Models In Stock
  • -50 to 1650°C (-58 to 3002°F) temperature range
  • 50:1 Field of View
  • Adjustable Emissivity: 0.10 to 1.00
  • 250 ms response time
OS310-Series
From C$ 481.71
Models In Stock
-20 to 1650 °C, 50:1 High Performance Infrared Thermometer OS310-Series OS310-Series
loading all models
Product Overview
  • Accuracy ± 1.5 °C (± 2.7 °F)
  • Accuracy Detail -50 to 0°C (-58 to 32°F): ± 3°C (± 5°F) 0 to 1650°C (32 to 3002°F): ± 1.5% or 1.5°C (± 2.7°F)
  • Alarm High and Low
  • Battery Size and Type Alkaline 9V
  • Display Monitoring Function Average, Max, Memory, Min
  • Display Type LCD
  • Emissivity, Max 1
  • Emissivity, Min 0.1
  • Field of view 50:1
  • Form factor Pistol Grip
  • Height 200 mm
  • Laser Dot
  • Length 60 mm
  • Operating Temperature, Max 40 °C
  • Operating Temperature, Min 0 °C
  • Process Temperature, Min -50 °C
  • Relative Humidity, Max 0.95
  • Relative Humidity, Min 10%
  • Relative Humidity Note 10 -95% RH non-condensed
  • Repeatability <1/3
  • Repeatability Note Less than 1/3 of basic error
  • Response Time <250 ms
  • Series ID OS310-Series
  • Spectral Response, Max 14 µm
  • Storage Temperature, Max 50 °C
  • Storage Temperature, Min -20 °C
  • Weight 270 g
  • Width 141 mm

Supported Applications

The Omega OS310-Series is a high-performance, handheld infrared thermometer designed for non-contact temperature measurement of metals, ceramics, and graphite. The series supports measurements across the range of -50 to 1650°C (-58 to 3002°F). It features an adjustable emissivity setting (0.10 to 1.00) to accommodate various surface materials.

Documented material applications include:

  • Metallic surfaces: Aluminum, Brass, Copper, Iron, Lead, Steel
  • Ceramics and minerals: Basalt, Brick, Carbon, Ceramic, Concrete, Limestone, Sand
  • Organic and composite materials: Asbestos, Asphalt, Paint, Paper, Plastic, Rubber, Textiles, Wood
  • Liquids and other media: Oil, Water, Frozen Food, Hot food, Ice, Skin, Snow

Operating Conditions & Performance

The OS310-Series operates within an ambient temperature range of 0 - 40°C (32 - 104°F) and relative humidity from 10% to 95% non-condensing. Storage is supported between -20 - 50°C (-4 - 122°F). The device utilizes a spectral response of 8 – 14 µm with a field of view ratio of 50:1.

Performance specifications are defined by temperature range:

  • -50 to 0°C (-58 to 32°F): Accuracy ±3°C (±5°F)
  • 0 to 1650°C (32 to 3002°F): Accuracy ±1.5% or 1.5°C (±2.7°F), whichever is better

The thermometer provides a resolution of 0.1°C or 0.1°F and achieves response times less than 250 ms. Repeatability is specified as less than 1/3 of basic error.

Configuration Options

The series includes the OS-310-501 model, a handheld unit powered by an included 9V battery. The device features a backlit LCD display that shows current temperature as well as minimum, maximum, average, and differential temperatures.

  • Target Aiming: Single laser dot for target location
  • Emissivity Adjustment: Configurable from 0.1 to 1.0 (or 0.10 to 1.00)
  • Alarm Settings: Available on the unit
  • Housing and Weight: Dimensions of 5.6 x 7.9 x 2.4 in (141 x 200 x 60 mm); weight is 270 g (9.5 oz) including battery

The unit ships complete with a hard case and manual.

Key Product Differences

The OS310-Series is represented by the single model number OS-310-501, described as a Handheld Infrared Thermometer with a 50:1 Field of View covering -50 to 1650°C. No alternative housings, output signal types, or sensor element variations are documented for this series.

Rated 4 out of 5 by from Accuracy lacking This device has a large range in temperature, and I like the simplicity of use. It is sturdy and a quality device, But the accuracy is not as displayed. I trust it to 20-30 degrees Manufacturer needs to work on accuracy. I am using it to measure the temperature of a Baily Gas Kiln and the chimney.
Date published: 2026-02-01
  • y_2026, m_8, d_10, h_10
  • bvseo_bulk, prod_bvrr, vn_bulk_3.0.48
  • cp_1, bvpage1
  • co_hasreviews, tv_0, tr_1
  • loc_en_CA, sid_OS-310-501, prod, sort_[SortEntry(order=HELPFULNESS, direction=DESCENDING), SortEntry(order=POSITIVE_FEEDBACK, direction=DESCENDING), SortEntry(order=SUBMISSION_TIME, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 7ms
  • REVIEWS, PRODUCT

Any way to get the default temp to read in Fahrenheit not Celsius on this model? It keeps restarting in Celsius..

Asked by: Drewzilla
Hi Drewzilla, Follow up from previous question on F/C selection; The Fahrenheit selection will not remain active through power cycle. It will still display Celsius , you need to press C/F key to switch it . Best regards Omega
Answered by: Norwalk AE_RS
Date published: 2025-12-29
  • y_2026, m_8, d_9, h_10CST
  • bvseo_bulk, prod_bvqa, vn_bulk_3.0.48
  • cp_1, bvpage1
  • co_hasquestionsanswers, tq_1
  • loc_en_CA, sid_OS-310-501, prod, sort_[SortEntry(order=NUM_ANSWERS, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getContent, 11ms
  • QUESTIONS, PRODUCT
en-CA
en_CA
Save Configuration