Netherlands 07077 03815 | International +31 (0) 7077 03815 | Welcome to our new website, found an issue or bug? Please report it here
PX1005
Product Specs
<< See all Product Models| Brand |
|
| Accuracy | ± 0.25 % |
| Accuracy Class | 0.1% to 0.25% |
| Accuracy Detail | Combined linearity, hysteresis and repeatability: ± 0.25% FSO, BSL [Note: Performance is determined at 25°C ± 1°C (77°F ± 2°F), open circuit, rated excitation, unless otherwise specified] |
| Applications | Aeronautics and Aerospace, Automotive, Fuel and Gas, Industrial and Manufacturing |
| Burst Pressure | 3x Full Scale |
| Burst Pressure Note | 3 times rated pressure |
| Compensated Temperature, Max | 27 °C |
| Compensated Temperature, Min | -196 °C |
| Electrical Connection | MIL-26482-I 10-6P |
| Electrical Output | 3 mV/V |
| Input Resistance | 400 Ω |
| Insulation Resistance | 100 MΩ |
| Materials, Wetted | 17-4 Stainless Steel |
| Mating Connector | MIL-26482-I 10-6S |
| Measurement Pressure Type | Sealed Gauge |
| Measurement Pressure, Min | 0 psi |
| Mounting | Case isolation ensures that the sensing element will not be affected by external stresses |
| Natural frequency | 35 kHz at 1000 psi decreasing logarithmically 5 kHz at 15 psi |
| Operating Temperature, Max | 149 °C |
| Operating Temperature, Min | -196 °C |
| Output Resistance | 400 Ω |
| Output Signal | Millivolt |
| Process Connection Note | Pressure Fitting: 7/16 - 20 Male per MS33656-4 |
| Process Connection Size | 1/4 in |
| Process Connection Type | AS5205/MS33656-4 Male |
| Proof Pressure | 2x Full Scale |
| Proof Pressure Note | 200% of rated pressure will not cause changes in performance beyond the specified tolerances |
| Relative Humidity | Per MIL-STD-810, method 507.1 |
| Residual Unbalance | 0 mV ± 5% FSO at zero pressure |
| Sensing Element | 4-active-arm Strain Gauge Bridge |
| Sensing Element Note | 4-active-arm bridge using sputter-deposited thin-film elements |
| Series ID | PX1005 |
| Shock | 100 g, 11 ms, half sine wave without calibration shift or damage |
| Shunt Calibration Value | Provisions for single-arm, external shunt calibration |
| Supply Voltage | 10 Vdc |
| Thermal Effect on Span error, Max | ± 0.01% FSO/°F |
| Thermal Effect on Zero error, Max | ± 0.01% FSO/°F, ± 0.01% FSO/°F |
| Vibration Sensitivity | 0.04% FSO/g |
| Vibration Sensitivity Note | At 35 g peak from 10 Hz to 2000 Hz (1⁄2” D.A.), the output shall not exceed 0.04% FSO/g for 15 psi units decreasing logarithmically to 0.003% FSO/g for 1000 psi units |
| Voltage Output, Min | 26 mV |
| Weight | 145 g |
| See All Specs | |
PDFs & Manuals