High Precision Calibration-Temperature Reference Chamber

고정밀 교정용 온도 기준 챔버
  • A 0°C (32°F) Thermoelectric “Refrigerator”
  • Eliminates Old Fashioned “Ice Bath”
  • For Versatile Use in the Factory, Laboratory, Instrument Shop
  • Calibration of All Temperature Instruments and Sensors
TRCIII-Series
From 8,229,278 원
고정밀 교정용 온도 기준 챔버 TRCIII-Series TRCIII-Series
loading all models
Product Overview
  • 정확도 ± 0.1 °C (± 0.18 °F)
  • 디스플레이 해상도 なし
  • 폼팩터 ポータブル・ベンチトップ
  • 빈도 60 Hz
  • 높이 305 mm
  • 길이 322 mm
  • 참조 우물 수 6
  • 사양 참고 우물 수 사양 참고 6ウェル、内径5⁄32インチ、深さ33⁄4インチ;ウェルは互いに熱的および電気的に接地されている;外径1⁄8インチのプローブ対応
  • 작동 온도, 최대 32 °C
  • 최소 작동 온도 2 °C
  • 기준 온도 0 °C
  • 기준 웰 깊이 3.75 in
  • 기준 웰 내경 0.1563 in
  • 시리즈 ID TRCIII-Series
  • Stability ± 0.04 °C
  • Stabilization Time 周囲温度および熱負荷により2時間または3時間
  • Supply Power Note 115/220 V、60 Hz、100 ワット
  • Weight 9.5 kg
  • Width 159 mm

Supported Applications

The Omega TRCIII Series is a high precision ice point™ Calibration Reference Chamber designed to maintain a stable reference temperature of 0°C (32°F) using freezing water. The series serves as an on-site calibration standard for thermocouple, RTD, and thermistor probes.

Operating Conditions & Performance

The TRCIII Series operates within an ambient temperature range of 2 to 32°C (35 to 90°F). The unit maintains a process reference point at 0°C (32°F) with accuracy values stated as ±0.1°C or better. Media compatibility is limited to distilled, deionized water; the unit must not be allowed to freeze completely, as damage will result.

Configuration Options

The series offers configurable power supply options and optional accessories for expanded calibration workflows:

  • Voltage Models: TRCIII (120 Vac operation) and TRCIII-220 (220 Vac operation).
  • Power Requirements: 115/220 V, 60 Hz, 100 Watts.
  • Physical Dimensions: 305 H x 159 W x 322 mm D (12.0 x 6.25 x 12.7").
  • Weight: 9.5 kg (21 lb).

The series supports the following optional configurations and accessories:

  • NIST Traceable Calibration Available.
  • Companion Dry Block Probe Calibrator (Model CL900/950).
  • TRP reference probes including TRP-K, TRP-J, TRP-T, TRP-E, TRP-C, TRP-D, TRP-B, TRP-S, and TRP-R.
  • OMEGACARE™ Extended Warranty Program (extends standard 1-year warranty to a total of 4 years).

The unit is supplied with male and female color-coded thermocouple connectors. White-coded Copper-Copper connectors are specified for electrical connection.

Key Product Differences

Variants within the TRCIII Series differ primarily by input voltage:

  • TRCIII: ice point™ reference 120 Vac operation.
  • TRCIII-220: ice point™ reference 220 Vac operation.

The CAL-3 option provides NIST-traceable calibration. TRP reference probes are available in standard lengths of 12" for Types J, K, T, and E, or 6" for Types R, S, B, C, G, and D; longer leads (e.g., -24, -36) are configurable at additional cost.