Skip to search Skip to main content Skip to navigation

Netherlands 07077 03815 | Intl +31 (0) 7077 03815 | Welcome to our new website, found an issue or bug? Please report it here

We noticed you are in , please visit your local site by clicking the below link for regional offers and support.

MCJ

Miniature Electronic Ice Point


Conversion of voltages from thermocouples to signals referenced at 0°C
Excellent stability of ± 1°C over the entire operating range of 10 to 45°C
Compatible with any standard offset connector or socket
Can be used in reverse mode to check the operation of thermometers and other instruments directly connectable with a thermocouple
Miniature Electronic Ice Point MCJ MCJ
loading all models
Product Overview
  • Accuracy ± 1 °C (± 1.8 °F)
  • Accuracy Detail 0.5°C from 15 to 30°C; 1°C from 10 to 45°C
  • Alloy Material (-) Iron
  • Alloy Material (+) Constantan
  • Color Options Black
  • Height 2.94 in
  • Impedance Output 200 Ω
  • Length 0.5 in
  • Operating Temperature, Max 45 °C
  • Operating Temperature, Min 10 °C
  • Reference Temperature 0 °C
  • Series ID MCJ
  • Storage Temperature, Max 75 °C
  • Storage Temperature, Min -25 °C
  • Storage Temperature Note -25 to 75°C (-13 to 167°F) (without battery)
  • Weight 57 g
  • Width 1 in

Supported Applications

The Omega MCJ Series Miniature Electronic Ice Point is a cold junction compensator designed to convert thermocouple voltages into 0°C referenced signals. It provides the electrical equivalent of an ice bath reference at 32°F (0°C) for use with digital multimeters (DMMs), voltage recorders, and other electronic devices that require temperature measurement via thermocouples.

The device is also utilized as a precision millivolt source to calibrate thermocouple meters. It supports nine standard thermocouple calibrations: J, K, T, E, R, S, G, C, and D.

Operating Conditions & Performance

The MCJ Series operates within an ambient temperature range of 10 to 45°C. Accuracy specifications are defined by the operating environment:

  • Accuracy: 0.5°C from 15 to 30°C; 1°C from 10 to 45°C.

The unit features a reference temperature setting of 0°C (32°F) and maintains high stability, performing better than ±0.2°C at any stabilized temperature over the full 10 to 50°C range for related series configurations. The output impedance is less than 250 Ω standard.

For storage without a battery, the device tolerates temperatures from -25 to 75°C (-13 to 167°F). The compact physical dimensions are 75 H x 25 W x 13 mm D (2.94 x 1 x 0.5"), with a weight of 57 g (< 2 oz).

Configuration Options

The MCJ Series is available in specific thermocouple calibration types, each identified by a distinct color code for quick identification:

  • J (Iron/Constantan): Black
  • K (Chromega®/Alomega®): Yellow
  • T (Copper/Constantan): Blue
  • E (Chromega®/Constantan): Violet
  • R and S: Green
  • C* (W-5% Re/W-26% Re):** Red
  • D* (W-3% Re/W-25% Re):** Red/white

The device connects via a standard thermocouple connector or jack and includes female-to-male adaptor pins. It is battery-powered, with replacement batteries available as the MCJ-BATT-B accessory.

Accessories and Replacement Parts

  • Battery: MCJ-BATT-B (Replacement battery)
  • Retro-fit Kits for Pre-1996 Units:
    • MCJ-BATT-R-A: For J, K, T, E calibrations
    • MCJ-BATT-R-B: For R, S calibrations
    • MCJ-BATT-R-D: For C, D calibrations

Note: Units manufactured prior to April 1995 may require retro-fit kits to replace mercury batteries. MCJs manufactured before 1996 use a different battery and compartment configuration.

Key Product Differences

The primary variation within the series is determined by the thermocouple calibration type (J, K, T, E, R/S, C, D), which dictates the internal compensation logic and color coding. The MCJ Series differs from the related CJ Series in that the MCJ is a miniature electronic ice point specifically designed for portability with dimensions of 75 H x 25 W x 13 mm D.

While the standard reference temperature setting is 0°C (32°F), notes indicate that CJs are available with reference junction temperatures other than 0°C, up to a maximum of 212°F, for an additional cost. The MCJ Series model numbers follow the format MCJ-, such as MCJ-J or MCJ-K.

The device is intended for use with instrumentation having an impedance of 50K ohms or greater to ensure accurate readings.