CDSN

Carbon Dioxide/Temperature transmitter, wall mount, universal current/voltage output, 2000 PPM

Choose a Variant

Measurement Type

AUD375.03 / each

Item# CDSN

Product Specs

<< See all Product Models
Brand
Carbon Dioxide Accuracy ± 40 PPM + 3% rdg
Carbon Dioxide Concentration, Max 2000 ppm
Carbon Dioxide Concentration, Min 0 ppm
Carbon Dioxide Response Time 120 s (τ90)
Display Type N/A
Electrical Connection Screw Terminal
Electrical Output 0 to 20 mA
Enclosure Ratings IP30, NEMA Type 1
GrossWeightValue 0.27
Materials ABS
Measurement Type Carbon Dioxide
Mounting Orientation N/A
Operating Temperature, Max 122 °F
Operating Temperature, Min 32 °F
Output Detail Field selectable; Current: 4-20 mA or 0-20 mA; Voltage: 0-5 V, 1-5 V, 0-10 V, or 2-10 V
Output Signal Current, Voltage
Relative Humidity 0 to 85% RH Non-Condensing
Sensor Life 15 Years
Sensor Type NDIR
Series ID Model-CDSN
Storage Temperature, Max 158 °F
Storage Temperature, Min -22 °F
Supply Power Note ± 20%, 50/60 Hz
Supply Power Type AC/DC
Supply Voltage 24 Vac/Vdc
Switch Type N/A
Temperature Accuracy ± 1 °C
Temperature Output Range N/A
Voltage Output Range 0 to 5 V, 1 to 5 V, 0 to 10 V, 2 to 10 V
Warm Up Time <2 mins Operational; 10 mins Max Accuracy
Country of Origin United States
See All Specs Show Less