy ± 0.15 % of reading, ± 1 °C
- Alarm modes High/Low, Latch/Unlatch, Manual reset
- Cutout Dimensions 92 × 92 mm
- Data Interface RS232
- Data Interface Note RS232 Transmit Only, 2 wire configuration, Tansmit & Ground connections, One way communications only
- Display Size 1.4 cm
- Display Type 4 Digit LCD
- Display Type Note LCD, backlight, PV, SV and status indicators, 4-digit, PV 1.4 cm (0.56")
- Form factor 1/4 DIN
- Height 89 mm
- Input Type Thermocouple
- Length 95 mm
- Material Construction Option N/A
- Measuring Range Type J: -50 to 1000°C (-58 to 1832°F) Type K: -50 to 1370°C (-58 to 2498°F) Type T: -120 to 400°C (-184 to 752°F) Type E: -50 to 759°C (-58 to 1382°F)
- Memory Backup Non-Volatile
- Number of Relays 1
- Operating Temperature, Max 50 °C
- Operating Temperature, Min 0 °C
- Power Consumption 1.8 W
- Resolution 0.1 (< 1000), 1 (> 1000)
- Retransmission/Analog Output SPDT Relay
- Retransmission/Analog Output Note 5 A SPDT relay @ 250 Vac resistive
- Series ID DP1001AM-Series
- Storage Temperature, Max 50 °C
- Storage Temperature, Min 0 °C
- Supply Power Note 9 Vdc, 200 mA (115 Vac adaptor included)
- Supply Voltage 9 Vdc
- Temperature Sensor Type N/A
- Weight 374 g
- Width 83 mm
Le scanner à 10 canaux à microprocesseur DP1001AM accepte le même type d'entrée thermocouple pour chaque entrée. Chaque canal peut être balayé manuellement ou automatiquement et les zones actives sont affichées. L'instrument est programmable pour les types d'entrée thermocouple J, K, T ou E, les degrés °C ou °F, et la vitesse de balayage entre 2 et 60 secondes. Chaque canal dispose de points de consigne d'alarme individuels, indépendants et programmables avec indicateur d'alarme. La mémoire et les points de consigne sont conservés lorsque l'appareil est hors tension.
Les communications RS232 (configuration à 2 fils, transmission/masse) peuvent surveiller jusqu'à dix canaux et un logiciel est fourni avec chaque appareil. Un relais à verrouillage de sortie unique de forme C est fourni pour activer une condition d'alarme sur n'importe quelle zone. L'instrument affiche la condition d'alarme pour chaque canal sur l'écran avant. Les alarmes sont à verrouillage et sont réinitialisées manuellement à partir du panneau avant.
Les communications RS232 (configuration à 2 fils, transmission/masse) peuvent surveiller jusqu'à dix canaux et un logiciel est fourni avec chaque appareil. Un relais à verrouillage de sortie unique de forme C est fourni pour activer une condition d'alarme sur n'importe quelle zone. L'instrument affiche la condition d'alarme pour chaque canal sur l'écran avant. Les alarmes sont à verrouillage et sont réinitialisées manuellement à partir du panneau avant.
PDFs & Manuals
Software & Drivers
Show Ratings & Reviews
What is the recommended calibration/maintenance frequency for Item Item# DP1001AM?
Thank you for connecting with OMEGA.
Recommended maintenance should be performed annually on a continuous basis. This will ensure maximum shelf life of the unit
Date published: 2022-01-26
Is there a way to change the alarm relay from latched to momentary?
Hi MH3050
Thank you for the request. DP1001AM has latching alarms only.
Best regards,
OMEGA
Date published: 2024-03-24
Can a delay be put on an output? I need it to turn the output on for maintained temperature not spikes.
Hi Jimmie
Thank you for the inquiry. This unit does not have delay function.
Best regards
OMEGA
Date published: 2022-02-03
Can I calibrate the DP1001AM?
Thank you for your question.
You cannot calibrate this unit in the field. Our 10 channel temperature scanner has an automatic calibration system internally, this is done by the microprocessor master control board. If the unit is not reading accurately then you need to contact our customer care team.
Date published: 2019-06-26
Alarm Type: Hi/Lo, what does this mean? High and Low window, band? Manual doesn't explain the setup? Thank you
Thank you for your inquiry! This simply means that each channel has the option for a high or low alarm which will trigger when that high or low value is exceeded by the process value.
Date published: 2021-04-01
I noticed a CJC (assuming that stands for Cold Junction Compensation) terminal on the back of the unit in the pinout diagram in the manual, but your manual doesn't mention it at all. could you expand on its setup/use(s).
Thank you for the inquiry. The CJC temperature sensor is located in the area of the copper wire to thermocouple terminations. The sensor output is used to determine the compensation needed to offset the electric potential generated at the copper wire to thermocouple termination.
Best regards
OMEGA
Date published: 2021-09-16
Is there a CAD step file for this module? If not how about a 2D drawing with the overall dimensions?
Thank you for your inquiry. Unfortunately, we do not have a CAD, Step File or 2-dimensional drawing available for this product. Dimensions: 89 H x 83 W x 95 mm D
(3.50 x 3.27 x 3.75") ; Panel Cutout: 92 x 92 mm (3.622 x 3.622")
Date published: 2022-08-04
Do you have mounting instruction? I can't find it the manual. What are the options that could be mount? Any accessory need?
Thank you for your inquiry. This instrument is a panel mount design for a standard 1/4 DIN panel opening (Panel Cutout: 92 x 92 mm (3.622 x 3.622 in), mounting hardware is included.
Date published: 2021-12-16