Need assistance? Call us: 1-800-663-4209
RD5100-Series
Universal Input 10 in (250 mm) Hybrid Chart Recorder
Registrador de gráficos híbrido de entrada universal de 10 in (250 mm)
Models In Stock
- High Speed Scanning at 36 Points/Sec and High-Speed Recording
- High Accuracy of 0.05%
- Various Industrial Values Can be Measured at the Same Time with Selectable Ranges
- Superior Ease of Operation
RD5100-Series
From
€
9.011,60
Models In Stock
Product Overview
- Accuracy ± 0.05 %
- Accuracy Detail Thermocouple K: ±0.05% ±0.5°C (±20 mV, ±40 mV), ±0.05% ±1°C (±80 mV), E: ±0.05% ±0.7°C (±20 mV, ±40 mV), ±0.05% ±1°C (±80 mV), J: ±0.05% ±0.7°C (±20 mV, ±40 mV), ±0.05% ±1°C (±80 mV), T: ±0.05% ±0.7°C (±20 Mv), R,B: ±0.05% ±1°C (±20 Mv), U,L: ±0.05% ±1°C, W-WRe26, PrRh40-PtRh20, NiMo-Ni: ±0.1% ±1°C, N: ±0.1% ±0.1°C, Platinel II: ±0.1% ±0.1°C (±20 Mv), ±0.1% ±1°C (±40 mV, ±80 mV), RTD: ±0.05% ±0.3°C, DC Voltage: ±0.05% ±1 digit
- Alarm Note Alarm Types: High limit, low limit, Alarm Setting Method: Individual setting for each point four levels/ channels
- Battery life up to 5 yr
- Battery Size and Type Li-ion
- Chart Length 20000 mm
- Chart Note Fan-fold type; overall width 318 mm (12.5"), total length 20 m (65.6'); effective recording width 250 mm (10") (analog recording)
- Chart Speed 1 to 1,500 mm/h
- Chart Width 250 mm
- CMRR 130 dB
- Common Mode Voltage 30 Vac
- Cutout Dimensions 388 x 248 mm
- Dielectric Strength Between Primary Terminals and Protective Conductor Terminals: 1 minute at 1500 Vac, Between Secondary Terminals and Protective Conductor Terminals:1 minute at 500 Vac, Between Primary Terminals and Secondary Terminals: 1 minute at 1500 Vac
- Display Note Color LCD panel RGB (640 x 240 dot), Size: 149.8 W x 57.4 mm H, Setting Display: Common to digital display Display Contents: Digital display, Channel Display: One-point/multiple points continuous/sequential indication change, Display Measuring Value of Each Channel: One-point/multiple points continuous/sequential indication change, Clock Display: Hour/Minute/ Second/Tag/Unit. Alarm Display: Occurrence CH No, data is displayed in red when alarm occurs
- Display Type LCD
- Electrical Connection Note Terminal Board: Detachable type, removable for wire connection
- Height 260 mm
- Input Note Input Correction: Zero/span correction and shift correction for each channel
- Input Resistance 1 MΩ
- Insulation Resistance Between Primary Terminals and Protective Conductor Terminals:20MΩ or more at 500 Vdc, Between Secondary Terminals and Protective Conductor Terminals: 20MΩ or more at 500 Vdc, Between Primary Terminals and Secondary Terminals: 20MΩ or more at 500 Vdc
- LED Indicators Record On: Lights during recording; LED, Key Lock: Lights during key lock, Alarm: Lights during alarm activated; LED, Chart End: Lights just before record ending, Fail: Lights during unit abnormal time
- Length 300 mm
- Loop Resistance, Max Thermocouple Inputs, DC Voltage Input (10 mV): 500Ω or less (without burnout), DC Voltage Input (Except 10 mV): 100Ω or less Resistance, Thermometer Inputs: 10Ω or less/ line, three lines are common, Pt100, JPt100
- Materials ABS, Soda glass, Steel
- Measurement Range Note Thermocouple: K: Reference Range: ±20 mV: -200 to 500°C, ±40 mV: -200 to 900°C, ±80 mV: -200 to 1370°C, E: Reference Range: ±20 mV: - -200 to 250°C, ±40 mV: -200 to 500°C, ±80 mV: -200 to 900°C, J : Reference Range: ±20 mV: - -200 to 350°C, ±40 mV: -200 to 700°C, ±80 mV: -200 to 1200°C, T: Reference Range: ±20 mV: - -200 to 400°C, R: Reference Range: ±20 mV: 0 to 1760°C, B: Reference Range: ±20 mV: 0 to 1300°C, N : Reference Range: ±20 mV: 0 to 600°C, ±40 mV: 0 to 1000°C, ±80 mV: 0 to 1300°C, W-WRe26: Reference Range: ±80 mV: 0 to 2315°C, W-WRe26: Reference Range: ±80 mV: 0 to 2315°C, PtRh20-PrRh40: Reference Range: ±20 mV: 0 to 1888°C, U : Reference Range: ±20 mV: -200 to 350°C, ±40 mV: -200 to 600°C, L : Reference Range: ±20 mV: -200 to 350°C, ±40 mV: -200 to 700°C, ±80 mV: -200 to 900°C, Nimo-Ni: Reference Range: ±80 mV: -50 to 1310°C, Platinel II: Reference Range: ±20 mV: 0 to 500°C, ±80 mV: 0 to 950°C, 0 to 1395°C RTD: Reference Range: 50Ω: -50 to 50°C , 100Ω: -100 to 130°C, 200Ω: -200 to 250°C, 300Ω: -200 to 550°C Voltage: ±10 mV, ±20 mV, ±40 mV, ±80 mV, ±1.5 V, ±2.5 V, ±5 V, ±10 V
