Taiwan 0809 090 375 | Intl +65 6415 5353 | Welcome to our new site - found a bug? Please report it here
OM-CP-PROCESS101A-Series
DC Current Data Logger
Models In Stock
- USB Cable and Software (Model OM-CP-IFC200) Sold Separately and Required to Program the Unit
- 10 Year Battery Life
- 4 Hz Reading Rate
- Multiple Start/Stop Function
OM-CP-PROCESS101A-Series
From
TWD
9,400
Models In Stock
Product Overview
- Alarm Programmable high and low limits; alarm is activated when current reaches or exceeds set limits
- Baud Rate 115200
- Calibration Accuracy ±0.05%
- Calibration Note Digital calibration through software, automatically recorded within device
- Conversion Time 133 ms
- Data Interface USB
- Electrical Connection Removable Screw Terminal
- Frequency Rejection 50 to 60 Hz
- Height 1.4 in
- Housing Depth 1.5 in
- Housing Height 2.9 in
- Housing Width 5.8 in
- Input Voltage 0 to 2.5 V
- LED Functionality Green LED Blinks: 10 second rate to indicate logging 15 second rate to indicate delay start mode;Red LED Blinks: 10 second rate to indicate low battery and/or full memory
- Length 0.6 in
- Materials Aluminum, ABS
- Measurement Type Current
- Memory 1,000,000 readings
- Memory Modes Yes
- Memory Note 1,000,000 readings; software configurable memory wrap 333,000 readings in multiple start/ stop mode
- Number of Channels 1
- Operating Temperature, Max 80 °C
- Operating Temperature, Min -40 °C
- Programmable Functions Security
- Real Time Recording The device may be used with PC to monitor and record data in real-time
- Relative Humidity 0 to 95% RH
- Sampling Rate Note 4 Hz to 1 reading every 24 hours
- Series ID OM-CP-PROCESS101A-Series
- Software XP SP3/Vista/7 and 8
- Start Modes Button Push, Delayed Start, Immediate Start
- Supply Power 3.6 V Lithium Battery
- Supply Power Note 10 Year Battery Life Typical, at a 15 Minute Reading Rate
- Supply Power Type Battery
- Temperature Coefficient < ±50 ppm/ºC
- Time Accuracy ±1 minute/month (stand alone data logging)
- Voltage Measurement Range N/A
- Weight 24 g
- Width 2.2 in
Supported Applications
The Omega OM-CP-PROCESS101A Series is a battery-powered DC current data logger designed for measuring and recording direct current signals. As part of the NOMAD® family, these devices capture process variables at user-specified intervals ranging from once every 2 seconds to once every 12 hours. The series supports real-time monitoring and recording when connected to a PC via USB.
Operating Conditions & Performance
The OM-CP-PROCESS101A operates within an ambient temperature range of -40 to 80°C (-40 to 176°F) and humidity levels from 0 to 95% RH non-condensing. The device features a nominal response time of 133 ms.
Performance specifications vary by input range as documented in the series data sheet:
- ±20 mA Range (Measurement: -2 to 30 mA): Resolution is 0.5 µA with an accuracy of ±0.05% FSR and overload protection up to ±316 mA.
- ±160 mA Range: Resolution is 5 µA with an accuracy of ±0.05% FSR and overload protection up to 1A.
- ±3A Range: Resolution is 100 µA with an accuracy of ±0.15% FSR and overload protection up to 6A.
All ranges share a common input voltage range of 0 to 2.5V. The device includes memory wrap functionality, allowing it to continue logging after reaching its 1 million reading storage capacity.
Power and Storage
- Power Supply: Powered by an included 3.6V lithium battery with a typical life of 10 years at a 15-minute reading rate.
- Status Indicators: A green LED blinks every 10 seconds to indicate logging or every 15 seconds in delay start mode; a red LED indicates low battery and/or full memory status.
Configuration Options
The series is available with three distinct measurement ranges, each configurable for specific current levels. The standard housing material is ABS Plastic with removable screw terminals for electrical connection.
- Housing Options: An optional weatherproof enclosure (OM-CP-WATERBOX101A) provides NEMA 4 (IP65) protection.
- Data Retrieval: Requires the OM-CP-IFC200 interface cable and Windows software for programming, data download, and real-time monitoring. The device connects via USB; a serial port is not used on this specific series model.
- Firmware & Security: Features field-upgradeable firmware and optional password protection.
Available Accessories and Options
- OM-CP-IFC200: Windows software and 1.8 m (6') USB interface cable.
- OM-CP-BAT105: Replacement 3.6V lithium battery.
- OM-CP-CONNECTOR-3: Replacement 3 position terminal block.
- NIST calibration certificate: Available for all ranges as an optional add-on (e.g., -CERT suffix).
Key Product Differences
The OM-CP-PROCESS101A Series is distinguished by its input range and the inclusion of a NIST calibration certificate. The base models are identified as follows:
- OM-CP-PROCESS101A-20MA: DC current data logger, ±20 mA range.
- OM-CP-PROCESS101A-160MA: DC current data logger, ±160 mA range.
- OM-CP-PROCESS101A-3A: DC current data logger, ±3A range.
Variants with the "-CERT" suffix (e.g., OM-CP-PROCESS101A-20MA-CERT) include a NIST calibration certificate. The series carries CE approval and offers an optional 5-year warranty.
Software & Drivers