Netherlands 07077 03815 | Intl +31 (0) 7077 03815 | Welcome to our new website, found an issue or bug? Please report it here
DPG409S
Sanitary, High Accuracy, Digital Pressure Gauge with Output
- Easy to Read 25 mm (1') Backlit LCD Display
- High 0.08% Accuracy Typical
- Hi and Lo Alarms Standard
- MIN/MAX/AVG Reading
Product Overview
- Accuracy Class < 0.1%
- Accuracy Detail ± 0.08% Best Straight Line Typ. (Linearity, Hysteresis and Repeatability combined)
- Battery life 9 mo
- Battery Size and Type 3.6V Li-ion
- Burst Pressure 6x Full Scale
- Cable Length 2.7 m
- Calibration Note 5-Pt NIST traceable Calibration with zero and span values, Calibrated in horizontal direction (Compound Gage models calibrated in positive direction only)
- Connection Location Lower
- Data Interface USB
- Data Interface Note USB connection for set-up, cable included
- Display Segment Height 25 mm
- Display Type 4 Digit LCD
- Display Type Note 4-digit LCD with user selectable backlighting, (9999 count) 25 mm (1') character height.
- Electrical Output 4 to 20 mA
- Electrical Output Note User selectable 0 to 5 Vdc, 0 to 10 Vdc of 4 to 20 mA (external 24 Vdc power required)
- Enclosure Material 316 Stainless Steel
- Enclosure Ratings IP65, NEMA Type 4X, Weatherproof
- Fill Fluid NEOBEE M5 food grade
- Isolation between Case and Output Terminations, Min 100MΩ @ 50 Vdc
- Materials, Wetted 316 Stainless Steel
- Number of Batteries 1
- Number of Digits 4
- Operating Temperature, Max 66 °C
- Operating Temperature, Min -18 °C
- Output Signal Current, Voltage
- Pressure Accuracy ± 0.08 %
- Pressure Cycles 250000
- Process Connection Note 11 ⁄2 and 2" Tri-GripTM Fitting (compatible with Tri-Clamp®): 600 psi using suitable clamp (600 psi on 11 ⁄2" Tri-Clamp® only)
- Process Connection Type Tri-Clamp
- Sample/Display Rate 0.25 to 10 sec
- Sample Display Rate Note User selectable from 0.25 to 10 sec. (preset to 1/sec)
- Sensor Dimension Note 72 to 89 mm L (2.8 to 3.5") below display
- Series ID DPG409S
- Settable Parameters Backlight, Min, Max, Zero, Clear
- Software Settings Software Settings (via USB) Update Rate: 4 options, 10, 5, 2, 1 times per second Units: psi, inHg, inH2O, bar, mbar, hPa units; must be specified at time of order Lock: Allows the lock-out of front panel functions Alarms: User selectable “High” and “Low” alarm limits (open collector) Backlight: On/off, 10 sec, 30 sec, 1 min, 5 min Wireless Transmitter (Optional): Channel number, transmission rate, alarms, sensor offset, chart recording, data logging Calibrate: Zero and Span Overrange Indication: “_ _ _ _”
- Span Setting ± 0.5% FS typical ± 1% maximum (For ranges ≤2.5 psi ± 1% Typical ± 2% maximum)
- Storage Temperature, Max 82 °C
- Storage Temperature, Min -40 °C
- Supply Power Type Battery
- Thermal Effects Thermal Effect (Over Compensated Range): Zero and Span shift over Compensated Range; Ranges: ≤2.5 psi: Zero: 1.00% Span: 0.70% ;5 psi: Zero: 0.80% Span: 0.50% ; > 5 psi: Zero: 0.50% Span: 0.50% ; 26B & 26HB Barometric Range: Zero: 0.90% Span: 0.50%
- Weight 900 g
- Weight Note 900 g (2 lb) Typical depending upon configuration
- Wireless Option Specifications Transmit Sample Rate: User programmable from 1 sample/2 min to 1 sample/2 sec Radio Frequency (RF) Transceiver Carrier: ISM 2.4 GHz RF Output: 10 dBm (10 mW) Range of RF Link: Outdoors (Line of Sight): Up to 120 m (400') Indoors/Urban: Up to 40 m (130') Included Software: Requires Windows® 7 (32 bit); XP, Vista Internal Battery: Two 3.6V lithium 8.4 Ah capacity (C-cell) included Data Transmission to Host: Pressure reading, ambient temperature reading, RF transmit strength and battery level
- Zero/Balance Error ± 0.5% FS typical ± 1% maximum (For ranges ≤2.5 psi ± 1% Typical ± 2% maximum)
Supported Applications
The Omega DPG409S Series is a Sanitary, High Accuracy Digital Pressure Gauge designed for wash-down and marine applications. The series incorporates integral sanitary pressure sensors with highly stable silicon wafer technology micro-machined to precision tolerances. Models are available in Gage, Absolute, Compound Gage, Vacuum (Negative Gage), and Barometric configurations.
