Passer à la recherche Passer au contenu principal Passer à la navigation

Besoin d'aide? Appelez-nous: 1-800-663-4209

Nous avons remarqué que vous êtes dans , veuillez visiter votre site local en cliquant sur le lien ci-dessous pour des offres et un support régionaux.

CL515-PLUS-series

Calibrateur et thermomètre à affichage auto-connecté RTD de haute précision

High Accuracy RTD Auto-Connect Display Calibrator & Thermometer
  • Easy to Use
  • Calibrate and Measure Most RTD Sensors
  • Comes Protected With a Rugged Rubber Boot
  • Calibrate Directly in Temperature (ºC or ºF)
CL515-PLUS-series
À partir de TWD 96,249
High Accuracy RTD Auto-Connect Display Calibrator & Thermometer CL515-PLUS-series CL515-PLUS-series
loading all models
Aperçu du produit
  • Accuracy ± 0.015 %
  • Accuracy Detail ± (0.015% of reading + 0.05 Ω)
  • Additional Features 1 Source Only
  • Battery life 50 hr
  • Display Type LCD
  • Display Type Note High contrast graphic liquid crystal display, LED backlighting for use in low lit areas
  • Height 41 mm
  • Length 143 mm
  • Operating Temperature, Min -25 °C
  • Operating Temperature Note Nominal Temperature: 23°C
  • Relative Humidity, Max 0.9
  • Relative Humidity, Min 10%
  • Relative Humidity Note 10% ≤RH ≤90% (0 to 35°C), non-condensing 10% ≤RH≤ 70% (35 to 60°C), non-condensing, Nominal Relative Humidity: 70%
  • Series ID CL515-PLUS-series
  • Supply Voltage 6 Vdc
  • Temperature Drift ± 0.01%
  • Temperature Drift Note ± 0.01% of span outside of 23°C ± 10°C (73°C ± 18°F)
  • Thermocouple Type N/A
  • Type of RTD Signal 100 Ω
  • Voltage Read Excitation: Current 0.5 mA nominal Normal Mode Rejection: 50/60 Hz, 50 dB Common Mode Rejection: 50/60 Hz, 120 dB
  • Voltage Source Accuracy: From 1 to 10.2 mA external excitation Current ± (0.015% of Reading + 0.05 Ω) Below: 1 mA of external excitation Current: ± (0.015% of reading + 0.025 mV/mA excitation current + 0.05 Ω) Resistance Ranges: 0.00 to 410.00, 410.1 to 4001.0 Ω Allowable Excitation Current Range <41Ω: 10.2 mA maximum; steady or pulsed/intermittent 410 to 4001Ω: 1 mA maximum; steady or pulsed/intermittent Pulsed Excitation: Current compatibility DC to 0.01 second pulse width 0.025 mV
  • Weight 0.34 kg
  • Weight Note 0.34 kg (12.1 oz) including boot and batteries
  • Width 76 mm

Supported Applications

The Omega CL515-PLUS series is a handheld device designed for checking, calibrating, and measuring Resistance Temperature Detector (RTD) sensors. The unit functions as an automated RTD calibrator capable of simulating resistance outputs to verify system performance.

Documented applications include compatibility with smart transmitters, PLCs, DCS systems, and multichannel recorders for calibration verification tasks.

Operating Conditions & Performance

The CL515-PLUS series provides RTD measurement accuracy of ±0.05 °C (±0.09 °F) and Ohms measurement accuracy of ±0.01 Ω. Resolution is specified as 0.1 °C, 0.1 °F, or 0.01 Ω.

Measurement ranges cover:

  • RTD Temperature: -200 to 850 °C and -328 to 1562 °F
  • Ohms (Source/Read): 0 to 400 Ω and 0 to 2000 Ω

The device supports automatic output stepping with selectable step sizes of 2, 3, 5, 11, or 21 steps. Step times are configurable at 5, 6, 7, 8, 9, 10, 15, 20, 25, 30, or 60 seconds.

Configuration Options

The series features a rugged design with a protective rubber boot and a backlit display. It automatically detects electrical connections for RTD sensors configured in 2-wire, 3-wire, or 4-wire arrangements.

Sensor Types Supported:

  • Platinum 100 (α=3850, α=3902, α=3926)
  • Platinum 1000 (α=3850, α=3750)
  • Copper 10 & 50 Ohm
  • Nickel 120 Ohm

Output and Display Options:

  • Outputs in °C, °F, or ohms (simulated resistance)
  • Automatic indication of connection type
  • Battery symbol on display indicates low battery status

Power and Configuration:

  • Powered by 4 x AA 1.5V Alkaline batteries (supplied)
  • Configuration menu includes Auto Off, Backlight control, Steps selection, and Step Time settings
  • Removing the batteries resets all stored configuration options to factory defaults