Australia +61-2-4272 2055 | Welcome to our new website, found an issue or bug? Please report it here
CMDP-Series
Chilled Mirror Dew Point/Temperature Transmitter
Models In Stock
- Précision
- Deux sorties 4-20 mA (configurables)
- Durable
- Temps de réponse rapide
CMDP-Series
From
TWD
87,191
Models In Stock
Product Overview
- Précision ± 0.5 °C (± 0.9 °F)
- Fonctions supplémentaires 1 NIST Traceable
- Vitesse de l'air, max 900 m/mn
- Interface de données RS232
- Température du point de rosée, max 60 °C
- Température du point de rosée, min -20 °C
- Remarque sur la température du point de rosée Point de rosée -20°C (@ 25°C) à 60°C (sans condensation) Point de rosée -40°C (@ 0°C) Sonde de température : (optionnelle) - 40°C à 120°C
- Puissance électrique 4 à 20 mA
- Température de fonctionnement, max 120 °C
- Température de fonctionnement, min -40 °C
- Détails de la sortie (2) sorties 4 à 20 mA à 24 Vcc (configurables). Paramètres par défaut : - 20 °C à 60 °C (sauf indication contraire sur le bon de commande). Plage de sortie analogique point de rosée (4-20 mA) sélectionnable par l'utilisateur : 2) 4-20 mA - 24 Vcc (configurable) Réglage usine - les deux sorties, -20 °C à 60 °C (sauf indication contraire sur la commande)
- Limites de pression Jusqu'à 29 psi
- Pression, min 0 psi
- Répétabilité ± 0,05 °C
- Résolution 0,0001
- ID de série CMDP-Series
- Type d'approvisionnement électrique CC
- Tension d'alimentation 24 Vcc
- Mesure de la température, max 60 °C
- Mesure de la température, min -40 °C
- Poids 5 lb
Supported Applications
The Omega CMDP Series is a family of low-cost chilled mirror dew point transmitters configured as duct, wall, pipe, or OEM units. The series utilizes primary measurement principles to provide stable, accurate, and repeatable measurements traceable to NIST.
- Duct mounted applications (CMDP-D)
- Wall mounted applications (CMDP-W, CMDPT-W)
- Pipe mounted applications (CMDP-P)
Operating Conditions & Performance
The series operates within a process temperature range of -40°C to +60°C. Measurement capabilities include dew point from -20°C (@ 25°C) to 60°C (non-condensing), and down to -40°C when the sensor body is at 0°C. An optional air temperature probe supports measurements from -40°C to 120°C.
- Accuracy: ± 0.5°C for both dew point and temperature
- Repeatability: ± 0.05°C
- Resolution: 0.01 %
- Air Velocity: 50 to 900m/min
The unit requires a power supply of 24Vdc +/-3V @1.0A and has a pressure rating of 2 ATM (per data sheet) or 5 ATM (per manual). Wetted materials include Nickel, ultem, and anodized aluminum.
Configuration Options
The CMDP Series offers configurable mounting styles and measurement types. The series supports two 4-20mA outputs that are factory scaled for -20°C to 60°C unless specified on the purchase order. An RS232 communication connection is available via a 12-pin terminal block.
Mounting Configurations
- Duct mounted (CMDP-D)
- Wall mounted (CMDP-W, CMDPT-W)
- Pipe mounted (CMDP-P)
Sensor Options
An air temperature probe is available as an optional accessory for wall and OEM configurations. The probe connects via 4 Wire flying leads.
Key Product Differences
The series distinguishes models by mounting location and measurement capability:
- CMDP-D: Duct mounted chilled mirror hygrometer (dew point only)
- CMDP-W: Wall mounted chilled mirror hygrometer (dew point only)
- CMDPT-W: Wall mount chilled mirror hygrometer with dew point/air temperature
- CMDP-P: Pipe mounted chilled mirror hygrometer (dew point only)
Maintenance involves cleaning the mirror using a wetted cotton swab to restore normal operation. Settable service cycles are available for balancing and cleaning.
PDFs & Manuals