- Measurement Type Thermocouple, RTD, Voltage
- Mounting Panel Mount
- NMRR 50 dB
- Operating Environment Reference Operating Condition: Ambient Temperature/Humidity Range: 21 to 25 °C (70 to 77°F), 45 to 65% RH,Power Voltage: 90 to 264V,Power Frequency: 50/60Hz ±2%,Attitude: Forward/Backward/left/right within 3°,Warm-Up Time: 1 hour or longer; Normal Operating Condition: Ambient Temperature/Humidity ,Range: 0 to 40 °C (32 to 104°F), 20 to 80% RH,Power Voltage: 90 to 264V,Power Frequency: 50/60Hz ±2%,Attitude: Forward/backward/left/right within 3°
- Operating Temperature, Max 40 °C
- Operating Temperature, Min 0 °C
- Power Consumption 100 VA
- Recording Color Analog Recording: Red, Purple-Red, Orange, Brown, Green, Yellow-Green, Blue-Green, Purple, Purple-Blue, black, Digital Recording and Logging: Black
- Recording Note Recording System: Raster scan system, 10-color wire dot printing; Compensation: Independent setting of zero spans are available, Operation: Record On/Off: Recording operation ON/OFF*, Data Print: Printing measuring data*, Feed: Fast-forwarding chart paper
- Relative Humidity 20 to 80% RH
- Resolution Input: 1/40,000; Display: Thermocouple, RTD: 0.1°C, DC Voltage: 1 µV(±10 mV, ±20 mV), 10 µV(±40 mV, ±80 mV), 100 µV(±1.25V, ±2.5V), 1 mV(±5V, ±10V)
- RTD Type 100 Ω pt or Jpt sensor; 3-wire Pt100
- Sampling Rate 0.1 sec/channel
- Sensor Break Detection Thermocouple Burn-Out: Select with/without burnout for each input
- Series ID RD5100-Series
- Storage Temperature, Max 60 °C
- Storage Temperature, Min -20 °C
- Supply Power Note Power Voltage: 90 to 264V, Power Frequency: 50/60Hz ±2%
- Supply Power Type AC
- Supply Voltage 100 to 240 Vac
- Switch Rating 3 A @ 100 to 240 Vac
- Switch Rating Note 100 to 240 Vac, 3 A resistance load
- Switch Type SPST NO
- Temperature Drift 0.1%/10°C
- Temperature Measurement Range K: -200 to 1370°C, E: -200 to 900°C, J: -200 to 1200°C, T: -200 to 400°C, R: 0 to 1760°C, B: 0 to 1300°C, N: 0 to 1300°C, W-WRe26: 0 to 2315°C, B: 0 to 1888°C, L: -200 to 900°C, Nimo-Ni: -50 to 1310°C, Platinel II: 0 to 1395°C, RTD: -200 to 550°C
- Thermocouple Type J, K, T, E, N, R, S, U, B, G, L, P, NiMo-Ni, PrRh40-PtRh20
- Transportation and Storage Details Transportation Condition: Ambient Temperature/Humidity ,Range: -20 to 60°C (-4 to 140°F), 5 to 90%RH (no dew condensation),Vibration: 10 to 60 Hz, 4.9 m (16')/ S2 (0.5G or less),Impact: 392 m (1.3')/S2 (approx. 40G or less); Storage Condition: Ambient Temperature: -20 to 60°C (-4 to 140°F), 5 to 90% RH (no dew condensation),
- Voltage Measurement Range ± 10 mV, ± 20 mV, ± 40 mV, ± 80 mV, ± 1.5 V, ± 2.5 V, ± 5 V, ± 10 V
- Weight 15 kg
- Width 400 mm
Los registradores de gráficos de la serie RD5100 son registradores híbridos de 250 mm (10") con entrada de rango múltiple. El registrador de alto rendimiento con diseño innovador proporciona alta precisión, escaneo de alta velocidad ±0,05%, 0,1 segundos para 36 puntos y grabación de alta velocidad, 1 línea en 3 segundos. Teclas de operación simples y funciones de configuración para PC mejoran drásticamente la usabilidad del sistema de registro.
PDFs & Manuals
Accessories
accessories
consumables
Z-fold chart paper 250 mm x 20 m (9.8 in x 65.6 ft), case of 15
TWD25,888 / each
7 In Stock