Skip to search Skip to main content Skip to navigation

Taiwan 0809 090 375 | Intl +65 6415 5353 | Welcome to our new site - found a bug? Please report it here

We noticed you are in , please visit your local site by clicking the below link for regional offers and support.

PHCN-85-Controller

pH/ORP Transmitter and Controller with TFT color display

  • The PHCN-85 is being replaced by the AM-2250 (controller) / AM-2250TX (transmitter)
  • 3.5', 320 × 240 TFT Color Display, Virtual Pointer or Digital Style, Multi-Parameter
  • Calibration with Automatic Buffer Recognition
  • Automatic or Manual Temperature Compensation
PHCN-85-Controller
From
pH/ORP Transmitter and Controller with TFT color display PHCN-85-Controller PHCN-85-Controller
loading all models
Product Overview
  • Cutout Dimensions 91 × 91 mm
  • Data Interface RS485
  • Display Type TFT
  • Display Type Note 3.5", 320 × 240 TFT Color Display, Virtual Pointer or Digital Style, Multi-Parameter
  • Enclosure Ratings IP65, NEMA Type 4X
  • Height 96 mm
  • Length 94 mm
  • Measurement Resolution pH: 0.01, ORP: mV, 0.1 mV
  • Mounting Orientation Panel Mount
  • Number of Relays None
  • Operating Humidity 0 to 85% RH
  • Operating Temperature, Max 50 °C
  • Operating Temperature, Min 0 °C
  • ORP Accuracy ± 5 mV
  • ORP Measurement Range -999 to +999 mV
  • pH Accuracy ± 0.1 pH
  • PH Measurement Range 0 to 14.00 pH
  • Power Consumption 5.5 W
  • Process Temperature, Max 0 °C
  • Process Temperature, Min 100 °C
  • Relay Note Contact photoelectric relay output, programmable; temperature, pH/ORP output can be set for pH/ORP/temperature or timing ON/OFF or PID control Relay: Resistance load 50 mA maximum AC/DC 30V
  • Series ID PHCN-85-Controller
  • Storage Temperature, Max 60 °C
  • Storage Temperature, Min -20 °C
  • Temperature Accuracy ± 0.5 °C
  • Temperature Resolution 0.1 °C
  • Temperature Sensor Type 1000 Ω RTD
  • Width 96 mm

Supported Applications

The Omega PHCN-85 is a panel-mount pH/ORP transmitter and controller designed for high-level water quality management, automatic running systems, and inline monitoring. Documented applications include:

  • High Purity Water: Management control of high pure or ultra-pure water in electronics, electric power, pharmaceuticals, fine chemicals, clinical medical science, and life-science research.
  • Industrial Process Solutions: Salinity analysis for water treatment, desalination, concentrate management, circulating cooling water, industrial coating, large-scale water treatment equipment, mixing, metallurgy, petrochemicals, and industrial cleaning.

Operating Conditions & Performance

The PHCN-85 measures pH from 0.00 to 14.00 with a resolution of 0.01 and an accuracy of ±0.1. ORP measurement ranges from -1999 to +1999 mV with a resolution of 0.1 mV and display accuracy of ±5 mV. Temperature is measured from 0 to 100°C (32 to 212°F) using a 1000 Ω RTD input, offering a resolution of 0.1°C and an accuracy of ±0.5°C.

Automatic temperature compensation functions within the range of 0 to 50°C (32 to 122°F). The unit operates in ambient temperatures from 0 to 50°C (32 to 122°F) with relative humidity ≤85% RH (non-condensing). Storage conditions are specified as -20 to 60°C (-4 to 140°F) at ≤85% RH. The enclosure provides a protection level of NEMA 4 (IP65) when the rear cover is installed.

Electrical and Output Specifications

  • Power Supply: DC 24V ±4V with internal automatic polarity identification; consumption <5.5 W.
  • Current Outputs (2): Isolated, adjustable, reversible 4 to 20 mA outputs supporting instrument or transmitter mode. Loop resistance is maximum 400 Ω at DC 24V with a resolution of ±0.1 mA.
  • Relay Outputs (3): Contact photoelectric relay outputs programmable for pH, ORP, temperature, timing ON/OFF, or PID control. Relay load is maximum 50 mA AC/DC at 30V.
  • Data Communication: RS485 Modbus-RTU protocol.

Configuration Options

The PHCN-85 features a 3.5 inch 320×240 TFT color screen with virtual pointer or digital style interface display and multi-parameter capability. Calibration supports offline/inline selection with automatic buffer recognition for up to 6-point buffer solution calibration.

Configurable Outputs

The three photoelectric switch outputs can be configured as:

  • pH, ORP, or temperature output control.
  • Timing output (calendar function with reserved timing).
  • PID adjustment to drive pulse metering pumps or extension controls.

Accessories and Options

The series supports the following accessories for specific installation needs:
  • pH Electrodes: PHE-7351-15-PT1K (Industrial pH electrode with temperature, in-line or submersible) and PHE-6510B (Submersible pH electrode).
  • ORP Electrode: ORE-6510B (Submersible ORP electrode).
  • Mounting Assembly: PHEH-65-10 (Required for ORE-6510B or PHE-6510B).
  • Buffer Solutions: PHA-4, PHA-7, and PHA-10 (pH 4.00, 7.00, and 10.00 respectively; 500 mL bottles).

Key Product Differences

The series is offered as a single model configuration: the PHCN-85 panel-mount pH/ORP transmitter and controller. The unit includes EMC design for anti-interference, password protection, and status check menus to view current transmitting status or mV values. Dimensions are 96 H × 96 W × 94 D mm (3.78 × 3.78 × 3.7") with a panel cutout of 91 × 91 mm (3.58 × 3.58"). The device carries CE approval and includes a one-year warranty.

Accessories

cables

MOUNTING CABLE ASSEMBLY
TWD4,665 / each
1 In Stock

consumables

10.00 pH buffer solution 500 mL (1 pint) bottle
TWD814 / each
68 In Stock
4.00 pH buffer solution 500 mL (1 pint) bottle
TWD739 / each
75 In Stock
7.00 pH buffer solution 500 mL (1 pint) bottle
TWD739 / each
53 In Stock

installation hardware

BNC TO TERMINAL ADAPTER FOR PHTX-100
TWD1,307 / each
14 In Stock
TEMP. COMP PT1K ASSEMBLY (*)
TWD12,379 / each
0 In Stock

sensors

Flat Surface In Line pH/ORP Electrodes
TWD10,061 / each
58 In Stock
Submersible Flat Surface pH/ORP Electrodes
TWD9,701 / each
27 In Stock
en-TW
en_TW
Save Configuration