Models In Stock
  • 5-Point NIST Traceable Calibration with
    59 kOhm Shunt Cal Data
  • High Accuracy
  • Aluminum for Light Weight and Durability
  • Calibrated in Tension
LCM105-LCM115
From C$ 925.80
Models In Stock
Metric, Lightweight, Aluminum S-Beam Load Cells LCM105-LCM115 LCM105-LCM115
loading all models
Product Overview
  • Bridge Resistance 350 ± 5 Ω
  • Cable Length 30 ft
  • Calibration Option 5 POINT NIST
  • Electrical Connection Cable
  • Electrical Output 3 mV/V
  • Enclosure Ratings IP65
  • End Rod MREC-M12M
  • Force Type Compression, Tension
  • Full Scale Deflection 0.02 in
  • Hysteresis ± 0.02%
  • Length 76 mm
  • Linearity ± 0.03%
  • Load Button MLBC-M12
  • Load Cell Capacity 500 kgf
  • Load Cell Type S-beam
  • Materials Aluminum
  • Operating Temperature, Max 93 °C
  • Operating Temperature, Min -40 °C
  • Overload, Max 300% FS
  • Process Temperature, Max 71 °C
  • Process Temperature, Min 17 °C
  • Repeatability ± 0.01%
  • Safe Overload 150% FS
  • Series ID LCM105-LCM115
  • Supply Power Note 10 Vdc, 15 Vdc maximum
  • Supply Voltage 10 Vdc
  • Thermal Span Effect 0.002% rdg/°C
  • Thermal Zero Effect 0.002% FSO/°C
  • Weight 0.5 kg
  • Width 25 mm
  • Wire Gauge 20 AWG
  • Wiring/Connector Table Green: (+) Output; White: (-) Output; Black: (-) Input; Red: (+) Input
  • Zero/Balance Error ± 1%

Supported Applications

The Omega LCM105 and LC105 series are aluminum 'S' Beam load cells designed for tension/compression applications. These lightweight, high-accuracy sensors are calibrated in tension and available in standard (LC) and metric (LCM) models to accommodate diverse measurement needs.

Operating Conditions & Performance

The series operates within an ambient temperature range of -40 to 93°C (-40 to 200°F), with a compensated temperature range of 17 to 71°C (60 to 160°F). Measurement ranges span from 0-25 lb to 0-1000 lb and 0-11 kgf to 0-455 kgf. Performance specifications include linearity of ±0.03% FSO, repeatability of ±0.01% FSO, and hysteresis of ±0.05% FSO for capacities ≤ 250 lbs or 100 kgf (±0.02% FSO for > 250 lbs or 100 kgf). The electrical output is 3 mV/V ±0.0075 mV/V with a bridge resistance of 350 ± 5 Ω, requiring an excitation voltage of 10 Vdc (maximum 15 Vdc).

Thermal effects are specified as Zero: 0.002% FSO/°C and Span: 0.002% rdg/°C. The load cells feature a Safe Overload rating of 150% of capacity and an Ultimate Overload rating of 300% of capacity, with full-scale deflection between 0.25 to 0.50 mm (0.010 to 0.020"). The series carries NIST Standard compliance and RoHS certification.

Configuration Options

The LCM105/LC105 series offers configurable process connections based on capacity ranges:

  • Low Ranges (≤ 250 lb / ≤ 100 kgf): Process connection of 1/4-20UNC-2B or metric M6 both ends.
  • High Ranges (> 250 lb / > 100 kgf): Threaded connections include ½-20 UNF for standard models and M12 x 1.75-6H for metric models.

Electrical connectivity is provided via a 4-Conductor Shielded Cable with lengths of 9 m (30'). Wire gauge varies by capacity: 24 AWG for ≤ 100 kgf/200 lb and 20 AWG for > 100 kgf/200 lb. The series supports an IP65 protection class.

Optional configurations include the LC115/LCM115 connector style, which changes the model number suffix without additional charge. Accessories available separately include:

  • PT06F10-6S: Mating connector for LC115/LCM115 series load cells.
  • CA-4PC24-2-015: Cable assembly with twist lock connector for LC115/LCM115 series load cells.

The units come complete with 5-point NIST-traceable calibration and a 59 kΩ shunt. Compatible meters include DPiS, DP41-S, DP25B-S (for standard models) and DP41-B, DP41-S, DP25B-S (for metric models).

Key Product Differences

The series distinguishes between standard inch-based units (LC105/LC115) and metric-based units (LCM105/LCM115). The LC115 and LCM115 variants differ from the base 105 models by offering a connector style option rather than a fixed cable, requiring separate purchase of mating connectors or twist-lock assemblies. Dimensions vary by capacity: low-range models feature an overall diameter of ø38 (1.50) mm (inch) and height of 64 (2.5) mm (inch), while high-range models have specific dimensional parameters A through E corresponding to their threaded connections.

Accessories

accessories

ROD END METRIC
C$45.26 / each
33 In Stock
TWIST LOCK CONNECTOR NON-ROHS
C$66.26 / each
505 In Stock

installation hardware

LOAD BUTTON
C$131.39 / each
0 In Stock
LOAD CELL
C$135.20 / each
0 In Stock
METRIC ROD END
C$45.26 / each
0 In Stock
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 5ms
  • REVIEWS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getContent, 4ms
  • QUESTIONS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;
en-CA
en_CA
Save Configuration