CMDP-Series

Chilled Mirror Dew Point/Temperature Transmitter

Models In Stock
  • Précision
  • Deux sorties 4-20 mA (configurables)
  • Durable
  • Temps de réponse rapide
CMDP-Series
From TWD 87,191
Models In Stock
Chilled Mirror Dew Point/Temperature Transmitter CMDP-Series CMDP-Series
loading all models
Product Overview
  • Précision ± 0.5 °C (± 0.9 °F)
  • Fonctions supplémentaires 1 NIST Traceable
  • Vitesse de l'air, max 900 m/mn
  • Interface de données RS232
  • Température du point de rosée, max 60 °C
  • Température du point de rosée, min -20 °C
  • Remarque sur la température du point de rosée Point de rosée -20°C (@ 25°C) à 60°C (sans condensation) Point de rosée -40°C (@ 0°C) Sonde de température : (optionnelle) - 40°C à 120°C
  • Puissance électrique 4 à 20 mA
  • Température de fonctionnement, max 120 °C
  • Température de fonctionnement, min -40 °C
  • Détails de la sortie (2) sorties 4 à 20 mA à 24 Vcc (configurables). Paramètres par défaut : - 20 °C à 60 °C (sauf indication contraire sur le bon de commande). Plage de sortie analogique point de rosée (4-20 mA) sélectionnable par l'utilisateur : 2) 4-20 mA - 24 Vcc (configurable) Réglage usine - les deux sorties, -20 °C à 60 °C (sauf indication contraire sur la commande)
  • Limites de pression Jusqu'à 29 psi
  • Pression, min 0 psi
  • Répétabilité ± 0,05 °C
  • Résolution 0,0001
  • ID de série CMDP-Series
  • Type d'approvisionnement électrique CC
  • Tension d'alimentation 24 Vcc
  • Mesure de la température, max 60 °C
  • Mesure de la température, min -40 °C
  • Poids 5 lb

Supported Applications

The Omega CMDP Series is a family of low-cost chilled mirror dew point transmitters configured as duct, wall, pipe, or OEM units. The series utilizes primary measurement principles to provide stable, accurate, and repeatable measurements traceable to NIST.

  • Duct mounted applications (CMDP-D)
  • Wall mounted applications (CMDP-W, CMDPT-W)
  • Pipe mounted applications (CMDP-P)

Operating Conditions & Performance

The series operates within a process temperature range of -40°C to +60°C. Measurement capabilities include dew point from -20°C (@ 25°C) to 60°C (non-condensing), and down to -40°C when the sensor body is at 0°C. An optional air temperature probe supports measurements from -40°C to 120°C.

  • Accuracy: ± 0.5°C for both dew point and temperature
  • Repeatability: ± 0.05°C
  • Resolution: 0.01 %
  • Air Velocity: 50 to 900m/min

The unit requires a power supply of 24Vdc +/-3V @1.0A and has a pressure rating of 2 ATM (per data sheet) or 5 ATM (per manual). Wetted materials include Nickel, ultem, and anodized aluminum.

Configuration Options

The CMDP Series offers configurable mounting styles and measurement types. The series supports two 4-20mA outputs that are factory scaled for -20°C to 60°C unless specified on the purchase order. An RS232 communication connection is available via a 12-pin terminal block.

Mounting Configurations

  • Duct mounted (CMDP-D)
  • Wall mounted (CMDP-W, CMDPT-W)
  • Pipe mounted (CMDP-P)

Sensor Options

An air temperature probe is available as an optional accessory for wall and OEM configurations. The probe connects via 4 Wire flying leads.

Key Product Differences

The series distinguishes models by mounting location and measurement capability:

  • CMDP-D: Duct mounted chilled mirror hygrometer (dew point only)
  • CMDP-W: Wall mounted chilled mirror hygrometer (dew point only)
  • CMDPT-W: Wall mount chilled mirror hygrometer with dew point/air temperature
  • CMDP-P: Pipe mounted chilled mirror hygrometer (dew point only)

Maintenance involves cleaning the mirror using a wetted cotton swab to restore normal operation. Settable service cycles are available for balancing and cleaning.