M12TXSS

Stainless Steel RTD probes with 4 to 20 mA analog output & Mounting Threads

Models In Stock
  • Programmable Transmitter
  • M12 Connection
  • 4 to 20 mA Output = 0 to 100°C Default Setting
  • -50 to 120°C (-58 to 248°F) Full Programmable Range
M12TXSS
From TWD 6,116
Models In Stock
Stainless Steel RTD probes with 4 to 20 mA analog output & Mounting Threads M12TXSS M12TXSS
loading all models
Product Overview
  • Accuracy Class A
  • Accuracy Detail Class A per IEC60751
  • Body Material Stainless steel AISI 316L
  • Downscale 3.6 mA
  • Electrical Output 4 to 20 mA
  • EMC In accordance to EN 61326
  • Enclosure Ratings IP67
  • Enclosure Ratings Note IP65 and IP67 in accordance to IEC60529
  • Factory Setting 0 to 100°C/ sensor break >21 mA (upscale)
  • Operating Temperature, Max 80 °C
  • Operating Temperature, Min -40 °C
  • Permissible Load 700
  • Permissible Load Note 700 @ 24 Vdc [RLo= (Vsupply-8.5)/0.020]
  • Probe Diameter 3 mm
  • Process Temperature, Max 120 °C
  • Process Temperature, Min -50 °C
  • Range Configurations Note It is possible set the input temperature range (span) by the M12TX-CONFIG-SETconfiguration kit (PC with OS Windows® required)
  • Relative Humidity, Max 1
  • Relative Humidity, Min 0%
  • Response Time < 50 ms
  • Response Time Note Diameter 3 mm (0.12") <3.5 seconds, diameter 6 mm (0.24") <13 seconds (test in water to IEC751- time for reaching 63.2% of instantaneous temperature change)
  • RTD Type 100 Ω Pt Sensor
  • Sensor Break Monitoring Note Selectable: Upscale (>21.0 mA) or downscale (<3.6 mA) action Sensor Short-Circuit: Fixed to downscale (<3.6 mA) action
  • Sensor Error Compensation Note Over 2 points (maximum 1% of span)
  • Series ID M12TXSS
  • Span, Min 50°C
  • Span Note 50°C (122°F) [if the zero value is set between one of these values: -40°C (-40°F), -20°C (-4°F), 0°C (32°F), 20°C (68°F), 40°C (104°F), the minimum span is 20°C (68°F) rather than 50°C (122°F)]
  • Supply Power Effects Negligible
  • Supply Voltage 8.5 to 32 Vdc
  • Temperature Influence, Max ± 0.3%
  • Temperature Influence Note Maximum of ± 0.3°C/25°C (77°F) or ± 0.3% of span/25°C (77°F)
  • Temperature Sensor Type RTD
  • Thermocouple Type N/A
  • Tmax. Electronic 80°C
  • Upscale 21 mA
  • Zero Setting Adjustable Range Any value between -50 to 50°C (-58 to 122°F)

Supported Applications

The Omega M12TXSS series consists of stainless steel RTD temperature transmitters designed for compact installations where space is limited. These devices integrate a Pt100 sensor with an analog transmitter to provide direct process monitoring in industrial environments requiring secure, threaded mounting.

  • Temperature measurement and signal conditioning via 4 to 20 mA output
  • Applications utilizing M12 male connector interfaces for electrical connection

Operating Conditions & Performance

The series is engineered for a process temperature range of -50 to +120°C (-58 to 248°F). The electronic section operates within an ambient temperature range of -40 to +80°C. Accuracy specifications include Class A sensor performance in accordance with IEC751 and a transmitter accuracy maximum of ±0.2°C or ±0.2% of span.

  • Response time: <3.5 seconds for the 3 mm diameter probe (tested in water to IEC751, reaching 63.2% of thermal gap)
  • Enclosure rating: IP65 and IP67 in accordance with IEC60529
  • Humidity tolerance: 0 to 100% relative humidity

Configuration Options

The M12TXSS series offers configurable parameters via the M12TX-CONFIG-SET kit, which includes a USB interface and software for Windows®. The transmitter supports input temperature range configuration between -50°C to 800°C (span), sensor break monitoring selection (upscale or downscale action), and measurement unit settings.

  • Probe lengths: 13 mm (0.51") or 24 mm (0.94"); diameter is fixed at 3 mm (0.1")
  • Housing material: Stainless steel AISI 316L with a thermowell construction
  • Process connections: Standard 1/8" NPT male thread or G1/8 metric thread options
  • Electrical output: 4 to 20 mA signal; permissible load is 700Ω @ 24 Vdc (calculated as R_load = (U - 8.5) / 0.022)
  • Power supply: 8.5 + 32 Vdc with polarity protection

Key Product Differences

The series is distinguished by probe insertion length and thread standard, resulting in specific model configurations:

  • M12TXSS-PT100-13MM variants: 13 mm (0.51") long with either 1/8" NPT or G1/8 threads
  • M12TXSS-PT100-24MM variants: 24 mm (0.94") long with either 1/8" NPT or G1/8 threads

Factory default parameters are set to a range of 0+150°C for most models, though the M12TXSS series defaults to 0+100°C (32 to 212°F) with upscale sensor break indication (>21 mA). The device is non-isolated and features an M12 male connector output.

Accessories

accessories

CONFIGURATION KIT FOR M12TX SENSORS
TWD17,249 / each
0 In Stock
CONF RTD PROBE W/M12 CONN
/ each
0 In Stock
This item is discontinued. Please contact Technical Sales Team for assistance.

cables

M12 Cables with Field Mountable Connectors for Resistance Sensors (RTDs)
TWD4,054 / each
4 In Stock
M12 Cables with Field Mountable Connectors for Resistance Sensors (RTDs)
TWD2,417 / each
20 In Stock
en-TW
en_TW
Save Configuration