Taiwan 0809 090 375 | Intl +65 6415 5353 | Welcome to our new site - found a bug? Please report it here
LCHD
Ultra Low Profile, Tension and Compression Load Cells
Models In Stock
- 5 Point NIST Traceable Calibration, Certificate Supplied
- 0.25% Interchangeability for Scale Applications
- All Stainless Steel Construction
- Rugged Heavy Duty Design
LCHD
From
TWD
54,417
Models In Stock
Product Overview
- Additional Features 1 IEEE 1451.4 module
- Calibration Option 5 POINT NIST
- Input Resistance 350 ± 10 Ω
- Interchangeability 0.0025
- Load Cell Type Pancake
- Materials 17-4 PH Stainless Steel
- Non-Linearity 0.001
- Operating Temperature, Max 75 °C
- Operating Temperature, Min -40 °C
- Output Resistance 350 ± 10 Ω
- Process Temperature, Max 71 °C
- Process Temperature, Min 16 °C
- Safe Overload 150% FS
- Series ID LCHD
- Supply Voltage 10 Vdc, 15 V max, 10 to 28 Vdc
- Thermal Span Effect ± 0.009% rdg/°C
- Thermal Zero Effect ± 0.009% FSO/°C
- Trim Pot Access None
- Zero/Balance Error ±1% of FSO
Supported Applications
The Omega LCHD Series consists of ultra low profile, tension and compression load cells designed for force measurement in severe environments. These pancake-style sensors utilize high-quality strain gages and precision techniques to ensure long-term stability. The series is calibrated in tension with capacities ranging from 5 lb (2.3 kgf) up to 500,000 lb (10,000 kgf). Designed for mounting via through holes and a threaded center shaft, the LCHD Series incorporates two stabilizing diaphragms welded to the sensing member to reduce off-center and side-load effects.
Operating Conditions & Performance
The series operates within an ambient temperature range of -65°F to 250°F (-54°C to 121°C), with a compensated temperature range of 60°F to 160°F (16°C to 71°C). Thermal effects are specified as Zero: 0.0036% FSO/°C and Span: 0.0036% FSO/°C. Performance specifications vary by capacity:
- Linearity: ±0.25% FSO for capacities ≤25 lb (10 kgf); ±0.10% FSO for ≥50 lb (20 kgf).
- Hysteresis: ±0.10% FSO for ≤25 lb; ±0.08% FSO for ≥50 lb.
- Repeatability: ±0.10% FSO for ≤25 lb; ±0.03% FSO for ≥50 lb.
The load cells feature a 350 Ω bonded foil strain gage bridge with an excitation voltage of 10 Vdc (maximum 15 V). Safe overload is rated at 150% of capacity, and ultimate overload is 300%. Full-scale deflection typically ranges from 0.025 to 0.076 mm (0.001" to 0.003").
Configuration Options
The LCHD Series offers a broad envelope of configurations including standard imperial models and metric variants (LCMHD). Key configurable dimensions include overall diameter, height, and threaded shaft sizes that scale with capacity:
- Capacities: 5 lb to 500,000 lb (2.3 kgf to 10,000 kgf).
- Housing Materials: 17-4 PH stainless steel for all models; 4340 alloy steel is noted in the manual for capacities ≤200 k.
- Electrical Connections: Includes a mating connector. Units up to 5,000 lb use PT06F-10-6S (or DT02H-10-6P); units from 7.5k to 500k use MS3102E-14S-6P.
- Output Signals: Standard output is 2 mV/V for ≤25 lb and 3 mV/V for ≥50 lb, with accuracy ±0.25% FSO.
The series supports several optional suffixes to modify functionality:
- .5E: 5 V Output (calibrated 0 to 5 Vdc in tension; bi-directional ±5 Vdc).
- -DUAL: Dual Bridge configuration.
- .TEDS: IEEE 1451.4 module support.
- -TC1 / -TC2: Extended temperature range options (-45 to 93°C and up to 163°C).
- .OL: Overload stops included.
Key Product Differences
The series distinguishes between standard imperial models (LCHD) and metric models (LCMHD), with specific model numbers corresponding to capacity ranges. For example, LCHD-5K represents a 5,000 lb unit, while LCMHD-10K corresponds to a 10,000 kgf unit. Tension plates are available as accessories for specific ranges (e.g., LCHD-TP225 for 50–1,000 lb). Models exceeding 300,000 lb capacity are calibrated at the 300,000 lb point.
Electrical pinouts follow a standard configuration: Pins A & B (+) Excitation; C & D (-) Excitation; E (-) Output; F (+) Output. The series includes NIST-traceable calibration and comes complete with the mating connector and 59 kΩ shunt data.