Operating Conditions & Performance
The DPG409S series features a rugged 316 Stainless Steel enclosure with an ABS center gasket, rated Weatherproof NEMA 4X (IP65). The device operates within an ambient temperature range of -18 to 66°C (0 to 150°F) and supports storage temperatures from -40 to 82°C (-40 to 180°F).
Performance specifications include a typical accuracy of ±0.08% Best Straight Line Typ., combining linearity, hysteresis, and repeatability. For ranges ≤2.5 psi, the specification is ±0.5% FS typical with a maximum of ±1%. The display features 4-digit resolution (9999 count) with 25 mm (1") character height on a backlit LCD.
Pressure measurement capabilities span from low differential ranges up to high-pressure limits:
- Gage Pressure: Ranges include 0 to 10 inH2O, 0 to 1 psi, and extend up to 0 to 600 psi (using suitable clamp for 1½" Tri-Clamp®).
- Absolute Pressure: Available from 0 to 5 psi up to 0 to 600 psi.
- Vacuum/Negative Gage: Ranges include 0 to -10 inH2O and 0 to -15 psi.
- Compound Gage: Available from ±10 inH2O up to ±15 psi.
- Barometric Pressure: Models cover 0 to 32.00 inHg, 16 to 32 inHg, and metric ranges such as 0 to 1100 hPa.
The burst pressure rating is specified as 4 x rated pressure up to the maximum rated flange pressure. The response time is user-selectable from 0.25 to 10 sec (preset to 1/sec).
Configuration Options
The series offers extensive configuration options for process connections, outputs, and power:
- Process Connections: Sanitary models feature 1½ inch or 2.0 inch Tri-Grip™ fittings (compatible with Tri-Clamp®). The fitting size is indicated in the model number: "15" for 1.5 inch and "20" for 2.0 inch.
- Wetted Materials: Constructed with 316L SS wetted parts and filled with NEOBEE M5 Food Grade Fill Fluid.
- Analog Output Signals: User-selectable outputs include 0 to 5 Vdc, 0 to 10 Vdc, or 4 to 20 mA. Model suffixes indicate the signal type: "V1" for 0-5 Vdc, "V2" for 0-10 Vdc, and "MA" for 4-20 mA.
- Wireless Option: Appending a "-W" to the model number indicates a wireless version. Wireless models operate on ISM 2.4 GHz with an RF output of 10 dBm (10 mW). Range is up to 120 m (400') outdoors line-of-sight or up to 40 m (130') indoors/urban.
- Power Supply: Internal power uses two included 3.6V lithium C-cell batteries with 8.4 Ah capacity. External DC voltage source operation is also supported.
Key Product Differences
The DPG409S series distinguishes itself from the standard DPG409 and metric DPGM409 families through its sanitary design, Tri-Grip™ fittings, and food-grade fill fluid. Within the DPG409S line, variations are defined by:
- Measurement Type Codes: Model suffixes denote G (Gage), A (Absolute), V (Vacuum/Negative Gage), CG (Compound Gage), and B/HB (Barometric).
- Fitting Size: The numeric code "15" or "20" in the model number specifies the 1.5 inch or 2.0 inch Tri-Grip™ fitting size.
- Wireless Capability: Models with the "-W" suffix include a wireless transmitter option, requiring specific receivers (UWTC-REC1, UWTC-REC2 variants) for data logging and charting.
PDFs & Manuals