Models In Stock
  • Maximum temperature of 450°C (850°F)
  • Minimum temperature of 5°C (9°F) above ambient
  • An independent over-temperature cutout, 470°C (875°F)
  • Heat up time 25° C to 400°C with stability at 12 minutes
DBCL-400
From TWD 141,802
Models In Stock
Ambient 5 °C to 450 °C Temperature Dry Block Calibrator DBCL-400 DBCL-400
loading all models
Product Overview
  • Accuracy ± 0.4 °C
  • Accuracy Detail ±0.4°C (50 to 400°C) ±0.7°F (122 to 752°F) ±0.7°C (400 to 450°C) ±1.3°F (752 to 850°F)
  • Cool-Down Time Note Cool down 400°C to 100°C: 20 minutes
  • Dimensions Note 8.75 x 8 x 8 inches/222.25 x 203.2 x 203.2 mm *excluding the carrying strap
  • Display Resolution 0.1°
  • Frequency 50/60 Hz
  • Heat-Up Time Note Heat up time 25° C to 400°C: 12 minutes
  • Height 222.25 mm
  • Length 203.2 mm
  • Operating Temperature, Max 40 °C
  • Operating Temperature, Min 5 °C
  • Power Consumption 900 W
  • Process Temperature, Max 450 °C
  • Process Temperature, Min 5 °C
  • Process Temperature Note 5°C/9°F above ambient to 450°C/850°F;Temperature Uniformity: 0.020°C at 200°C & 0.030°C at 400°C
  • Relative Humidity, Max 0.95
  • Relative Humidity Note Up to 95% relative humidity, non-condensing
  • Series ID DBCL-400
  • Temperature Overload Protection Fan Cooling: Automatic; Overtemperature Limit: 470 °C
  • Temperature Stability ±0.050°C (50 to 400°C) ±0.090°F (122 to 752°F) (After 15 minutes)
  • Weight 5 kg
  • Width 203.2 mm

Supported Applications

The Omega DBCL-400 Ambient 5 °C to 450 °C Temperature Dry Block Calibrator provides a safe, dry, constant temperature source for checking and calibrating a wide range of temperature sensors, systems, indicators, and thermometers. Designed as both a benchtop unit and a portable field instrument weighing 11 lbs (5 Kg), the series utilizes a machined aluminum block as the heat transfer medium to establish stable thermal environments.

Operating Conditions & Performance

The DBCL-400 operates within an ambient temperature range of 5°C/9°F to 40°C/104°F and supports a measurement range from 5°C/9°F above ambient up to 450°C/850°F. An over-temperature limit protection circuit is set at 470°C/875°F. The unit features an automatic fan cooling system with the following thermal response characteristics:

  • Heat up time (25°C to 400°C): 12 minutes
  • Cool down (400°C to 100°C): 20 minutes

Accuracy specifications vary by temperature range: ±0.4°C (50 to 400°C) / ±0.7°F (122 to 752°F), and ±0.6°C (400 to 450°C) / ±1.0°C (752 to 850°F). Stability after 15 minutes is specified as ±0.050°C (50 to 400°C) / ±0.090°C (122 to 752°F). Well-to-well radial uniformity is documented at 0.015°C at 100°C and 0.052°C at 300°C. The display resolution is 0.1°. Quoted specifications apply for an ambient temperature range of 10°C/50°F to 30°C/86°F; performance may deteriorate outside this range while maintaining safe operation.

Configuration Options

The series supports two primary power supply configurations: 230V, 50/60Hz, 900W and 120V, 50/60Hz, 900W. Electrical connections utilize a power cable with plug; wire color codes are Live (Brown for 240V / Black for 120V), Neutral (Blue for 240V / White for 120V), and Earth ground (Green/yellow or Green). The unit requires earth grounding to ensure proper electrical safety.

Configuration options include various inserts, accessories, and replacement parts:

  • Inserts: Multiwell insert with 1/8", 3/16", 1/4", 5/16" & 3/8" holes (DBCL-400-3041); Blank insert (DBCL-400-3047); Inserts configured for specific hole combinations including 5 x 1/4" (DBCL-400-3043), mixed sizes such as 1 x 9/16" & 1 x 1/4" (DBCL-400-3048), and Blackbody Source insert for IR pyrometers (DBCL-400-3129).
  • Accessories: Unit carrying strap (4168), Insert extractor (4153), Carrying case (DBCL-3052), and region-specific power cables including UK 240V with 13amp plug (DBCL-UKCABLE) and Euro style R/A Schuko plug (4164).
  • Replacement Parts: Heaters, temperature controllers, PRTs, solid state relays, fuses (4 amp for 240V units; 8 amp for 120V units), heater blocks, and cooling fans.

The unit dimensions are 8.75 x 8 x 8 inches / 222.25 x 203.2 x 203.2 mm (excluding carrying strap). Immersion depth is specified as 4.5" (114.3mm).

Key Product Differences

The DBCL-400 series distinguishes between two voltage variants: the DBCL-400-120 for 120V operation and the DBCL-400-240 for 230V/240V operation. Units marked 230V on the rating plate operate at 220V, while units marked 120V operate at 110V; in both cases where voltage deviates from nominal ratings, the heating rate degrades by approximately 8%. The series includes a fused plug for UK use fitted with a 5 AMP fuse. No operator-maintainable parts are located within the equipment.

Accessories

accessories

SOFT SIDED CARRYING CASE W/ POCKETS FOR
TWD21,831 / each
0 In Stock
INSERT TO ACCEPT 5 X 1/4 INCH DIAMETER
TWD20,939 / each
2 In Stock
INSERT TO ACCEPT 2 X 1/4 AND 2 X 3/8 IN
TWD20,939 / each
0 In Stock
INSERT TO ACCEPT 2 X 1/4 AND 2 X 1/2 IN
TWD20,939 / each
2 In Stock
INSERT TO ACCEPT 1 X 1/4 INCHES DIAMETE
TWD20,939 / each
0 In Stock
INSERT BLANK
TWD17,177 / each
3 In Stock
INSERT TO ACCEPT 1 X 9/16 INCH AND 1 X 1
TWD20,939 / each
1 In Stock
INSERT TO ACCEPT 1 X 5/8 INCH AND 1 X 1/
TWD20,939 / each
0 In Stock
INSERT TO ACCEPT 1 X 11/16 INCH AND 1 X
TWD20,939 / each
2 In Stock
INSERT TO ACCEPT 1 X 3/4 INCH AND 1 X 1/
TWD20,939 / each
1 In Stock
en-TW
en_TW
Save Configuration