Skip to search Skip to main content Skip to navigation

Netherlands 07077 03815 | Intl +31 (0) 7077 03815 | Welcome to our new website, found an issue or bug? Please report it here

We noticed you are in , please visit your local site by clicking the below link for regional offers and support.

RD5111

Hybrid Chart Recorder, 250 mm (10")

€ 10.567,38 / each

Item# RD5111

In Stock

Lead Time (If not in stock): 16 weeks

Product Specs

<< See all Product Models
Brand
Accuracy ± 0.05 %
Accuracy Detail Thermocouple K: ±0.05% ±0.5°C (±20 mV, ±40 mV), ±0.05% ±1°C (±80 mV), E: ±0.05% ±0.7°C (±20 mV, ±40 mV), ±0.05% ±1°C (±80 mV), J: ±0.05% ±0.7°C (±20 mV, ±40 mV), ±0.05% ±1°C (±80 mV), T: ±0.05% ±0.7°C (±20 Mv), R,B: ±0.05% ±1°C (±20 Mv), U,L: ±0.05% ±1°C, W-WRe26, PrRh40-PtRh20, NiMo-Ni: ±0.1% ±1°C, N: ±0.1% ±0.1°C, Platinel II: ±0.1% ±0.1°C (±20 Mv), ±0.1% ±1°C (±40 mV, ±80 mV), RTD: ±0.05% ±0.3°C, DC Voltage: ±0.05% ±1 digit
Alarm Note Alarm Types: High limit, low limit, Alarm Setting Method: Individual setting for each point four levels/ channels
Battery Size and Type Li-ion
Battery life up to 5 yr
CMRR 130 dB
Chart Length 20000 mm
Chart Note Fan-fold type; overall width 318 mm (12.5"), total length 20 m (65.6'); effective recording width 250 mm (10") (analog recording)
Chart Speed 1 to 1,500 mm/h
Chart Width 250 mm
Common Mode Voltage 30 Vac
Cutout Dimensions 388 x 248 mm
Data Interface USB
Data Interface Note Inside of front door, USB1.1, full speed 12 mbps, bulk transfer, Control transfer and Parameter setting for exclusive application
Dielectric Strength Between Primary Terminals and Protective Conductor Terminals: 1 minute at 1500 Vac, Between Secondary Terminals and Protective Conductor Terminals:1 minute at 500 Vac, Between Primary Terminals and Secondary Terminals: 1 minute at 1500 Vac
Display Note Color LCD panel RGB (640 x 240 dot), Size: 149.8 W x 57.4 mm H, Setting Display: Common to digital display Display Contents: Digital display, Channel Display: One-point/multiple points continuous/sequential indication change, Display Measuring Value of Each Channel: One-point/multiple points continuous/sequential indication change, Clock Display: Hour/Minute/ Second/Tag/Unit. Alarm Display: Occurrence CH No, data is displayed in red when alarm occurs
Display Type LCD
Electrical Connection Note Terminal Board: Detachable type, removable for wire connection
GrossWeightValue 0.15
Height 260 mm
Input Note Input Correction: Zero/span correction and shift correction for each channel
Input Resistance 1 MΩ
Insulation Resistance Between Primary Terminals and Protective Conductor Terminals:20MΩ or more at 500 Vdc, Between Secondary Terminals and Protective Conductor Terminals: 20MΩ or more at 500 Vdc, Between Primary Terminals and Secondary Terminals: 20MΩ or more at 500 Vdc
LED Indicators Record On: Lights during recording; LED, Key Lock: Lights during key lock, Alarm: Lights during alarm activated; LED, Chart End: Lights just before record ending, Fail: Lights during unit abnormal time
Length 300 mm
Loop Resistance, Max Thermocouple Inputs, DC Voltage Input (10 mV): 500Ω or less (without burnout), DC Voltage Input (Except 10 mV): 100Ω or less Resistance, Thermometer Inputs: 10Ω or less/ line, three lines are common, Pt100, JPt100
Materials ABS, Soda glass, Steel
