Need assistance? Call us: 1-800-663-4209
HX300A-Series
Humidity/Temperature Transmitters
Models In Stock
- Fast, Stable and Accurate
- Watertight Enclosure
- Compact, Easy to Use
- Excellent Long-Term Stability
HX300A-Series
From
C$
423.81
Models In Stock
Product Overview
- Accuracy ±2.5 % RH, ±0.3°C
- Electrical Connection Cable Gland for 5 to 10 mm Wire with Screw Terminal block
- Enclosure Ratings IP65
- Long Term Stability Better than 1% RH per year (typical)
- Materials, Wetted ABS
- Operating Temperature, Max 50 °C
- Operating Temperature, Min 0 °C
- Response Time 15 sec
- Sensor Material Sintered filter
- Series ID HX300A-Series
- Supply Power Type DC
- Supply Voltage 12 to 40 Vdc
- Temperature Compensation 0.008% RH/°C (effect @ 05 RH)
- Weight 150 g
Supported Applications
The Omega HX300A Series consists of wall-mount transmitters designed for industrial environments requiring fast, stable, and accurate measurements. The series employs a highly stable thin-film polymer capacitor to sense relative humidity and a high-accuracy, thin-film RTD (PT100 Ω) compliant with IEC 751 and DIN 43760 standards to accurately sense temperature.
Operating Conditions & Performance
The HX300A Series operates within an ambient temperature range of 0 to 50°C (32 to 122°F). Measurement capabilities include a humidity range of 0 to 100% RH @ 0 to 50°C (32 to 122°F) and a temperature measurement range of 0 to 100°C (32 to 212°F).
Performance specifications at 25°C include an accuracy of ±2.5% RH, ±0.3°C with long-term stability better than 1% RH per year (typical). The response time is <15 seconds (90% @ 25°C in moving air at 0.5 m/sec), and temperature compensation is rated at 0.008% RH/°C (effect @ 05 RH).
The unit features a sintered filter for sensor protection and utilizes an ABS plastic watertight enclosure to NEMA 4 (IP65). Electrical connections are provided via liquid-tight nylon cable bushing fitting 5 to 10 mm dia. (0.2 to 0.4") cables.
Key Specifications
- Housing: ABS plastic watertight enclosure to NEMA 4 (IP65)
- Sensors - Humidity: Thin-film capacitor
- Sensors - Temperature: RTD PT100 Ω, IEC 751, DIN 43760
- Weight: Approx 150 g (5.3 oz)
- Approvals/Certifications: CE, NIST-traceable calibration available via CAL-3-HU accessory
Configuration Options
The series supports multiple output configurations and sensor combinations to address specific monitoring needs. Power supply requirements are 12 to 40 Vdc (24 Vdc recommended).
Output Signals
- Current Output: 4 to 20 mA, 2-wire with ±15% scalable zero and span adjustment.
- Voltage Output: 0 to 1 V (primary series specification).
Sensor Configurations
- Humidity only models (HX302A family)
- Temperature and Humidity combined models (HX303A family)
Accessories & Options
The series is offered with various power supply options including:
- PSR-24S: Regulated power supply, US plug, 90 to 264 Vac input, 24 Vdc output, 400 mA, screw terminals, UL.
- PSR-24L: Regulated power supply, US plug, 90 to 264 Vac input, 24 Vdc output, 400 mA, stripped leads, UL.
- PSR-24L-230: Regulated power supply, European plug, 230 Vac input, 24 Vdc output, 400 mA, stripped leads, CE.
- PSU-93: Unregulated power supply, 16 to 23 Vdc, 300 mA max, screw terminal.
Key Product Differences
The HX300A Series is differentiated by sensor type and output signal format. The model number guide follows the pattern: HX300A - <MODEL> - <OUTPUT TYPE>.
- HX302AC: Wall mount RH transmitter, 4 to 20 mA current output.
- HX302AV: Wall mount RH transmitter, 0 to 1 voltage output.
- HX303AC: Wall mount RH/temperature transmitter, 4 to 20 mA current outputs.
- HX303AV: Wall mount RH/temperature transmitter, 0 to 1 voltage outputs.
Show Ratings & Reviews
Are these compatible in a corrosive environment?
Thank you for the inquiry. No, these are not compatible in a corrosive environment.
Date published: 2023-08-28
Is there a procedure/manual to use the calibration buttons for the HX302AC?
Hi Michael
Please refer to manual for procedure;
https://assets.omega.com/manuals/M5072.pdf
Best regards
OMEGA
Date published: 2025-08-14
Are these designed to be installed in a bulkhead or throughpanel configuration? If so what is the recommended method for accomplishing this?
Hi TimMcMan,
Thank you for the inquiry. The HX302C is designed for mounting on a wall. Manual supplies detail including mounting hole location.
Best regards
OMEGA
Date published: 2022-12-05
Is this safe to use in a commercial combi-oven for creating jerky? I need to record the percentage of humidity and temperature. Oven temperature would be no more than 170F and can generate its own steam.
Thank you for your inquiry. We believe the best assistance would come from speaking to us directly at 888-826-6342 ask for the Temperature Department Mon. – Fri. (HOURS 8:30AM - 5:30PM EST).
Date published: 2024-08-20
Is there a more extensive manual for the HX303AV model? This a jumper on it to change the output range to 0 to 5V but that is not mentioned in the online manual. I'm seeing 0.4V for ambient temp reading and 0.9V for RH. Seems wrong?
Thank you for the inquiry. Manual for the HX303A; https://assets.omega.com/manuals/M5072.pdf. For more trouble shooting assistance please forward email to TEMP@OMEGA.COM.
Best regards
OMEGA
Date published: 2024-06-25
the data that I am getting to my PLC is not coming through as a percentage but rather a 4 digit number. Is there a formula that I need to apply to the data to get the information that I need? I doubt 2894 is the correct RH. Any help is greatly appreciated
Hi James,
You'll need to check the PLC scaling and adjust as needed. The 4 to 20 mA outputs of the HX303AC are scaled to represent 0 to 100 % RH and 0 to 100 C.
Best regards
OMEGA
Date published: 2023-03-01