- Internal Electric Pressure Pump Generates Pressure from -0.8 to 20 bar (-12 to 300 psi)
- Accuracy to 0.025% FS
- Source and Measure 4 to 20 mA
- 24 Vdc Loop Power Supply to Power Device Under Test
Product Overview
- Accuracy Detail ±0.015% of reading ±2 counts (simulation and measurement); Voltage Accuracy: ±0.015% of reading ±2 counts (measurement only)
- Additional Features 1 Interchangable Pressure Module
- Battery Size and Type Alkaline AA
- Construction Details Electrically operated pneumatic pump generates -0.8 to 20 bar (-12 to 300 psi)
- Current Accuracy ± 0.015 %
- Electrical Connection Banana Plug
- Electrical Connection Note RTD Connection: 4 pin LEMO style ; External Module: 6 pin LEMO style
- Form factor Handheld
- Functions Pressure Source & Pressure Reference & Signal Indication
- Height 200 mm
- Length 60 mm
- Operating Temperature, Max 50 °C
- Operating Temperature, Min -10 °C
- Pressure Accuracy ± 0.025 %
- Process Connection Size 1/8 in
- Process Connection Type NPT Female
- Pump Type Pnuematic
- Series ID PCL819
- Storage Temperature, Max 60 °C
- Storage Temperature, Min -20 °C
- Temperature Accuracy ± 0.1 °C
- Temperature Accuracy Note ±0.1°C (0.2°C) (measurement only)
- Voltage Accuracy ±0.015%
- Weight 1.2 kg
- Width 100 mm
Supported Applications
The Omega PCL819 is a handheld pressure calibrator featuring an integral electrically operated pneumatic pump. It combines high-performance measurement with portability to calibrate virtually any pressure device, including transmitters and switches. The series supports Gage and Absolute pressure measurements across multiple ranges. Additionally, the unit provides inputs for Current (mA), Voltage, Switch Contacts, and Temperature (RTD) calibration.
For extended range requirements beyond the internal sensor limits, the PCL819 is compatible with the PCL-PM series external pressure modules via a required adaptor. These external modules support ranges from 5 to 5000 psig as well as differential and compound configurations.
Operating Conditions & Performance
The internal isolated pressure sensor delivers an accuracy of ±0.025% FS when operating between 15 to 35°C (59 to 95°F). Outside this temperature band, add 0.002%/°C to the uncertainty. The electrically operated pneumatic pump generates vacuum or pressure from -0.8 to 20 bar (-12 to 300 psi).
Electrical measurement performance includes:
- mA (simulation and measurement): ±0.015% of reading ±2 counts.
- Voltage (measurement only): ±0.015% of reading ±2 counts.
- Temperature: ±0.1°C (±0.2°C) for measurement only.
The unit operates in ambient temperatures from -10 to 50°C (14 to 122°F). Storage temperature is rated from -20 to 60°C (-4 to 140°F).
Power and Battery Life
The PCL819 is powered by 8 standard "AA" Alkaline batteries. Minimum battery life specifications are:
- 100 pump cycles to 300 psi.
- 1,000 cycles to 30 psi.
A rechargeable battery kit (PCL819-BATT-KIT) is available as an accessory option. The unit includes a 24 Vdc Loop Power Supply to power the Device Under Test.
Configuration Options
The PCL819 series offers configurable pressure ranges and measurement types through specific model variants:
- Gage Pressure Models:
- PCL819-30: -0.8 to 2 bar (-12 to 30 psi)
- PCL819-150: -0.8 to 10 bar (-12 to 150 psi)
- PCL819-300: -0.8 to 20 bar (-12 to 300 psi)
- Absolute Pressure Models:
- PCL819-30A: 0 to 2 bar (0 to 30 psi)
- PCL819-150A: 0 to 10 bar (0 to 150 psi)
Process connection is provided via a 1/8 NPT female port. Electrical connections utilize standard banana jacks for test leads, while RTD inputs use a 4 pin LEMO style connector and external modules connect via a 6 pin LEMO style interface.
Housing and Physical Specifications
The handheld unit measures 200 H x 100 W x 60 mm D (8 x 4 x 2 3/8") and weighs 1.2 kg (2.5 lb). The housing is contoured to fit a technician's hand with balanced weight distribution.
Key Product Differences
The primary distinction within the PCL819 series lies in the pressure range and measurement type (Gage vs. Absolute) defined by the model suffix. All models share the same internal electric pump capability, electrical input functions, and physical dimensions.
Accessories
The following accessories are available to support specific operational needs:
- PCL819-TRAP: Spare liquid and dirt trap.
- PCL-PMA: Pressure module adaptor, required to use PCL-PM series external pressure modules.
PDFs & Manuals
Show Ratings & Reviews
is the display selectable for different scales? I.E. inH2o, Bar ,PSI
There are 15 units of measure that can be selected by the user during setup. They are: psi, bar, mbar, kPa, MPa, kgcm2, mmH2O @ 4°C, mmH2O @ 20°C,
cmH2O @ 4°C, cmH2O @ 20°C, inH2O @ 4°C, inH2O @ 20°C, inH2O
@ 60°F, mmHg @ 0°C, inHg @ 0°C
Date published: 2021-12-28
Can this device be powered by a wall outlet instead of battery power?
No.
Per page 42 in manual:
Power Requirements
Battery
12 VDC
Eight (8) AA alkaline, Lithium or NiMh cells