PCL1000-Series

Handheld Multifunction Calibrator with Pressure Modules

  • Excellent Accuracy ±0.6°C T/C, ±0.3°C RTD 0.015% Electrical 0.05% Pressure
  • Read and Source Modes
  • 10 T/C Types  8 RTD types plus 400 Series Thermistor
  • Stores up to 9 setpoints for each output function
PCL1000-Series
From C$ 3,178.77
Handheld Multifunction Calibrator with Pressure Modules PCL1000-Series PCL1000-Series
loading all models
Product Overview
  • Battery life 25 hr
  • Battery Size and Type Alkaline AAA
  • Continuity Test 0 to 400.0 Ω; audible tone under 10 Ω
  • Data Interface RS232
  • Frequency CPM: Source 2 to 1000.0; measure 0 to 1000.0,Hz: Source 0 to 1000; measure 0 to 1000.0,kHz: Source 1 to 10.0; measure 0 to 10.00
  • Height 221 mm
  • Length 58 mm
  • Loop Supply Limits Voltage Drive: 1 mA
  • Current, Max 0 to 24 mA
  • Resistance 3200 Ω
  • RTD Input Note RTD Types: JIS, Ni120, Cu10; RTD IEX Range: 0.01 to 3 mA; RTD Frequency Response: 10 ms
  • RTD Type Pt 385 (100, 200, 500, 1 kΩ)
  • Series ID PCL1000-Series
  • Thermistor Type Series 400
  • Thermocouple Type J, K, T, E, N, R, S, U, B, L
  • Voltage Note Low: 0 to 20.000 Vdc; read and source High: 0 to 250 Vdc or Vac; read only
  • Weight 0.9 kg
  • Width 107 mm

Supported Applications

The Omega PCL1000 is a high-accuracy multifunction calibrator designed to read and source thermocouples, RTDs, thermistors, voltage, current, frequency, resistance, and pressure signals. It combines the functions of a precision calibrator and multimeter for process device calibration tasks.

Documented measurement types include:

  • Thermocouple (Types J, K, T, E, R, S, N, B, L, U)
  • RTD (Pt 385: 100, 200, 500, 1 kΩ; Pt 392: JIS, Ni120, Cu10)
  • Thermistor (Series 400)
  • Voltage and Current
  • Frequency (CPM, Hz, kHz)
  • Resistance
  • Pressure via optional PCL-PM series modules

Operating Conditions & Performance

The unit operates with a power supply of 4 "AA" batteries (included), providing typical battery life of 25 hours. Physical dimensions are 221 H x 107 W x 58 mm D (8.7 x 4.2 x 2.3") and weight is 0.9 kg (1.9 lb). A protective rubber boot is included.

Performance specifications vary by measurement type:

  • Voltage Accuracy: ±0.015% rdg Electrical
  • T/C Type J Accuracy: ±0.3°C ±10 μV (±0.4°C total)
  • RTD Pt100 Accuracy: ±0.1°C ±0.075 Ω (±0.3°C total)
  • Resistance Accuracy: 400 Ω: ±0.1 Ω; 3200 Ω: ±1.0 Ω
  • Pressure Accuracy: 0.25% FS Pressure (via PCL-PM modules)
  • mA Compliance: Up to 1000 Ω load
  • Voltage Drive: 1 mA
  • RTD IEX Range: 0.01 to 3 mA
  • RTD Frequency Response: 10 ms
  • Continuity: 0 to 400.0 Ω; audible tone under 10 Ω

Configuration Options

The PCL1000 series supports a broad range of electrical and temperature measurement capabilities with optional pressure expansion.

Measurement Ranges

  • Voltage: Low: 0 to 20.000 Vdc (read and source); High: 0 to 250 Vdc or Vac (read only)
  • Current: 0 to 24.000 mA
  • Resistance: 0 to 400.00 Ω and 400 to 3200 Ω
  • Frequency (CPM): Source 2 to 1000.0; measure 0 to 1000.0
  • Frequency (Hz): Source 0 to 1000; measure 0 to 1000.0
  • Frequency (kHz): Source 1 to 10.0; measure 0 to 10.00
  • Pressure: Ranges from 5 to 5000 psig via optional PCL-PM modules, including vacuum, absolute, compound, and differential pressures.

Configuration & Interfaces

  • User Interface: Direct keyboard entry with storage of up to 9 setpoints per output.
  • Communication: RS232 interface available. Note: Pressure module uses RS232 output; computer connection via RS232 is not possible while a pressure module is connected.

Included Accessories

  • Test leads
  • Type K thermocouple with cable
  • Protective rubber boot
  • 4 "AA" alkaline batteries
  • Calibration certificate and operator's manual

Key Product Differences

The base model PCL1000 is a high-accuracy multifunction calibrator. Pressure capability requires the addition of specific accessories:

  • PCL-PMA: Required pressure module adaptor for connecting PCL-PM modules.
  • 700MA: Required adaptor to RS232 port for pressure module communication.
  • PCL1000-SCC: Soft carrying case (optional).

The PCL-PM series modules are sold separately and required for pressure applications. The base unit does not include a built-in pressure sensor; it interfaces with external modules via the RS232 port using the specified adaptors.

  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 4ms
  • REVIEWS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getContent, 60ms
  • QUESTIONS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;
en-CA
en_CA
Save Configuration