-5 °C to 125 °C Temperature Dry Block Calibrator
- Operates from -5 to 125°C (23 to 257°F)
- Rugged, Portable Benchtop Design
- Fast, Accurate, Stable Readings
- Standard and Metric Well Designs Available
CL1500-Series
From
C$
5,768.07
Product Overview
- Accuracy ± 0.8 °C (± 1.5 °F)
- Additional Features 1 Portable
- Data Interface RS232
- Display Resolution 0.1°
- Heat-Up Time Note -5°C to Ambient: 1 min Ambient to 100°C: 3 min 100 to 125°C: 5 min
- Height 79 mm
- Length 203 mm
- Operating Temperature, Max 45 °C
- Operating Temperature, Min 0 °C
- Power Consumption 200 W
- Process Temperature, Max 125 °C
- Process Temperature, Min -5 °C
- Process Temperature Note -5 to 125°C (23 to 257°F) at 23°C (73°F) Ambient;Temperature Uniformity: ±0.55°C
- Relative Humidity, Max 0.8
- Relative Humidity, Min 0%
- Relative Humidity Note 0 to 80% RH non-condensing
- Sensor Type RTD
- Sensor Type Note Platinum RTD, 100 Ω, 0.00385, Class A
- Series ID CL1500-Series
- Stabilization Time 5 min
- Supply Power Note 115 or 230V versions
- Temperature Stability ±0.25°C
- Weight 2.8 kg
- Width 206 mm
Supported Applications
The Omega CL1500 Series is a portable, rugged bench-top dry block calibrator designed to test and calibrate temperature probes of various diameters. It functions as a hot/cold calibration source featuring a built-in precision PID digital controller for laboratory and field use.
Operating Conditions & Performance
The CL1500 Series operates within an ambient environmental range of 0 to 45°C (32 to 113°F) with humidity between 0 to 80% RH non-condensing. The unit provides a temperature calibration range of -5 to 125°C (23 to 257°F) at 23°C (73°F) Ambient.
Performance specifications include an accuracy of ±0.8°C (±1.5°F), stability of ±0.25°C (±0.45°F), and well-to-well uniformity of ±0.55°C (1°F). The display resolution is 0.1 degree, with a stabilization time of 5 min.
Heating times are documented as -5°C to Ambient: 1 min; Ambient to 100°C: 3 min; and 100 to 125°C: 5 min. Cooling times include 125 to 100°C: 1 min, 100 to 0°C: 8 min, and 0 to -5°C: 4 min.
Configuration Options
The series is available in two well styles: Standard (CL1500) with wells for 1/16, 1/8, 5/32, 3/16, 1/4, and 3/8" diameter probes; and Metric (CL1500M) with wells for 1.5, 2, 3, 4.5, 6, and 8 mm diameter probes. The calibration block features 6 holes to accommodate these varying diameters.
Power supply options include 90 to 240 Vac at 200 W. Specific models are configured for 115 Vac or 230 Vac operation. Each unit includes a power cord, operator's manual, RS232 cable (with an included RS232 to USB converter), software, and a NIST traceable calibration certificate.
Available accessories include:
- HH806W - Wireless thermocouple thermometer
- HHM290 - Digital multimeter/infrared thermometer with laser sighting dot/circle switch
- TJ36-CASS-316U-12-SMPW-M - Heavy-duty 12" transition, joint probe, Type K, 3/16" diameter ungrounded with sub-mini connector
The unit features a custom fabricated metal enclosure and an internal control sensor of Platinum RTD, 100 Ω, 0.00385, Class A.
Key Product Differences
Variants differ primarily by well style (Standard vs. Metric) and voltage configuration:
- CL1500: Hot/cold dry block calibrator with 1/16, 1/8, 5/32, 3/16, 1/4, and 3/8" diameter wells.
- CL1500M: Hot/cold dry block calibrator with 1.5, 2, 3, 4.5, 6, and 8 mm diameter wells.
The model number guide indicates that the suffix "-230" is added for 230 Vac models (e.g., CL1500-230). The series includes a well depth of 114.3 mm (4.5") and overall dimensions of 206 W x 79 H x 203 mm D (8.1 x 3.1 x 8"). Weight is listed as 2.8 kg (6.3 lb).
PDFs & Manuals
Software & Drivers
Show Ratings & Reviews
Rated 3 out of
5
by
CJ AAQD from
Hot/Cold Benchtop Dry Block Calibrator wrong unit
I only have one issue , the temperature of the calibrator is in Fahrenheit, I ordered in Celsius.
Date published: 2021-02-26
How many probes can it hold/calibrate at once? I have 11 thermocouples I was hoping to place into a dry block calibrator at once.
Thank you for the inquiry. CL1500 has a 6 port calibration block. Sizes 3/8 down to 1/16 inch dia. Refer to figure 2 in manual. DBCL-400 block calibrator can be used with optional blank well. You can have the blank well drilled by 3rd party machine shop to match your probes.
Please email TEMP@OMEGA.COM for more details.
Best regards
OMEGA
Date published: 2023-02-06
What is the optional blank well part number? I don't see it as an accessory or selectable option.
Thank you for the inquiry. The CL1500 does not have a removeable well. The DBCL-400 does offer removeable well.
Best regards
OMEGA
Date published: 2023-03-08