Measurement Range Note Thermocouple: K: Reference Range: ±20 mV: -200 to 500°C, ±40 mV: -200 to 900°C, ±80 mV: -200 to 1370°C, E: Reference Range: ±20 mV: - -200 to 250°C, ±40 mV: -200 to 500°C, ±80 mV: -200 to 900°C, J : Reference Range: ±20 mV: - -200 to 350°C, ±40 mV: -200 to 700°C, ±80 mV: -200 to 1200°C, T: Reference Range: ±20 mV: - -200 to 400°C, R: Reference Range: ±20 mV: 0 to 1760°C, B: Reference Range: ±20 mV: 0 to 1300°C, N : Reference Range: ±20 mV: 0 to 600°C, ±40 mV: 0 to 1000°C, ±80 mV: 0 to 1300°C, W-WRe26: Reference Range: ±80 mV: 0 to 2315°C, W-WRe26: Reference Range: ±80 mV: 0 to 2315°C, PtRh20-PrRh40: Reference Range: ±20 mV: 0 to 1888°C, U : Reference Range: ±20 mV: -200 to 350°C, ±40 mV: -200 to 600°C, L : Reference Range: ±20 mV: -200 to 350°C, ±40 mV: -200 to 700°C, ±80 mV: -200 to 900°C, Nimo-Ni: Reference Range: ±80 mV: -50 to 1310°C, Platinel II: Reference Range: ±20 mV: 0 to 500°C, ±80 mV: 0 to 950°C, 0 to 1395°C RTD: Reference Range: 50Ω: -50 to 50°C , 100Ω: -100 to 130°C, 200Ω: -200 to 250°C, 300Ω: -200 to 550°C Voltage: ±10 mV, ±20 mV, ±40 mV, ±80 mV, ±1.5 V, ±2.5 V, ±5 V, ±10 V
Measurement Type Thermocouple, RTD, Voltage
Mounting Panel Mount
NMRR 50 dB
Number of Channels 12
Number of Relays 12
Operating Environment Reference Operating Condition: Ambient Temperature/Humidity Range: 21 to 25 °C (70 to 77°F), 45 to 65% RH,Power Voltage: 90 to 264V,Power Frequency: 50/60Hz ±2%,Attitude: Forward/Backward/left/right within 3°,Warm-Up Time: 1 hour or longer; Normal Operating Condition: Ambient Temperature/Humidity ,Range: 0 to 40 °C (32 to 104°F), 20 to 80% RH,Power Voltage: 90 to 264V,Power Frequency: 50/60Hz ±2%,Attitude: Forward/backward/left/right within 3°
Operating Temperature, Max 40 °C
Operating Temperature, Min 0 °C
Power Consumption 100 VA
RTD Type 100 Ω pt or Jpt sensor; 3-wire Pt100
Recording Color Analog Recording: Red, Purple-Red, Orange, Brown, Green, Yellow-Green, Blue-Green, Purple, Purple-Blue, black, Digital Recording and Logging: Black
Recording Note Recording System: Raster scan system, 10-color wire dot printing; Compensation: Independent setting of zero spans are available, Operation: Record On/Off: Recording operation ON/OFF*, Data Print: Printing measuring data*, Feed: Fast-forwarding chart paper
Relative Humidity 20 to 80% RH
Resolution Input: 1/40,000; Display: Thermocouple, RTD: 0.1°C, DC Voltage: 1 µV(±10 mV, ±20 mV), 10 µV(±40 mV, ±80 mV), 100 µV(±1.25V, ±2.5V), 1 mV(±5V, ±10V)
Sampling Rate 0.1 sec/channel
Sensor Break Detection Thermocouple Burn-Out: Select with/without burnout for each input
Series ID RD5100-Series
Storage Temperature, Max 60 °C
Storage Temperature, Min -20 °C
Supply Power Note Power Voltage: 90 to 264V, Power Frequency: 50/60Hz ±2%
Supply Power Type AC
Supply Voltage 100 to 240 Vac
Switch Rating 3 A @ 100 to 240 Vac
Switch Rating Note 100 to 240 Vac, 3 A resistance load
Switch Type SPST NO
Temperature Drift 0.1%/10°C
Temperature Measurement Range K: -200 to 1370°C, E: -200 to 900°C, J: -200 to 1200°C, T: -200 to 400°C, R: 0 to 1760°C, B: 0 to 1300°C, N: 0 to 1300°C, W-WRe26: 0 to 2315°C, B: 0 to 1888°C, L: -200 to 900°C, Nimo-Ni: -50 to 1310°C, Platinel II: 0 to 1395°C, RTD: -200 to 550°C
Thermocouple Type J, K, T, E, N, R, S, U, B, G, L, P, NiMo-Ni, PrRh40-PtRh20
Transportation and Storage Details Transportation Condition: Ambient Temperature/Humidity ,Range: -20 to 60°C (-4 to 140°F), 5 to 90%RH (no dew condensation),Vibration: 10 to 60 Hz, 4.9 m (16')/ S2 (0.5G or less),Impact: 392 m (1.3')/S2 (approx. 40G or less); Storage Condition: Ambient Temperature: -20 to 60°C (-4 to 140°F), 5 to 90% RH (no dew condensation),
Voltage Measurement Range ± 10 mV, ± 20 mV, ± 40 mV, ± 80 mV, ± 1.5 V, ± 2.5 V, ± 5 V, ± 10 V
Weight 15 kg
Width 400 mm
Country of Origin Japan
See All Specs Show Less
Hybrid Chart Recorder, 250 mm (10") RD5111 RD5111

Accessories

ZAILA DATA ANALYSIS SOFTWARE
€ 173,38 / each
0 In Stock
en-NL
en_NL
Save Configuration