Product Overview
- AC Adaptor Operating Temperature 0 to 40 °C
- Accuracy Detail RH: 10 to 90%: ± 2.75%, 5 to 10% and 90 to 95%: ± 3%, 0 to 5% and 95 to 100%: ± 4%, Temperature: ± 0.5°C for 5 to 45°C; up to ± 1°C for 0 to 5°C and 45 to 70°C
- Configuration Internal Web UI
- Data Interface Ethernet
- Data Interface Note Ethernet : (RJ45) Fixed or auto negotiating 10/100BASE-T, AUTO MDI/MDIX Supported Protocols: TCP, UDP, SNMP, SMTP, NTP, ARP, ICMP, DHCP, DNS, HTTP, and Telnet
- Electrical Connection Sub-D Male Miniature Connector
- Form factor Wall Bracket Mount
- Height 1.08 in
- Hysteresis ± 1% RH
- Input Type External
- Length 3.67 in
- Measurement Type Relative Humidity, Temperature
- Non-Linearity ± 3%
- Number of Measurements 2
- Relative Humidity Accuracy ± 2.75% RH
- Repeatability ± 0.1%
- Response Time Temperature: 5 to 30 sec, RH: 8 sec
- Response Time Note Temperature: 5 to 30 seconds, tau 63%, RH: 8 seconds, tau 63%
- RH Measurement Range 0 to 100% RH
- RH Resolution 0.1%, 12 bit
- Samples,Max 85k Samples, Interval 1 sec
- Sensor Type Digital
- Series ID ITHX-W3-Series
- Software OPC Server; macro for data logging in Excel program; compatible with Windows® operating systems
- Storage Temperature, Max 125 °C
- Storage Temperature, Min -40 °C
- Supply Current 200 mA
- Supply Power Note Safety-Qualified AC Power Adaptor, Nominal Output: 9 Vdc @ 0.5 A , Input (iBT(H)X-W, iBTX-SD): 100 to 240 Vac, 50/60 Hz (included), Switching Power Supply (iBT(H)X-D): Sold separately (iDRN-PS-1000), Back-up Battery (iBTX-SD): 9V alkaline (included)
- Supply Power Type DC
- Supply Voltage 5 Vdc
- Temperature Accuracy ± 0.5 °C
- Temperature Measurement, Min 0 °C
- Temperature Repeatability ± 0.1 °C
- Temperature Resolution 0.1 °C
- Weight 0.68 lb
- Width 2.6 in
Supported Applications
The Omega iTHX-W3 Series is a temperature and humidity virtual chart recorder designed for network-enabled monitoring. The product overview identifies applications including clean rooms, computer rooms, HVAC systems, pharmaceutical storage, hospitals, laboratories, semiconductor labs, electronic assembly, warehousing, museums, manufacturing, greenhouses, and farm animal shelters.
Operating Conditions & Performance
The series measures temperature, relative humidity, and dew point. Measurement ranges vary by configuration: the standard iTHX-W3 supports -40 to 124°C (-40 to 254°F) for temperature and 0 to 100% RH. A narrow range option is available for 20 to 30°C (68 to 86°F). Accuracy specifications are tiered by temperature: ±0.5°C (±1°F) from 5 to 45°C (41 to 113°F); ±1°C (up to ±1°C) from 0 to 5°C and 45 to 70°C (32 to 41°F and 113 to 158°F); and ±1.5°C (up to ±1.5°C) for -40 to 5°C and 45 to 124°C (-40 to 41°F and 113 to 255°F). Humidity accuracy is listed as ±2.75% from 10 to 90% RH, ±3% from 5 to 10% and 90 to 95% RH, and ±4% for 0 to 5% and 95 to 100% RH.
Resolution is specified as 0.1°C (14 bit) with a repeatability of ±0.1°C. Response time ranges from 5 to 30 seconds, tau 63%. Operating temperature for the iTHX-W3 unit is 0 to 70°C (32 to 158°F), while the operating process temperature for industrial probe models iTHX-W3-2 or iTHX-W3-5 extends from -40 to 85°C (-40 to 185°F). Storage temperature capability is -40 to 125°C (-40 to 257°F).
Configuration Options
The series offers configurable probe types and lengths via specific model numbers. The iTHX-W3 includes a commercial wand probe on a 152 mm (6") cable with a DB9 connector. Industrial configurations include the iTHX-W3-2 with a 51 mm (2") probe on a 0.9 m (3') cable, and the iTHX-W3-5 with a 137 mm (5") probe on a 3 m (10') cable.
Connectivity is provided via Ethernet (RJ45) supporting fixed or auto-negotiating 10/100BASE-T. The device functions as an embedded web server delivering real-time data and live updated charts, with support for TCP/IP packets. Supported protocols include TCP, UDP, SNMP, SMTP, NTP, ARP, ICMP, DHCP, DNS, HTTP, and Telnet.
Power is supplied via a 5 Vdc @ 200 mA input using an included safety-qualified AC power adaptor (100 to 240 VAC, 50/60 Hz). The enclosure utilizes a Valox® 364 PBT case with a wall mount bracket. Approvals include CE and FCC-B; NIST-traceable calibration certificates are available as an option.
Key Product Differences
The series distinguishes between commercial wand probes and industrial probe configurations based on the model suffix:
- iTHX-W3: Features a commercial wand probe with 152 mm (6") cable length.
- iTHX-W3-2: Equipped with an industrial 51 mm (2") probe on a 0.9 m (3') cable; supports extended process temperature range of -40 to 85°C (-40 to 185°F).
- iTHX-W3-5: Equipped with an industrial 137 mm (5") probe on a 3 m (10') cable; supports extended process temperature range of -40 to 85°C (-40 to 185°F).
Replacement probes and accessories are available, including the iTHP-W-6 wand probe, iTHP-2-DB9 industrial probe (2"), and iTHP-5-DB9 industrial probe (5"). Probe caps include porous polyethylene options for wet environments and a 5 µm porosity stainless steel cap for dusty or pressurized (<35 psi) environments.
PDFs & Manuals
Software & Drivers
Accessories
accessories
installation hardware
sensors
Show Ratings & Reviews